Peter Huber Kältemaschinenbau Unichiller P007 OLÉ
Peter Huber Kältemaschinenbau Unichiller P007 OLÉ — Recirculating chillers with Noise level 58 dB.
The Unichiller P007 OLÉ is a compact recirculating chiller with a plastic pump (type B, non-adjustable) and G3/4 AG male connections, rated for -20 to 40 °C. Its 'P' designation marks it for high-pressure-drop loops, where the pump still delivers 6 l/min at 2.0 bar. It serves externally closed cooling circuits requiring 0.2–0.7 kW of heat removal.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Controller | Safety class | Temperature range, °C | Interface | Power supply |
|---|---|---|---|---|---|---|
| Unichiller P007 OLÉ Current | 58 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Unichiller 015 | 63 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 015-H | 63 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller 015-H OLÉ | 63 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 015w-H | 54 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 015w-H OLÉ | 54 | OLÉ | I / NFL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 017T | 64 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 017T-H | 64 | Pilot ONE | I / NFL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 017Tw-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller 020T | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 020T-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller 020Tw | 60 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 020Tw-H | 60 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 022-H | 65 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 022-H OLÉ | 65 | OLÉ | I / NFL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 022 OLÉ | 65 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 022w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 022w-H | — | Pilot ONE | III / FL | -10–100 | — | — |
| Unichiller 022w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 025-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 025-H OLÉ | 64 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller 025 OLÉ | 64 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller 025T | 74 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025T-H | 74 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025Tw-H | — | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 025w-H OLÉ | 60 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 025w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P007 | 58 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208-240V 1~/2~ 50/60Hz |
| Unichiller P007-H | — | Pilot ONE | III / FL | -20–100 | — | — |
| Unichiller P007-H OLÉ | 58 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Unichiller P010 | 57 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller P010-H | 57 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010-H OLÉ | 57 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010 OLÉ | 57 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 110-120V 1~ 50/60Hz |
| Unichiller P010w | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010w-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012 | 59 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller P012-H | 59 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012-H OLÉ | 59 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012 OLÉ | 59 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012w-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012w-H OLÉ | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208V 2~ 60Hz |
| Unichiller P012w OLÉ | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P015 | 63 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller P015-H | 63 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P015-H OLÉ | 63 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P015 OLÉ | 63 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P015w | 54 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller P015w-H | 54 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P015w-H OLÉ | 54 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50Hz |
| Unichiller P022 | 65 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022-H | 65 | Pilot ONE | I / NFL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022-H OLÉ | 65 | OLÉ | I / NFL | -10–100 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller P022 OLÉ | 65 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P022w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P022w-H | 60 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller P025 | 64 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P025-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P025-H OLÉ | 64 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025 OLÉ | 64 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P025w-H OLÉ | 60 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
Full specifications
| Noise level | 58 dB(A) |
| Pump connection | G3/4 AG male |
| Pump speed adjustable | no |
| Pump type | B |
| Temperature range | -20...40 °C |
| Weight | 59 kg |
| Power supply | 208-240V 1~/2~ 50/60Hz |
| Controller | OLÉ |
| Display resolution | 0, 1 K |
| Digital interface | USB (Device), RS232 Interface |
| Temperature display | OLED Display |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
- The P007 OLÉ is the smallest-capacity member of the Unichiller series shown here; siblings 015/015-H (with and without OLÉ controller) are rated 63 dB(A) versus 58 dB(A) for this unit.
- The 'P' in the model name designates suitability for applications with high pressure drop (Operation Manual, Function Description).
- Pump is type B, non-adjustable speed, with G3/4 AG male connections; enclosure protection IP20, safety class I / NFL, weight 59 kg.
- Controller is the OLÉ with OLED display, 0.1 K display resolution, USB (Device) and RS232 interfaces.
Thermal and hydraulic behavior
- Cooling capacity is strongly set-point dependent: 0.7 kW at 15 °C, 0.55 kW at 0 °C, 0.4 kW at -10 °C and 0.2 kW at -20 °C. Published figures are valid at 20 °C room temperature and fall as ambient rises.
