Peter Huber Kältemaschinenbau Unichiller 180w-H
Peter Huber Kältemaschinenbau Unichiller 180w-H — Recirculating chillers.
The Unichiller 180w-H is a water-cooled recirculating chiller with a pressure pump (E1, non-variable speed) and Pilot ONE controller, covering -20 to 100 °C. It is the 'w' (water-cooled) variant of the 180 platform, distinguished from air-cooled siblings by its G3/4 cooling-water circuit. It serves externally closed process loops requiring up to 54 l/min delivery.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Pump type | Safety class | Temperature range, °C | Power supply | Expansion volume, m³ | Minimum fill volume, m³ |
|---|---|---|---|---|---|---|
| Unichiller 180w-H Current | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 110T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 130T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 160T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.01 | — |
| Unichiller 180-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 180Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 180w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 200T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 200Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 210T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 230-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w-H | E1 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller 250T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 260T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 260Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 300T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 300T-H | D3 | III / FL | -20–100 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 300Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 300Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 350T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 350Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 500Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 700Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 800Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller GL 1000w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.05 | — |
| Unichiller GL 200w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 250w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 300w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 400w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 750w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.05 | — |
| Unichiller P180 | E2 | I / NFL | -20–40 | — | 0.048 | 0.018 |
| Unichiller P180-H | E2 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w | E2 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230 | E2 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
Full specifications
| Expansion volume | 48 l |
| Heating capacity | 2 kW |
| Maximum delivery flow rate | 54 l/min (pressure pump max; also 3, 5 bar) |
| Minimum fill volume | 18 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | E1 |
| External sensor connection | Pt100 |
| Temperature range | -20...100 °C |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Water-cooled condenser with G3/4 male cooling-water connections; refrigerant R-449A (A1, H280), CO2 equivalent 2.1 t.
- Pressure pump type E1, speed not adjustable, G1 1/4 AG male pump connection; max delivery flow rate 54 l/min.
- Pilot ONE controller with Ethernet, USB (host and device) and RS232; external Pt100 sensor input.
- Within the Unichiller series, the 'w' suffix denotes water cooling (cf. air-cooled 110T/110T-H siblings); the 180 platform is the larger-capacity modification.
Working range and behavior
- Temperature range -20...100 °C; heating capacity 2 kW.
- Cooling water inlet must be 5...20 °C; higher cooling-water temperature reduces cooling power and may increase required cooling-water flow.
- Ambient operating range 5...40 °C; IP20, indoor use only.
- Minimum fill volume 18 l; expansion volume 48 l.
Installation, control and operation
- Three-phase 400 V 3~ 50 Hz supply; devices below 63 A are delivered with cable but without plug.
- Voltage tolerance ±5% only with simultaneous frequency tolerance ±2%.
- EMC Class A, Group 1 to EN55011 — intended for non-domestic environments.
- Power-failure response configurable (OFF/Standby, ON, Power Failure Auto); Process Safety mode releases maximum cooling on OT trigger.
- Cooling-water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.
Integration & Operating Notes
The 400 V 3~ 50 Hz supply must stay within ±5% voltage only when frequency is simultaneously within ±2%; certain combinations are not permitted. Verify the site feed before commissioning, and note that units below 63 A arrive with a cable but no plug, so the customer must fit one.
The unit is designed for cooling-water intake of 5...20 °C. Higher inlet temperature lowers cooling power and may require a higher cooling-water flow rate, so size the water circuit for the actual site conditions. Insulate internal cooling-water pipes when water temperature falls below the condensation point.
Free-standing models require a wall clearance of 50 to 200 cm on either side. The unit is IP20 and must be installed indoors in a dry, protected location with illuminance of at least 300 lx and at altitudes up to 2000 m. It must not be installed or started inside an ATEX zone.
The chiller is designed for externally closed applications only; open baths or unsealed loops are outside its intended use. Models with 'P' in the name are the ones suited to high pressure drops — this 180w-H carries the standard E1 pressure pump, so verify the pressure drop of the connected loop against the pump curve.
The Pilot ONE allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles — select the mode that matches the process risk.
Any work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The R-449A charge has a CO2 equivalent of 2.1 t, which may trigger periodic leak checks under F-gas rules; keep records accordingly.
Close the cooling-water supply to the unit for any downtime, including overnight, to prevent water damage and corrosion. Before switching on, activate parking brakes on casters and/or unscrew the leveling feet so the unit cannot move during operation.
Selection & Integration Q&A
Pick the 180w-H when a 5–20 °C cooling-water supply is available and higher capacity is needed; the 'w' denotes water cooling via G3/4 connections. Air-cooled 110T variants avoid a water circuit but are limited to their own capacity and noise class (74 dB(A) for 110T, 65 dB(A) for 110Tw).
Provide a 400 V 3~ 50 Hz feed within ±5% V / ±2% Hz; units below 63 A come with cable but no plug. Process side needs G1 1/4 male pump fittings, cooling water G3/4 male, and at least 18 l fill volume with 48 l expansion volume.
Install indoors only, 5–40 °C ambient, IP20, at least 300 lx, up to 2000 m altitude, with 50–200 cm clearance per side. Not permitted in ATEX zones. Shut off cooling water for any downtime, including overnight, and secure casters/leveling feet before start-up.
Refrigerant-circuit work must be done by approved refrigeration contractors; the R-449A charge (CO2 equivalent 2.1 t) may require periodic leak checks. The three-phase plug is not included, and the customer must supply compatible cooling-water and process fittings for the G3/4 and G1 1/4 male connections.