Peter Huber Kältemaschinenbau Unichiller 110Tw
Peter Huber Kältemaschinenbau Unichiller 110Tw — Recirculating chillers with Noise level 65 dB.
The Unichiller 110Tw is a water-cooled recirculating chiller designed for externally closed applications. It operates from -20 to 40 °C using the Pilot ONE controller and a fixed-speed C3 pump. Its 65 dB(A) noise level distinguishes it from the 74 dB(A) siblings, making it suitable for noise-sensitive installations.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Pump type | Safety class | Temperature range, °C | Power supply | Expansion volume, m³ | Minimum fill volume, m³ |
|---|---|---|---|---|---|---|
| Unichiller 110Tw Current | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 110T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 130T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 160T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.01 | — |
| Unichiller 180-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 180Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 180w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180w-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 200T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 200Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 210T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 230-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w-H | E1 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller 250T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 260T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 260Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 300T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 300T-H | D3 | III / FL | -20–100 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 300Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 300Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 350T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 350Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 500Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 700Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 800Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller GL 1000w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.05 | — |
| Unichiller GL 200w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 250w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 300w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 400w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 750w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.05 | — |
| Unichiller P180 | E2 | I / NFL | -20–40 | — | 0.048 | 0.018 |
| Unichiller P180-H | E2 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w | E2 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230 | E2 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
Full specifications
| Noise level | 65 dB(A) |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | C3 |
| External sensor connection | Pt100 |
| Temperature range | -20...40 °C |
| Weight | 225 kg |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water consumption | 89 l/h (cooling water 15°C, flow -10°C) cooling water 15°C, flow 0°C: 191 l/h · cooling water 15°C, flow 15°C: 400 l/h |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and positioning
- The Unichiller 110Tw is a water-cooled chiller intended for externally closed circuits. It uses refrigerant R-449A (1.75 kg, 2.4 t CO2e) and is classified as pressure equipment category I.
- Within the Unichiller series, the 'w' suffix indicates water cooling. The 110Tw is the quiet variant (65 dB(A)) compared to the 110T, 110T-H, and 130T (all 74 dB(A)). The 110Tw-H offers the same noise level but includes a heater.
- The unit is not explosion-protected and must not be installed in ATEX zones. It is for indoor use only, with a minimum illuminance of 300 lx and maximum altitude of 2000 m.
Behavior under load and media
- The temperature range is -20 to 40 °C, with an ambient temperature limit of 5 to 40 °C. Cooling water consumption varies with flow temperature: 191 l/h at 0 °C, 400 l/h at 15 °C, and 89 l/h at -10 °C, based on a cooling water supply of 15 °C.
- The cooling water circuit requires a differential pressure of at least 1 bar and a maximum pressure of 6 bar. The cooling water intake temperature should be between 5 °C and 20 °C; below the condensation point, internal cooling water pipes must be insulated.
- The C3 pump delivers a maximum of 96 l/min at 5.6 bar, with flow decreasing as pressure rises. The pump speed is not adjustable. The unit requires a minimum filling capacity of 6.5 l and has an expansion volume of 5.5 l.
Installation, connection, and control
- Electrical supply is 400 V 3~ 50 Hz with a maximum current of 13.5 A and a 3x16 A fuse. The unit is delivered with a cable but without a plug. Voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%.
- Pump connections are G1 1/4 male; cooling water connections are G3/4 male. Maximum tightening torques for pump connections must be observed, and connections must be counterheld.
- The Pilot ONE controller provides Ethernet, USB (Host and Device), and RS232 interfaces. A Pt100 external sensor connection is available. The power-failure response can be configured as OFF/Standby, ON, or Power Failure Auto. Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached.
- Installation requires a wall clearance of 50 to 200 cm per side for maintenance. Leveling feet must be extended and parking brakes applied before operation. Electrical lines directly below the unit must be protected against liquid ingress.
Integration & Operating Notes
Free-standing installation requires a wall clearance of 50 to 200 cm on either side to allow maintenance access. Insufficient clearance can prevent servicing of the refrigerant circuit and pump.
The unit draws up to 13.5 A at 400 V 3~ 50 Hz and requires a 3x16 A fuse. It is delivered with a cable but without a plug, so the installer must provide a compatible plug. Voltage and frequency tolerances are only valid in combination (±5% voltage with ±2% frequency).
The cooling water supply must have a differential pressure between 1 and 6 bar and an intake temperature between 5 °C and 20 °C. Below the condensation point, internal cooling water pipes must be insulated. The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE, and EPDM, so only compatible water may be used.
The unit is designed for externally closed circuits. Open baths or circuits that do not suit the expansion tank design must not be used. The minimum filling capacity is 6.5 l, and the expansion volume is 5.5 l.
The Pilot ONE controller allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached, instead of switching off all poles.
Work on the refrigerant circuit must only be performed by approved refrigeration/air-conditioning system contractors. The refrigerant is R-449A with a GWP of 1397 and a charge of 1.75 kg (2.4 t CO2e), so F-gas regulations apply.
The cooling water supply must be shut off even for short downtimes such as overnight to prevent water damage and corrosion.
Selection & Integration Q&A
Choose the 110Tw when water cooling is available and low noise (65 dB(A)) is required. The 110T is air-cooled and louder (74 dB(A)). The 110Tw-H adds a heater for higher temperature control but has the same noise level.
Cooling water: 5–20 °C, differential pressure 1–6 bar, connection G3/4 male. Process circuit: pump connections G1 1/4 male, minimum filling volume 6.5 l, expansion volume 5.5 l. Ensure water quality is compatible with copper, stainless steel 1.4401, MS, PA, PPE, PTFE, and EPDM.
Install indoors only, at 5–40 °C ambient, with 50–200 cm side clearance for maintenance. Leveling feet must be extended and parking brakes applied. Do not install in ATEX zones. Electrical lines below the unit must be protected from liquid ingress.
Refrigerant circuit work must be by certified contractors. Shut off cooling water during downtimes. A plug for the 400 V cable must be provided on site. No other accessories are specified in the provided documents.