- Pump curve: 21 l/min at 0.5 bar, 17 l/min at 1.0 bar, 11 l/min at 1.5 bar, 6 l/min at 2.0 bar. Loop pressure drop must be checked against this curve before selection.
- Operating temperature range -20 to 40 °C; ambient range 5 to 40 °C.
Installation, control and operation
- Supply is 208–240 V, 1~ or 2~, 50/60 Hz, max current 6.5 A, fuse 10–16 A. Voltage and frequency tolerances apply simultaneously; -5% voltage with +2% frequency is not permitted.
- Single/two-phase units ship with power cable and country-specific plug, which must be specified at ordering.
- The unit is designed for externally closed applications only; it is not explosion-protected and must not be installed in an ATEX zone.
- Refrigerant is R-449A (A1, H280), charge 0.3 kg, GWP 1397, 0.4 t CO2 equivalent; room size and charge must be checked at installation.
- Minimum filling capacity 3.8 l, expansion tank 1.7 l. For sub-zero duty, water should be mixed with glysantine or ethylene glycol for freeze protection; the plastic pump is not resistant to acetone or polyglycols.
Integration & Operating Notes
Flow collapses from 21 l/min at 0.5 bar to 6 l/min at 2.0 bar. Although the 'P' designation targets high-pressure-drop applications, the actual circuit resistance must be calculated and matched to the curve, otherwise the required cooling capacity at the set point cannot be transferred.
Voltage and frequency tolerances must be applied simultaneously, and the combination of -5% voltage with +2% frequency is explicitly not allowed. Verify the site supply against both limits before commissioning, and size upstream protection between 10 A and 16 A.
Free-standing installation requires a wall clearance of 50 to 200 cm on either side. The unit is IP20 and must be used indoors only, at illuminance of at least 300 lx and up to 2000 m altitude, within an ambient window of 5 to 40 °C.
The unit is not explosion-protected and must not be installed or started up within an ATEX zone. Refrigerant charge (0.3 kg R-449A) and room size must also be checked at installation to stay within the permitted refrigerant quantity per m³ of ambient air.
The type B plastic pump is not resistant to acetone or polyglycols (depending on manufacturer), so these fluids must not be used as heat transfer media. For sub-zero operation, water should be mixed with glysantine or ethylene glycol for freeze protection.
The power-failure response is configurable as OFF/Standby, ON, or Power Failure Auto via the Pilot ONE controller. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles.
Any work on the coolant/refrigerant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The unit is classified to DIN 12876 with NFL/FL identification depending on whether the thermal fluid is combustible; the rating is stated on the data sheet.
The cooling water supply must be shut off even for short downtimes such as overnight to prevent water damage and corrosion. Electrical lines directly below a water-cooled unit must be protected against liquid ingress from condensation.
Selection & Integration Q&A
Pick the P007 OLÉ for lower heat loads (0.2–0.7 kW depending on set point) and high-pressure-drop loops, where its 'P' designation applies. The 015/015-H siblings are louder (63 vs 58 dB(A)) and belong to a different capacity class; select them when the load exceeds the P007 envelope.
Check loop pressure drop against the pump curve (21 l/min at 0.5 bar down to 6 l/min at 2.0 bar) and provide at least 3.8 l filling volume with 1.7 l expansion allowance. Supply must be 208–240 V, 1~ or 2~, 50/60 Hz, 6.5 A max, fused 10–16 A; specify the country-specific plug at ordering.
Indoors only, 5–40 °C ambient, at least 300 lx illuminance, up to 2000 m altitude, with 50–200 cm wall clearance per side. Not permitted in ATEX zones. Leveling feet or parking brakes must be engaged before start-up, and cooling water shut off even overnight.
The base unit ships with a power cable and country-specific plug only if specified at ordering. Heat transfer fluid (water with glysantine or ethylene glycol for freeze protection) and the external loop are not included. Refrigerant circuit service requires approved contractors; no routine interval is stated in the supplied digests.