Peter Huber Kältemaschinenbau Unichiller 015-H
Peter Huber Kältemaschinenbau Unichiller 015-H — Recirculating chillers with Noise level 63 dB.
The Unichiller 015-H is an air-cooled recirculating chiller with an immersion pump and Pilot ONE controller, rated -20 to 100 °C. Within the 015 series it is the standard (non-OLÉ) variant; the -H suffix denotes this configuration. It serves externally closed cooling loops, delivering up to 29 l/min at 1 bar through G3/4 male connections.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Controller | Safety class | Temperature range, °C | Interface | Power supply |
|---|---|---|---|---|---|---|
| Unichiller 015-H Current | 63 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller 015 | 63 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 015-H OLÉ | 63 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 015w-H | 54 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 015w-H OLÉ | 54 | OLÉ | I / NFL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 017T | 64 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 017T-H | 64 | Pilot ONE | I / NFL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 017Tw-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller 020T | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 020T-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller 020Tw | 60 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 020Tw-H | 60 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 022-H | 65 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 022-H OLÉ | 65 | OLÉ | I / NFL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 022 OLÉ | 65 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 022w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 022w-H | — | Pilot ONE | III / FL | -10–100 | — | — |
| Unichiller 022w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 025-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 025-H OLÉ | 64 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller 025 OLÉ | 64 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller 025T | 74 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025T-H | 74 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025Tw-H | — | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 230V 1~ 50Hz |
| Unichiller 025w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller 025w-H OLÉ | 60 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller 025w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P007 | 58 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208-240V 1~/2~ 50/60Hz |
| Unichiller P007-H | — | Pilot ONE | III / FL | -20–100 | — | — |
| Unichiller P007-H OLÉ | 58 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Unichiller P007 OLÉ | 58 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Unichiller P010 | 57 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller P010-H | 57 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010-H OLÉ | 57 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010 OLÉ | 57 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 110-120V 1~ 50/60Hz |
| Unichiller P010w | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P010w-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012 | 59 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208V 2~ 60Hz |
| Unichiller P012-H | 59 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012-H OLÉ | 59 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012 OLÉ | 59 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012w-H | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50/60Hz |
| Unichiller P012w-H OLÉ | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208V 2~ 60Hz |
| Unichiller P012w OLÉ | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Unichiller P015 | 63 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller P015-H | 63 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P015-H OLÉ | 63 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P015 OLÉ | 63 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P015w | 54 | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 220-240V 1~/2~ 50Hz |
| Unichiller P015w-H | 54 | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P015w-H OLÉ | 54 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50Hz |
| Unichiller P022 | 65 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022-H | 65 | Pilot ONE | I / NFL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022-H OLÉ | 65 | OLÉ | I / NFL | -10–100 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller P022 OLÉ | 65 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P022w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P022w-H | 60 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P022w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
| Unichiller P025 | 64 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P025-H | 64 | Pilot ONE | III / FL | -10–100 | Ethernet, USB (Host u. D | 220-240V 1~ 50Hz |
| Unichiller P025-H OLÉ | 64 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025 OLÉ | 64 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025w | 60 | Pilot ONE | I / NFL | -10–40 | Ethernet, USB (Host u. D | 208-230V 1~/2~ 60Hz |
| Unichiller P025w-H OLÉ | 60 | OLÉ | III / FL | -10–100 | USB (Device), RS232 Inte | 220-240V 1~ 50Hz |
| Unichiller P025w OLÉ | 60 | OLÉ | I / NFL | -10–40 | USB (Device), RS232 Inte | 208-230V 1~/2~ 60Hz |
Full specifications
| Noise level | 63 dB(A) |
| Maximum delivery flow rate | 29 l/min (1 bar) |
| Pump connection | G3/4 AG male |
| Pump speed adjustable | Yes |
| Pump type | Immersion pump |
| External sensor connection | Pt100 |
| Temperature range | -20...100 °C |
| Weight | 55 kg |
| Fuse rating | 16 A |
| Power supply | 220-240V 1~/2~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Air-cooled refrigeration circuit using R-290 (A3, H220) natural refrigerant; IP20 enclosure, 55 kg, free-standing with casters/leveling feet.
