Peter Huber Kältemaschinenbau Unichiller 160T-H
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 160T-H

Peter Huber Kältemaschinenbau Unichiller 160T-H — Recirculating chillers.

The Unichiller 160T-H is a recirculating chiller with a 16 kW cooling capacity at 15 °C, designed for externally closed applications. It operates from -10 to 100 °C with a 2 kW heater and a fixed-speed C3 pump delivering up to 99 l/min. This modification is distinguished within the series by its higher capacity and heating function, serving as a central cooling source for process loops.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Pump typeSafety classTemperature range, °CPower supplyExpansion volume, m³Minimum fill volume, m³
Unichiller 160T-H Current C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 110T C3I / NFL-10–40400V 3~ 50Hz0.007—
Unichiller 110T-H C3III / FL-10–100400V 3~ 50Hz0.007—
Unichiller 110Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 130T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 130T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 130Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 160T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 160Tw C3I / NFL-20–40400V 3~ 50Hz0.01—
Unichiller 180-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 180T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 180T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 180Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 180Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 180w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 180w-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 200T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 200Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 210T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 210T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 210Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 230-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 230w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 230w-H E1III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller 250T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 250T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 250Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 260T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 260T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 260Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 260Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 300T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 300T-H D3III / FL-20–100460V 3~ 60Hz0.0390.014
Unichiller 300Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 300Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 350T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 350Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 500Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw-H D3III / FL-20–100400V 3~ 50Hz0.007—
Unichiller 700Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 800Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller GL 1000w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.05—
Unichiller GL 200w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 250w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 300w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 400w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 750w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.05—
Unichiller P180 E2I / NFL-20–40—0.0480.018
Unichiller P180-H E2III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller P180w E2I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller P180w-H E2III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller P230 E2I / NFL-20–40460V 3~ 60Hz0.0480.018
Unichiller P230-H E2III / FL-20–100460V 3~ 60Hz0.0480.018

Full specifications

Core Performance & Functional Specifications
Expansion volume 39 l
Heating capacity 2 kW
Minimum fill volume 14 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type C3
External sensor connection Pt100
Temperature range -10...100 °C
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and Series Positioning

The Unichiller 160T-H is a free-standing, air-cooled recirculating chiller built for indoor installation. It is equipped with a 2 kW heater, enabling both cooling and heating within the -10 to 100 °C range. The pump is a fixed-speed C3 type with G1 1/4 male connections, delivering a maximum flow of 99 l/min and a maximum pressure of 5.9 bar. The unit uses R-449A refrigerant and has a 39 l expansion volume with a 14 l minimum fill. Compared to sibling models such as the Unichiller 110T-H, this model offers higher cooling capacity (16 kW at 15 °C) and is intended for larger heat loads. The 'T' designation indicates a standard (non-water-cooled) variant, while the 'H' denotes the integrated heater.

Performance and Operating Behavior

Cooling capacity is strongly temperature-dependent: 16 kW at 15 °C, 8.8 kW at 0 °C, and 4 kW at -10 °C. This means low-temperature duty requires careful sizing to avoid insufficient capacity. The unit is designed exclusively for externally closed circuits; open baths or tanks are not permitted. The Pilot ONE controller manages temperature and offers configurable power-failure response (OFF/Standby, ON, or Power Failure Auto). For exothermic processes, the Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached, rather than shutting down all poles.

Installation, Control and Operation

The chiller requires a 400 V three-phase 50 Hz supply and is delivered with a cable but without a plug (for currents below 63 A). It must be installed indoors in a dry location (IP20), with a wall clearance of 50–200 cm per side for maintenance. Ambient temperature must remain between 5 and 40 °C; cooling capacity is rated at 20 °C room temperature and may derate above that. The unit is EMC class A group 1 and may cause radio interference in residential areas. Refrigerant charge and room size must be checked to avoid exceeding the refrigerant limit per m³. Leveling feet or parking brakes must be engaged before operation. The Pilot ONE provides Ethernet, USB (host and device), and RS232 interfaces for remote control and data logging.

