Peter Huber Kältemaschinenbau Unichiller 015-H OLÉ
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 015-H OLÉ

Peter Huber Kältemaschinenbau Unichiller 015-H OLÉ — Recirculating chillers with Noise level 63 dB.

The Unichiller 015-H OLÉ is a single-phase, air-cooled recirculating chiller with an immersion pump and a rated operating window of -20 to 100 °C. Its 1.5 kW cooling capacity at 15 °C derates to 0.3 kW at -20 °C, so low-temperature duty must be sized against the load. It serves externally closed process loops where a compact, fixed-speed hydraulic circuit is sufficient.

Noise level
63 dB

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Noise level, dBControllerSafety classTemperature range, °CInterfacePower supply
Unichiller 015-H OLÉ Current 63OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 015 63Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 015-H 63Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller 015w-H 54Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 015w-H OLÉ 54OLÉI / NFL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 017T 64Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 017T-H 64Pilot ONEI / NFL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 017Tw-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller 020T —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 020T-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller 020Tw 60Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 020Tw-H 60Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 022-H 65Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 022-H OLÉ 65OLÉI / NFL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 022 OLÉ 65OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 022w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 022w-H —Pilot ONEIII / FL-10–100——
Unichiller 022w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 025-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 025-H OLÉ 64OLÉIII / FL-10–100USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller 025 OLÉ 64OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller 025T 74Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025T-H 74Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025Tw-H —Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 025w-H OLÉ 60OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 025w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P007 58Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208-240V 1~/2~ 50/60Hz
Unichiller P007-H —Pilot ONEIII / FL-20–100——
Unichiller P007-H OLÉ 58OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Unichiller P007 OLÉ 58OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Unichiller P010 57Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller P010-H 57Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P010-H OLÉ 57OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P010 OLÉ 57OLÉI / NFL-20–40USB (Device), RS232 Inte110-120V 1~ 50/60Hz
Unichiller P010w —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P010w-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012 59Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller P012-H 59Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012-H OLÉ 59OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P012 OLÉ 59OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P012w-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012w-H OLÉ —OLÉIII / FL-20–100USB (Device), RS232 Inte208V 2~ 60Hz
Unichiller P012w OLÉ —OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P015 63Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller P015-H 63Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P015-H OLÉ 63OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P015 OLÉ 63OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P015w 54Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller P015w-H 54Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P015w-H OLÉ 54OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50Hz
Unichiller P022 65Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022-H 65Pilot ONEI / NFL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022-H OLÉ 65OLÉI / NFL-10–100USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller P022 OLÉ 65OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P022w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P022w-H 60Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller P025 64Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P025-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P025-H OLÉ 64OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025 OLÉ 64OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P025w-H OLÉ 60OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz

Full specifications

Core Performance & Functional Specifications
Noise level 63 dB(A)
Maximum delivery flow rate 29 l/min (1 bar)
Pump connection G3/4 AG male
Pump speed adjustable no
Pump type Immersion pump
Temperature range -20...100 °C
Dimensions, Weight & Mechanical Design
Weight 61 kg
Electrical, Power & Drive Specifications
Fuse rating 16 A
Power supply 220-240V 1~ 50Hz
Interfaces, Connections & Control
Controller OLÉ
Display resolution 0, 1 K
Digital interface USB (Device), RS232 Interface
Temperature display OLED Display
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series position

Thermal behavior under load

Control, connection and operation

Integration & Operating Notes

Mounting
Allow 50–200 cm wall clearance per side for service access

Free-standing installation requires this side clearance so maintenance can be performed; the datasheet also directs the installer to respect the space requirements for airflow, since air-cooled units lose capacity if the condenser is starved.

Unichiller Operation Manual, Installation
Mounting
Leveling feet and parking brakes must be set before start-up

The unit must be immobilized before it is switched on: activate the caster parking brakes and/or unscrew the leveling feet. An unsecured 61 kg chiller can move during operation and cause crushing injury.

Unichiller Operation Manual, Commissioning
Electrical
Country-specific plug must be specified at order

Single-phase devices in the 100–240 V class are delivered with a power cable and a country-specific plug; the plug type is not universal and must be stated when ordering to avoid rework at commissioning.

Unichiller 015-H OLÉ Technical Data Sheet, Specifications
Process Limits
Only externally closed loops are permitted

The manual states the unit is designed for externally closed applications. Open baths or circuits that expose the fluid to atmosphere fall outside the intended use and should not be connected to this chiller.

Unichiller Operation Manual, Function Description
Process Limits
Avoid acetone and polyglycols with the plastic pump

The standard plastic pump is not resistant to acetone or polyglycols depending on manufacturer. For sub-zero duty, mix water with glysantine or ethylene glycol for freeze protection rather than running pure water.

Unichiller 015-H OLÉ Technical Data Sheet, Specifications
Process Limits
Do not exceed pump-connection torque; counterhold fittings

When attaching hoses to the G3/4 AG male pump connection, the maximum tightening torque from the manual table must not be exceeded and the connection must be counterheld to avoid transmitting torque into the pump housing.

Unichiller Operation Manual, Installation
Control & I/O
Set power-failure response before unattended operation

The OLÉ controller can be configured for OFF/Standby, ON, or Power Failure Auto after mains loss. The datasheet also lists an autostart function, so the chosen mode determines whether the process restarts automatically or stays off after recovery.

Unichiller Operation Manual, Normal Operation
Control & I/O
Pt100 process sensor is factory-fit only

An optional Pt100 sensor connection (#10519) can display, but not control, process temperature. It is available only factory fitted at additional charge and cannot be retrofitted in the field, so it must be decided at ordering.

Unichiller 015-H OLÉ Technical Data Sheet, Specifications
Environment
Indoor use only, 5–40 °C ambient, at least 300 lx

The unit is IP20 and may only be installed indoors at an illuminance of at least 300 lx and up to 2000 m altitude. It must not be installed or started inside an ATEX zone because it is not explosion-protected.

Unichiller Operation Manual, Installation
Environment
Check refrigerant charge against room volume

With 0.35 kg of R-449A (GWP 1397, 0.5 t CO2 equivalent), the refrigerant limit per m³ of ambient air must be checked against the room size at installation to avoid exceeding the permissible concentration.

Unichiller Operation Manual, Installation
Maintenance
Refrigerant circuit work is restricted to approved contractors

Any work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. Plan service access accordingly, since the 50–200 cm side clearance is what makes this work possible.

Unichiller Operation Manual, Maintenance

Selection & Integration Q&A

When should I choose the 015-H OLÉ rather than the 015w or a plain 015-H?

Pick the 015-H OLÉ when you need the OLÉ controller with OLED display, USB and RS232 and can accept 63 dB(A). The 015w is the quieter 54 dB(A) sibling for noise-sensitive rooms; the 015-H is the same frame without the OLÉ controller.

What is not included in base delivery that I must plan for?

The country-specific plug must be specified at order, and the Pt100 process sensor connection (#10519) is a factory-fit option only, not retrofittable. Hoses and mating fittings for the G3/4 AG male pump connection are not part of the chiller.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF
PDF
Unichiller 015-H OLÉ Technical Data Sheet
Datasheet ·87,8 КБ ·2 pp.
View PDF