- Immersion pump, speed-adjustable, G3/4 AG male pump connection; external Pt100 sensor input; Pilot ONE controller with Ethernet, USB (host and device) and RS232.
- Sibling models 015, 015 OLÉ and 015-H OLÉ share the 63 dB(A) noise figure; the 015w is the quieter variant at 54 dB(A). The -H designation distinguishes this build from the base 015.
Behavior and media
- Operating range -20 to 100 °C; minimum filling capacity 3.8 l with 1.7 l expansion volume.
- Pump curve: 29 l/min at 1 bar, 21 l/min at 0.5 bar — the achievable flow depends on the external loop's pressure drop.
- Plastic pump is not resistant to acetone or polyglycols; for freeze protection the manufacturer recommends mixing water with glysantine or ethylene glycol rather than pure water.
- Cooling capacity is quoted at 20 °C room temperature and falls as ambient rises; the unit is air-cooled and needs airflow clearance.
Installation, control and operation
- Supply 220-240 V 1~/2~ 50 Hz, max current 14.5 A, 16 A fuse; single/two-phase devices ship with power cable and country-specific plug (must be specified at ordering). Voltage/frequency tolerance envelope applies (-5% voltage with +2% frequency not allowed).
- Power-failure response configurable via Pilot ONE as OFF/Standby, ON, or Power Failure Auto. Process Safety mode releases maximum cooling capacity when the OT trigger value is reached instead of switching off all poles.
- EN55011 Class A Group 1 — intended for industrial electromagnetic environments.
Integration & Operating Notes
The unit is designed for externally closed applications, so the hydraulic circuit must be a sealed loop with the external Pt100 sensor placed in the process. Open-tank or atmospheric operation is outside the intended use and invalidates the rating.
The 220-240 V 1~/2~ 50 Hz supply must carry a maximum current of 14.5 A protected by a 16 A fuse. Combined voltage/frequency deviation outside the permitted envelope (-5% voltage with +2% frequency explicitly not allowed) must be avoided at the installation point.
Free-standing installation requires 50 to 200 cm wall clearance on either side for maintenance access, and the air-cooled condenser needs unobstructed airflow. Leveling feet must be unscrewed/activated and parking brakes applied before switch-on.
The unit is not explosion-protected and must not be installed or started in an ATEX zone. It is for indoor use only at an illuminance of at least 300 lx and up to 2000 m altitude; refrigerant charge versus room volume must be checked against the limit per m³ of ambient air.
The plastic immersion pump is not resistant to acetone or polyglycols (depending on manufacturer). For freeze protection, mix water with glysantine or ethylene glycol rather than using pure water, and plan the fluid charge within the 3.8 l minimum filling and 1.7 l expansion volume.
The Pilot ONE allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling capacity when the OT trigger value is reached instead of switching off all poles — select the mode that matches the process risk.
Any work on the coolant/refrigerant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The refrigerant is R-290 (A3, H220), flammable, which affects site safety classification and handling requirements.
Selection & Integration Q&A
Pick the 015-H for the standard build at 63 dB(A). Choose the 015w if 54 dB(A) noise matters more. OLÉ variants (015 OLÉ, 015-H OLÉ) are separate configurations; confirm their differences against the datasheet before substituting, since this card covers only the 015-H.
The pump connection is G3/4 AG male, so hoses/fittings must match that thread and be counterheld at the maximum tightening torque. The loop must be externally closed, with the external Pt100 sensor placed in the process, and its pressure drop checked against 29 l/min at 1 bar (21 l/min at 0.5 bar).
Indoor only, ≥300 lx, max 2000 m altitude, ambient 5-40 °C, no ATEX zone, IP20 (protect from water). Allow 50-200 cm wall clearance per side for service and airflow. Cooling capacity is rated at 20 °C room temperature and drops as ambient rises, so verify the heat load at the actual site ambient.
Single/two-phase units ship with a power cable and country-specific plug that must be specified at ordering; heat-transfer fluid and mating hoses are not part of the unit. Refrigerant-circuit work is restricted to approved refrigeration contractors; plan fluid preparation (water with glysantine or ethylene glycol) and periodic maintenance access per the manual.