Peter Huber Kältemaschinenbau Unichiller P230-H
Peter Huber Kältemaschinenbau Unichiller P230-H — Recirculating chillers.
The Unichiller P230-H is an air-cooled, externally closed recirculating chiller with a fixed-speed E2 pressure pump and G1 1/4 male connections, rated -20 to 100 °C. Its 'P' designation targets high-pressure-drop process loops, distinguishing it from the 110T/110Tw siblings. It serves as a central cooling source for closed process circuits requiring up to 23 kW at 20 °C.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Pump type | Safety class | Temperature range, °C | Power supply | Expansion volume, m³ | Minimum fill volume, m³ |
|---|---|---|---|---|---|---|
| Unichiller P230-H Current | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller 110T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 130T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 160T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.01 | — |
| Unichiller 180-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 180Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 180w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180w-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 200T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 200Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 210T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 230-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w-H | E1 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller 250T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 260T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 260Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 300T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 300T-H | D3 | III / FL | -20–100 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 300Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 300Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 350T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 350Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 500Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 700Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 800Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller GL 1000w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.05 | — |
| Unichiller GL 200w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 250w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 300w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 400w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 750w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.05 | — |
| Unichiller P180 | E2 | I / NFL | -20–40 | — | 0.048 | 0.018 |
| Unichiller P180-H | E2 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w | E2 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230 | E2 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.048 | 0.018 |
Full specifications
| Expansion volume | 48 l |
| Heating capacity | 2 kW |
| Minimum fill volume | 18 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | E2 |
| External sensor connection | Pt100 |
| Temperature range | -20...100 °C |
| Power supply | 460V 3~ 60Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Air-cooled refrigeration circuit using R-449A (A1, H280; 2.1 t CO2 equivalent) with an integrated 2 kW heater and a 48 l expansion volume.
- Fixed-speed E2 pressure pump (speed not adjustable) with G1 1/4 AG male connections; the 'P' in the model name marks it as suited to applications with high pressure drop, unlike the 110T/110Tw siblings.
- Pilot ONE controller with Ethernet, USB host/device and RS232; external Pt100 sensor input; protection IP20; safety class III/FL.
Behavior under load
- Cooling capacity is strongly temperature-dependent: 23 kW at 20 °C, 13.5 kW at 0 °C, 9 kW at -10 °C and 5.5 kW at -20 °C, so sizing must use the capacity at the actual operating temperature.
- The 2 kW heater supports the upper part of the -20…100 °C range; the unit is intended only for externally closed applications.
- Ambient temperature must remain 5…40 °C; capacity may drop above 20 °C, and air-flow clearance must be respected.
Installation, control and operation
- Requires 460 V 3~ 60 Hz supply; voltage/frequency tolerance combination of -5% voltage with +2% frequency is not allowed. External branch circuit protection to UL 489 is required; three-phase units below 63 A are delivered with cable but without plug.
- Power-failure response is configurable (OFF/Standby, ON, Power Failure Auto); Process Safety mode releases maximum cooling when the OT trigger is reached.
- Minimum filling capacity 18 l; pump connections must be counterheld and maximum tightening torques observed.
Integration & Operating Notes
The compressor and pump must not be operated with -5% voltage combined with +2% frequency; verify site supply quality before commissioning, because this combination is explicitly not allowed and can stress the refrigeration circuit.
Three-phase devices drawing less than 63 A arrive with a cable but no plug, and external branch circuit protection according to UL 489 must be provided by the installer; plan the panel and receptacle accordingly.
Free-standing installation requires 50 to 200 cm wall clearance on either side for maintenance, and the air-cooled condenser needs unobstructed airflow; insufficient clearance reduces cooling performance and risks overheating.
The unit is not explosion-protected and must not be installed or started inside an ATEX zone; it is for indoor use at illuminance of at least 300 lx and up to 2000 m altitude, with IP20 requiring protection from water and dust.
The chiller is designed only for externally closed applications; the loop must provide at least 18 l filling capacity and account for the 48 l expansion volume, and the G1 1/4 male pump connections must be counterheld at the stated maximum torque.
The Pilot ONE allows OFF/Standby, ON or Power Failure Auto after mains loss, and Process Safety mode releases maximum cooling at the overtemperature trigger instead of switching off; select the behavior that matches the exothermic risk of the process.
Any work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors; the R-449A charge of 2.1 t CO2 equivalent may also trigger leak-check and record-keeping obligations under refrigerant regulations.
Selection & Integration Q&A
Pick the P230-H for high-pressure-drop process loops, since the 'P' marks suitability for such applications, and when the -20…100 °C window with up to 23 kW at 20 °C fits. The 110T/110Tw siblings differ in noise level (74 vs 65 dB(A)) and should be compared on that basis.
A 460 V 3~ 60 Hz supply with UL 489 branch protection and a suitable plug, since the unit arrives with cable but no plug below 63 A. The process loop must be closed, match G1 1/4 male threads, and hold at least 18 l with 48 l expansion volume.
Install indoors, outside ATEX zones, at 5…40 °C ambient with at least 300 lx lighting and up to 2000 m altitude. Keep 50–200 cm wall clearance per side for maintenance and airflow, and activate leveling feet or parking brakes before start-up.
The plug and UL 489 breaker are site scope, and refrigerant-circuit service must be performed by approved contractors. The manual covers maintenance and shutdown procedures; hoses, couplings and thermal fluid for the G1 1/4 connections are not part of the unit delivery.