Integration & Operating Notes

Mounting
Maintain 50–200 cm side clearance for service access

The manual requires a wall clearance of 50 to 200 cm on either side of free-standing models to allow maintenance and ensure adequate airflow. Insufficient clearance can lead to overheating and reduced cooling capacity, and will hinder service access to the refrigerant circuit.

Operation Manual, Section 5.2 Installation Site
Electrical
Provide a 400 V 3~ 50 Hz supply with ±5% voltage tolerance

The unit requires a three-phase 400 V 50 Hz supply. Voltage tolerance is ±5% with a simultaneous frequency tolerance of ±2%, but not all combinations are permitted. The customer must provide a suitable disconnector or plug, as the unit is delivered with a cable but without a plug for currents below 63 A.

Technical Data, Electrical Specifications; Operation Manual, Section 4.2 Electrical Connection
Process Limits
Use only in externally closed circuits; open tanks are prohibited

The chiller is designed exclusively for externally closed applications. Connecting an open bath or tank will cause fluid evaporation, corrosion, and potential pump damage. Ensure the external circuit is sealed and compatible with the thermal fluid.

Operation Manual, Section 3.2 Intended Use; Technical Data, Application Note
Cooling & Media
Cooling capacity derates significantly at low temperatures

Cooling capacity drops from 16 kW at 15 °C to 4 kW at -10 °C. When sizing for low-temperature processes, the heat load must be reduced accordingly or a larger chiller selected. Capacity is rated at 20 °C ambient; higher ambient temperatures may further reduce performance.

Technical Data, Performance Curves; Operation Manual, Section 3.1 Performance Data
Safety
Configure power-failure response to match process risk

The Pilot ONE controller allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. For processes where automatic restart is hazardous, select OFF/Standby. For exothermic reactions, enable Process Safety mode to release maximum cooling on overtemperature instead of shutting down.

Operation Manual, Section 6.3 Power Failure; Section 6.4 Process Safety
Maintenance
Refrigerant circuit work requires certified contractors

Any work on the refrigerant circuit must be performed by approved refrigeration/air-conditioning system contractors. The refrigerant R-449A (GHS H280) requires proper ventilation and handling. Regular maintenance should include checking the refrigerant charge and room size to avoid exceeding the limit per m³.

Operation Manual, Section 7.1 Maintenance; Technical Data, Refrigerant

Selection & Integration Q&A

When should I choose the Unichiller 160T-H over the 110T-H or 110Tw-H?

Choose the 160T-H for higher cooling capacity (16 kW at 15 °C) and a 2 kW heater. The 110T-H offers lower capacity (likely around 11 kW) and is suitable for smaller loads. The 110Tw-H is water-cooled and produces less noise (65 dB(A) vs 74 dB(A)), making it preferable where noise or heat rejection to air is a concern.

What are the requirements for connecting the chiller to my process loop?

The pump connection is G1 1/4 male, so use mating hoses or adapters with G1 1/4 female threads. Ensure the external circuit is closed and can withstand the pump's maximum pressure of 5.9 bar. The minimum fill volume is 14 l, and the expansion volume is 39 l; verify your loop volume is compatible.

What ambient and installation conditions must be met for reliable operation?

Install indoors in a dry location (IP20), with ambient temperature between 5 and 40 °C. Maintain 50–200 cm clearance on each side for maintenance. Avoid ATEX zones. The unit is EMC class A group 1, so it may cause radio interference in residential areas; additional shielding may be needed.

What maintenance and accessories are required, and what is not included?

The unit is delivered with a cable but without a plug; you must provide a suitable disconnector. Refrigerant circuit work must be done by certified contractors. Regular maintenance includes checking refrigerant charge and room size. No hoses, adapters, or thermal fluid are included; these must be sourced separately to match G1 1/4 connections.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller 160T-H Technical Data
Datasheet ·185,9 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF