Peter Huber Kältemaschinenbau Unichiller 180w-H
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 180w-H

Peter Huber Kältemaschinenbau Unichiller 180w-H — Recirculating chillers.

The Unichiller 180w-H is a water-cooled recirculating chiller with a pressure pump (E1, non-variable speed) and Pilot ONE controller, covering -20 to 100 °C. It is the 'w' (water-cooled) variant of the 180 platform, distinguished from air-cooled siblings by its G3/4 cooling-water circuit. It serves externally closed process loops requiring up to 54 l/min delivery.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Pump typeSafety classTemperature range, °CPower supplyExpansion volume, m³Minimum fill volume, m³
Unichiller 180w-H Current E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 110T C3I / NFL-10–40400V 3~ 50Hz0.007—
Unichiller 110T-H C3III / FL-10–100400V 3~ 50Hz0.007—
Unichiller 110Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 130T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 130T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 130Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 160T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 160T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 160Tw C3I / NFL-20–40400V 3~ 50Hz0.01—
Unichiller 180-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 180T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 180T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 180Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 180Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 180w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 200T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 200Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 210T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 210T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 210Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 230-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 230w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 230w-H E1III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller 250T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 250T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 250Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 260T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 260T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 260Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 260Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 300T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 300T-H D3III / FL-20–100460V 3~ 60Hz0.0390.014
Unichiller 300Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 300Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 350T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 350Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 500Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw-H D3III / FL-20–100400V 3~ 50Hz0.007—
Unichiller 700Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 800Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller GL 1000w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.05—
Unichiller GL 200w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 250w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 300w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 400w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 750w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.05—
Unichiller P180 E2I / NFL-20–40—0.0480.018
Unichiller P180-H E2III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller P180w E2I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller P180w-H E2III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller P230 E2I / NFL-20–40460V 3~ 60Hz0.0480.018
Unichiller P230-H E2III / FL-20–100460V 3~ 60Hz0.0480.018

Full specifications

Core Performance & Functional Specifications
Expansion volume 48 l
Heating capacity 2 kW
Maximum delivery flow rate 54 l/min (pressure pump max; also 3, 5 bar)
Minimum fill volume 18 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type E1
External sensor connection Pt100
Temperature range -20...100 °C
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Working range and behavior

Installation, control and operation

Integration & Operating Notes

Electrical
Three-phase feed must be within ±5% V and ±2% Hz together

The 400 V 3~ 50 Hz supply must stay within ±5% voltage only when frequency is simultaneously within ±2%; certain combinations are not permitted. Verify the site feed before commissioning, and note that units below 63 A arrive with a cable but no plug, so the customer must fit one.

Unichiller 180w-H Datasheet, page 2
Cooling & Media
Cooling water must enter between 5 °C and 20 °C

The unit is designed for cooling-water intake of 5...20 °C. Higher inlet temperature lowers cooling power and may require a higher cooling-water flow rate, so size the water circuit for the actual site conditions. Insulate internal cooling-water pipes when water temperature falls below the condensation point.

Unichiller 180w-H Datasheet, page 2
Mounting
Keep 50–200 cm clearance per side for maintenance access

Free-standing models require a wall clearance of 50 to 200 cm on either side. The unit is IP20 and must be installed indoors in a dry, protected location with illuminance of at least 300 lx and at altitudes up to 2000 m. It must not be installed or started inside an ATEX zone.

Unichiller Operation Manual, page 25
Process Limits
Only externally closed circuits are permitted

The chiller is designed for externally closed applications only; open baths or unsealed loops are outside its intended use. Models with 'P' in the name are the ones suited to high pressure drops — this 180w-H carries the standard E1 pressure pump, so verify the pressure drop of the connected loop against the pump curve.

Unichiller Operation Manual, page 34
Operation
Configure power-failure response before unattended runs

The Pilot ONE allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles — select the mode that matches the process risk.

Unichiller Operation Manual, pages 18–19
Maintenance
Refrigerant circuit work requires certified contractors

Any work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The R-449A charge has a CO2 equivalent of 2.1 t, which may trigger periodic leak checks under F-gas rules; keep records accordingly.

Unichiller Operation Manual, page 17; Datasheet, page 1
Operation
Shut off cooling water even for overnight downtime

Close the cooling-water supply to the unit for any downtime, including overnight, to prevent water damage and corrosion. Before switching on, activate parking brakes on casters and/or unscrew the leveling feet so the unit cannot move during operation.

Unichiller Operation Manual, pages 28, 31

Selection & Integration Q&A

When should I choose the 180w-H over the air-cooled 110T/110T-H siblings?

Pick the 180w-H when a 5–20 °C cooling-water supply is available and higher capacity is needed; the 'w' denotes water cooling via G3/4 connections. Air-cooled 110T variants avoid a water circuit but are limited to their own capacity and noise class (74 dB(A) for 110T, 65 dB(A) for 110Tw).

What supply and mating connections must be ready before installation?

Provide a 400 V 3~ 50 Hz feed within ±5% V / ±2% Hz; units below 63 A come with cable but no plug. Process side needs G1 1/4 male pump fittings, cooling water G3/4 male, and at least 18 l fill volume with 48 l expansion volume.

What site and duty-cycle limits apply to the 180w-H?

Install indoors only, 5–40 °C ambient, IP20, at least 300 lx, up to 2000 m altitude, with 50–200 cm clearance per side. Not permitted in ATEX zones. Shut off cooling water for any downtime, including overnight, and secure casters/leveling feet before start-up.

What maintenance obligations and delivery exclusions should be planned?

Refrigerant-circuit work must be done by approved refrigeration contractors; the R-449A charge (CO2 equivalent 2.1 t) may require periodic leak checks. The three-phase plug is not included, and the customer must supply compatible cooling-water and process fittings for the G3/4 and G1 1/4 male connections.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller 180w-H Datasheet
Datasheet ·18,2 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF