Peter Huber Kältemaschinenbau Unichiller 200Tw-H
Peter Huber Kältemaschinenbau Unichiller 200Tw-H — Recirculating chillers.
The Unichiller 200Tw-H is a water-cooled recirculating chiller with a 20 kW cooling capacity at 15 °C and a -20…100 °C working range, controlled by the Pilot ONE. It is the high-capacity modification in the Unichiller series, distinguished by its 200-size refrigeration circuit and 2 kW heating stage, intended for externally closed process loops requiring stable temperature control.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Pump type | Safety class | Temperature range, °C | Power supply | Expansion volume, m³ | Minimum fill volume, m³ |
|---|---|---|---|---|---|---|
| Unichiller 200Tw-H Current | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 110T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 130T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 160T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.01 | — |
| Unichiller 180-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 180Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 180w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180w-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 200T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 210T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 230-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w-H | E1 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller 250T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 260T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 260Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 300T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 300T-H | D3 | III / FL | -20–100 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 300Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 300Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 350T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 350Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 500Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 700Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 800Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller GL 1000w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.05 | — |
| Unichiller GL 200w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 250w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 300w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 400w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 750w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.05 | — |
| Unichiller P180 | E2 | I / NFL | -20–40 | — | 0.048 | 0.018 |
| Unichiller P180-H | E2 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w | E2 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230 | E2 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
Full specifications
| Expansion volume | 22 l |
| Heating capacity | 2 kW |
| Minimum fill volume | 15 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | D3 |
| External sensor connection | Pt100 |
| Temperature range | -20...100 °C |
| Weight | 374 kg |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling capacity | 20 kW (at 15 °C) at 0 °C: 11 kW · at -10 °C: 5 kW · at -20 °C: 3 kW |
| Cooling water connection | G3/4 male |
| Cooling water consumption | 445 l/h (cooling water at 15 °C, flow at -20 °C) cooling water at 15 °C, flow at -10 °C: 610 l/h · cooling water at 15 °C, flow at 0 °C: 860 l/h · cooling water at 15 °C, flow at 15 °C: 725 l/h |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The 200Tw-H is a water-cooled, floor-standing chiller with a 374 kg net weight and 760×800×1615 mm footprint. It uses refrigerant R-449A (2.35 kg, 3.3 t CO₂e) and is classified safety class III / FL per DIN 12876.
- Within the Unichiller series, the 200Tw-H is the largest sibling listed; the 110T/110Tw variants are smaller-capacity units. The 'Tw' designation indicates water cooling, and the '-H' suffix denotes the integrated 2 kW heating stage.
- The hydraulic circuit includes a D3 pump (non-speed-adjustable) with G1 1/4 AG male process connections and G3/4 male cooling-water connections. Minimum fill volume is 15 l; expansion volume is 22 l.
Performance and load behavior
- Cooling capacity is strongly temperature-dependent: 20 kW at 15 °C, 11 kW at 0 °C, 5 kW at -10 °C, and 3 kW at -20 °C. Required cooling-water flow at 15 °C cooling-water inlet rises from 445 l/h (at -20 °C process) to 860 l/h (at 0 °C process), reflecting the higher heat rejection at warmer process temperatures.
- The pump delivers up to 210 l/min at 4.7 bar maximum; flow falls sharply at higher system pressure drop (e.g., 31 l/min at 4.5 bar), so external circuit pressure drop must be matched to the pump curve.
- The unit is intended for externally closed applications only; open external baths are not permitted.
Installation, control and operation
- The Pilot ONE controller provides Ethernet, USB (host and device) and RS232 interfaces, plus a Pt100 external sensor connection. Power-failure response is configurable (OFF/Standby, ON, or Power Failure Auto).
- Process Safety mode releases maximum cooling capacity when the adjustable overtemperature trigger is reached, rather than switching off all poles—critical for exothermic processes.
- Electrical supply is 400 V 3~ 50 Hz; three-phase units under 63 A are supplied with cable but without a plug. Voltage/frequency tolerances are ±5%/±2%, but -5% voltage with +2% frequency is not allowed.
- Cooling water must be 5–20 °C inlet, 1–6 bar differential pressure; if cooling water falls below the condensation point, internal cooling-water pipes must be insulated. Electrical lines directly below the unit must be protected against condensate.
Integration & Operating Notes
Operating with cooling water outside this window risks reduced capacity or condensation; if the cooling water temperature falls below the dew point, the internal cooling-water pipes must be insulated to prevent condensate damage.
The unit is designed exclusively for externally closed applications; an open bath will cause fluid loss and unsafe operation. The external loop must hold at least 15 l, and the 22 l expansion volume must be accommodated in the circuit design.
The supply must remain within ±5% voltage and ±2% frequency, but the specific combination of -5% voltage and +2% frequency is explicitly prohibited and may cause malfunction or damage to the compressor and controls.
Because the current draw is below 63 A, the unit arrives with a power cable but without a plug; a suitable 400 V 3~ plug must be procured and fitted on site by a qualified electrician before commissioning.
Free-standing installation requires this clearance for maintenance access and airflow; failing to provide it will obstruct service work on the refrigeration circuit and may affect heat rejection from the water-cooled condenser.
The unit is not explosion-protected and must not be installed in an ATEX zone. It is restricted to indoor locations with a minimum illuminance of 300 lx and a maximum altitude of 2000 m; ambient temperature must remain between 5 and 40 °C.
Before switching on, the casters' parking brakes must be applied and/or the leveling feet unscrewed to stabilize the 374 kg unit; otherwise vibration or accidental movement during operation can cause crushing injuries or connection leaks.
In Process Safety mode, the controller releases maximum cooling capacity when the adjustable overtemperature trigger is reached instead of cutting all poles; this prevents runaway temperatures in exothermic processes but requires the OT trigger to be set appropriately for the application.
The Pilot ONE can be set to OFF/Standby, ON, or Power Failure Auto after a mains interruption; the choice determines whether the process restarts automatically or remains safe, and must be aligned with the process hazard analysis.
Any intervention on the R-449A circuit (2.35 kg, 3.3 t CO₂e) must be performed by certified refrigeration/air-conditioning contractors; unauthorized work risks refrigerant release and voids compliance with H280 safety obligations.
The cooling water supply must be closed during short idle periods to prevent water damage and corrosion inside the unit; this is especially important for water-cooled models where residual pressure can cause leaks.
Selection & Integration Q&A
Choose the 200Tw-H when you need up to 20 kW cooling at 15 °C and a 2 kW heating stage for externally closed loops; the 110T/110Tw siblings offer lower capacity and are air-cooled (110T) or water-cooled (110Tw) without the -H heating option. The 200Tw-H is the high-capacity water-cooled variant.
Cooling water: 5–20 °C inlet, 1–6 bar differential pressure, G3/4 male connections; flow required up to 860 l/h at 0 °C process. Process loop: G1 1/4 AG male pump connections, minimum 15 l fill, 22 l expansion volume, and the loop must be externally closed.
Indoor use only, 5–40 °C ambient, at least 300 lx illuminance, max 2000 m altitude, IP20 (dry location). Maintain 50–200 cm wall clearance per side for maintenance. Not for ATEX zones. Leveling feet and parking brakes must be engaged before start-up.
A 400 V 3~ plug must be procured (cable supplied without plug). Refrigerant circuit work requires approved contractors. Cooling water must be shut off during downtime. No external sensor is included; a Pt100 can be connected. Regular maintenance per manual; no consumables listed in base delivery.