Peter Huber Kältemaschinenbau Unichiller 350T
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 350T

Peter Huber Kältemaschinenbau Unichiller 350T — Recirculating chillers.

The Unichiller 350T is an air-cooled, CFC/HCFC-free recirculating chiller rated 23 kW at 0 °C and 8 kW at -20 °C, covering -20…40 °C on a 460 V/60 Hz three-phase supply. It is the large-capacity member of the Unichiller series, intended for externally closed process loops requiring high heat rejection.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Pump typeSafety classTemperature range, °CPower supplyExpansion volume, m³Minimum fill volume, m³
Unichiller 350T Current D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 110T C3I / NFL-10–40400V 3~ 50Hz0.007—
Unichiller 110T-H C3III / FL-10–100400V 3~ 50Hz0.007—
Unichiller 110Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 130T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 130T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 130Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 160T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 160T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 160Tw C3I / NFL-20–40400V 3~ 50Hz0.01—
Unichiller 180-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 180T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 180T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 180Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 180Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 180w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 180w-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 200T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 200Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 210T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 210T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 210Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 230-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 230w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 230w-H E1III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller 250T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 250T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 250Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 260T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 260T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 260Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 260Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 300T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 300T-H D3III / FL-20–100460V 3~ 60Hz0.0390.014
Unichiller 300Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 300Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 350Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 500Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw-H D3III / FL-20–100400V 3~ 50Hz0.007—
Unichiller 700Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 800Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller GL 1000w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.05—
Unichiller GL 200w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 250w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 300w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 400w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 750w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.05—
Unichiller P180 E2I / NFL-20–40—0.0480.018
Unichiller P180-H E2III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller P180w E2I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller P180w-H E2III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller P230 E2I / NFL-20–40460V 3~ 60Hz0.0480.018
Unichiller P230-H E2III / FL-20–100460V 3~ 60Hz0.0480.018

Full specifications

Core Performance & Functional Specifications
Expansion volume 39 l
Minimum fill volume 14 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type D3
External sensor connection Pt100
Temperature range -20...40 °C
Dimensions, Weight & Mechanical Design
Weight 640 kg
Electrical, Power & Drive Specifications
Current consumption 27 A
Power supply 460V 3~ 60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling capacity 15 kW (at 35 °C)
at 0 °C: 23 kW · at -10 °C: 14 kW · at -20 °C: 8 kW
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

Behavior under load

Installation, control and operation

Integration & Operating Notes

Electrical
Provide UL 489 branch protection and final plug on site

The 460 V three-phase supply draws 27 A and requires 3×32 A fusing; external branch circuit protection according to UL 489 must be installed by the customer. Because the unit is below 63 A it ships with a cable but no plug, so the final connection hardware is a site responsibility.

Unichiller 350T Datasheet, pages 1–2
Electrical
Verify supply quality before commissioning

Voltage may deviate ±5 % and frequency ±2 %, but the two tolerances must not be combined in the unfavorable direction. Supply quality outside this envelope risks compressor and electronics damage, so the site supply should be measured before startup.

Unichiller 350T Datasheet, page 2
Mounting
Keep 50–200 cm clearance per side for service access

Free-standing installation requires a wall clearance of 50 to 200 cm on either side for maintenance. The air-cooled condenser also needs unobstructed airflow for heat rejection; restricted ventilation reduces cooling capacity and can cause overheating.

Unichiller Operation Manual, page 25; Datasheet page 1
Mounting
Check floor loading and access route for 640 kg unit

At 905 × 2172 × 1900 mm and 640 kg net, the installation route and floor must be verified before delivery. Leveling feet must be unscrewed/activated and parking brakes applied before the unit is switched on to prevent movement during operation.

Unichiller 350T Datasheet, page 1; Operation Manual page 31
Process Limits
Externally closed loops only; not for high pressure drop

The 350T is designed for externally closed applications. Unlike models with “P” in the name, it is not the variant intended for high pressure-drop process loops, so circuit pressure loss must be checked against the D3 pump characteristic before selection.

Unichiller Operation Manual, page 34
Process Limits
Do not install inside an ATEX zone

The unit is not explosion-protected and must not be installed or started up within an ATEX zone. Refrigerant charge and room size must also be checked at installation to stay within the permissible refrigerant quantity per m³ of ambient air.

Unichiller Operation Manual, pages 13 and 15
Control & I/O
Configure power-failure response for the process risk

The Pilot ONE allows the power-failure response to be set to OFF/Standby, ON or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling when the overtemperature trigger is reached instead of switching off all poles; the chosen logic must match the process hazard analysis.

Unichiller Operation Manual, pages 18–19
Maintenance
Refrigerant circuit work restricted to approved contractors

Any work on the refrigerant circuit must be carried out by approved refrigeration/air-conditioning contractors. The R-449A charge has a CO2 equivalent of 11.2 t, which triggers F-gas record-keeping and periodic leak-check obligations in many jurisdictions.

Unichiller Operation Manual, page 17; Datasheet page 1
Environment
Indoor, dry location with at least 300 lx lighting

IP20 protection means the unit must be installed indoors, protected from water and dust. The manual additionally requires illuminance of at least 300 lx and a site no higher than 2000 m altitude; the ambient temperature must stay within 5…40 °C.

Unichiller Operation Manual, page 25; Datasheet page 2

Selection & Integration Q&A

When should I choose the Unichiller 350T instead of the 110T or 110Tw variants?

Pick the 350T when the process heat load needs the larger capacity (23 kW at 0 °C, 8 kW at -20 °C). The 110-class siblings are smaller machines; the Tw versions are the quieter 65 dB(A) build versus 74 dB(A) on the 110T, so choose those when acoustic limits dominate.

What electrical infrastructure and protection must be provided for the 350T?

A 460 V 3~ 60 Hz supply with 27 A consumption and 3×32 A fusing, plus external branch protection to UL 489. The unit ships with a cable but no plug (below 63 A), so the final connection is site-supplied; verify ±5 % voltage and ±2 % frequency are not combined unfavorably.

What mounting and environmental limits apply to this chiller?

Install indoors only (IP20), at 5…40 °C ambient and at least 300 lx lighting, below 2000 m altitude, never in an ATEX zone. Keep 50–200 cm clearance per side for service, ensure airflow for the air-cooled condenser, and secure leveling feet/parking brakes before startup.

What maintenance obligations and site-supplied items come with the refrigerant charge?

The R-449A charge has a CO2 equivalent of 11.2 t, triggering F-gas record-keeping and leak checks; refrigerant work is restricted to approved contractors. The base delivery excludes the final plug and UL 489 breaker, and the external loop must suit the 14 l minimum fill and 39 l expansion volume.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller 350T Datasheet
Datasheet ·18,0 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF