Peter Huber Kältemaschinenbau Unichiller 250T
Peter Huber Kältemaschinenbau Unichiller 250T — Recirculating chillers.
The Unichiller 250T is an air-cooled, externally closed recirculating chiller with a 6–18 kW cooling capacity window across −20 to +40 °C, built around a fixed-speed D3 pressure pump and Pilot ONE controller. It is the larger-capacity member of the Unichiller T series, distinguished from the 110T siblings by its higher heat-rejection envelope rather than by any change in hydraulic architecture.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Pump type | Safety class | Temperature range, °C | Power supply | Expansion volume, m³ | Minimum fill volume, m³ |
|---|---|---|---|---|---|---|
| Unichiller 250T Current | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 110T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 130T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 160T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.01 | — |
| Unichiller 180-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 180Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 180w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180w-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 200T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 200Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 210T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 230-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w-H | E1 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller 250T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 260T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 260Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 300T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 300T-H | D3 | III / FL | -20–100 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 300Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 300Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 350T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 350Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 500Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 700Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 800Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller GL 1000w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.05 | — |
| Unichiller GL 200w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 250w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 300w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 400w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 750w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.05 | — |
| Unichiller P180 | E2 | I / NFL | -20–40 | — | 0.048 | 0.018 |
| Unichiller P180-H | E2 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w | E2 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230 | E2 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
Full specifications
| Expansion volume | 39 l |
| Minimum fill volume | 14 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | D3 |
| External sensor connection | Pt100 |
| Temperature range | -20...40 °C |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling capacity | 15 kW (at 25 °C) at 0 °C: 18 kW · at -10 °C: 11 kW · at -20 °C: 6 kW |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
- Air-cooled refrigeration circuit using R-449A (A1, H280; GWP 1397), rejecting heat to the room rather than to a water supply.
- Stainless-steel pump (type D3, non-variable speed) with G1 1/4 male connections; one matching hose coupling supplied.
- 39 l expansion volume and 14 l minimum fill define the allowable external loop charge.
- Within the series, the 250T sits above the 110T/110T-H/110Tw/110Tw-H siblings; the T suffix denotes the standard (non-high-pressure-drop) hydraulic variant, so loops with high pressure drop are better served by a 'P' model rather than by this unit.
Behavior across the operating envelope
- Cooling capacity is not flat: 18 kW at 0 °C, 15 kW at 25 °C, 11 kW at −10 °C and 6 kW at −20 °C. Capacity is rated at 20 °C room temperature and falls as ambient rises toward the 40 °C limit.
- Temperature stability is 0.5 K at −10 °C; processes needing tighter control require a different model or added measures.
- Pump curve is finite: 210 l/min maximum flow and 4.7 bar maximum delivery pressure, with flow decreasing as loop resistance increases. The external circuit must be checked against this curve.
- The unit is designed for externally closed applications only; an open external circuit risks evaporation, contamination and overflow of the atmospheric expansion tank.
Installation, control and operation
- Supply is 400 V 3~ 50 Hz; three-phase units below 63 A ship with cable but without a plug, so site termination must be planned.
- Voltage tolerance ±5% and frequency ±2%, but the combination of −5% voltage with +2% frequency is explicitly not allowed.
- Pilot ONE controller provides Ethernet, USB host/device and RS232; a Pt100 external sensor input is available.
- Power-failure response is selectable (OFF/Standby, ON, Power Failure Auto); Process Safety mode releases maximum cooling when the overtemperature trigger is reached instead of switching all poles off.
- EMC Class A, Group 1 to EN55011 — industrial environment only; may interfere in residential areas. IP20, indoor use only.
Integration & Operating Notes
Air-cooled rejection means restricted airflow directly reduces capacity and can overheat the compressor. The manual requires 50 to 200 cm wall clearance on either side for maintenance, and leveling feet or caster brakes must be engaged before power-up to prevent the unit moving.
The drive and compressor tolerate ±5% voltage and ±2% frequency individually, but the specific pairing of −5% voltage and +2% frequency is prohibited. Verify the actual site supply combination before commissioning, since an out-of-tolerance combination can damage the compressor or pump.
Because the current draw is below 63 A, the 400 V 3~ 50 Hz unit is delivered with a cable only. The installer must provide a suitable plug or hard-wired termination and a matching protective device on site.
The unit is intended only for externally closed applications. Minimum fill is 14 l and expansion volume is 39 l; underfilling causes pump cavitation and poor control, while an open circuit risks evaporation and overflow of the atmospheric expansion tank.
Installation is restricted to indoors with illuminance of at least 300 lx and altitude up to 2000 m. IP20 offers no water protection, and operation outside the 5–40 °C ambient window impairs cooling or damages components. The unit is not explosion-protected and must not be placed in an ATEX zone.
Any work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. R-449A is A1 with H280 pressurised-gas hazard and GWP 1397, so leak detection, handling and disposal must follow those precautions and be reflected in service planning.
Selection & Integration Q&A
Pick the 250T when the heat load needs the 6–18 kW window across −20 to +40 °C. The 110T/110Tw siblings are lower-capacity; the Tw variants differ in noise level (65 vs 74 dB(A)). If the loop has high pressure drop, a 'P' model is the correct choice, not this standard T variant.
A 400 V 3~ 50 Hz supply within ±5% voltage and ±2% frequency (the −5%/+2% combination is prohibited), with a plug or termination since the unit ships cable-only. The external circuit must be closed, charged to at least 14 l with 39 l expansion allowance, and fitted to G1 1/4 male threads — only one hose coupling is included.
Indoors only, 5–40 °C ambient, ≥300 lx illuminance, up to 2000 m altitude, with 50–200 cm clearance per side for airflow and maintenance. Leveling feet or caster brakes must be engaged before start. The unit is not ATEX-rated and must not be installed in a hazardous zone.
Refrigerant-circuit service must be done by approved refrigeration contractors; R-449A (GWP 1397, H280) handling and leak checks are part of that schedule. The base delivery includes only one G1 1/4 hose coupling, so additional hoses, fittings and heat-transfer fluid must be sourced separately. Cooling-water shutoff applies only to water-cooled variants, not this air-cooled unit.