Peter Huber Kältemaschinenbau Unichiller 350Tw-H
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 350Tw-H

Peter Huber Kältemaschinenbau Unichiller 350Tw-H — Recirculating chillers.

The Unichiller 350Tw-H is a water-cooled recirculating chiller with a 400 V 3~ 50 Hz supply, a D3 pump delivering up to 210 l/min at 4.7 bar, and a -20...100 °C working range. It distinguishes itself within the series by its high-capacity 350 size and water-cooled condenser, requiring a 1–6 bar cooling-water differential. It is intended for externally closed process loops where heat loads up to 25 kW at 0 °C must be rejected.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Pump typeSafety classTemperature range, °CPower supplyExpansion volume, m³Minimum fill volume, m³
Unichiller 350Tw-H Current D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 110T C3I / NFL-10–40400V 3~ 50Hz0.007—
Unichiller 110T-H C3III / FL-10–100400V 3~ 50Hz0.007—
Unichiller 110Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 130T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 130T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 130Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 160T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 160T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 160Tw C3I / NFL-20–40400V 3~ 50Hz0.01—
Unichiller 180-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 180T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 180T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 180Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 180Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 180w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 180w-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 200T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 200Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 210T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 210T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 210Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 230-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 230w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 230w-H E1III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller 250T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 250T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 250Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 260T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 260T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 260Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 260Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 300T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 300T-H D3III / FL-20–100460V 3~ 60Hz0.0390.014
Unichiller 300Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 300Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 350T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 500Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw-H D3III / FL-20–100400V 3~ 50Hz0.007—
Unichiller 700Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 800Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller GL 1000w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.05—
Unichiller GL 200w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 250w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 300w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 400w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 750w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.05—
Unichiller P180 E2I / NFL-20–40—0.0480.018
Unichiller P180-H E2III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller P180w E2I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller P180w-H E2III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller P230 E2I / NFL-20–40460V 3~ 60Hz0.0480.018
Unichiller P230-H E2III / FL-20–100460V 3~ 60Hz0.0480.018

Full specifications

Core Performance & Functional Specifications
Expansion volume 22 l
Heating capacity 2 kW
Minimum fill volume 15 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type D3
External sensor connection Pt100
Temperature range -20...100 °C
Dimensions, Weight & Mechanical Design
Weight 387 kg
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling capacity 15 kW (at 15 °C)
at 0 °C: 25 kW · at -10 °C: 16 kW · at -20 °C: 10 kW
Cooling water connection G3/4 male
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, control and operation

Integration & Operating Notes

Process Limits
Cooling-water inlet must stay between 5 and 20 °C

Operating outside this window invalidates the published cooling capacities and can cause low-temperature or high-temperature faults. If the inlet is below the ambient dew point, insulate all cooling-water pipes to prevent condensation on the circuit and on electrical components.

Unichiller 350Tw-H Technical Datasheet, page 2
Electrical
Supply deviations must not combine opposite signs

The unit accepts ±5% voltage and ±2% frequency, but a simultaneous -5% voltage with +2% frequency (or vice versa) is not permitted. Verify site power quality before connection; a 3×20 A fuse and a suitable plug (not supplied) are required on the 400 V 3~ 50 Hz feed.

Unichiller 350Tw-H Technical Datasheet, page 2
Mounting
Keep 50–200 cm clearance on either side for service

Free-standing installation requires this wall clearance to allow maintenance access to the refrigerant circuit and pump. The unit must be indoors, at 5–40 °C ambient, with at least 300 lx illuminance and at a site no higher than 2000 m altitude. Leveling feet must be extended and parking brakes applied before start-up.

Unichiller Operation Manual, page 25 and page 31
Process Limits
External circuit must be closed and match pump curve

The chiller is designed only for externally closed applications; an open loop will not work with the atmospheric expansion tank. Size the external circuit so its pressure drop at the required flow intersects the D3 pump curve (210 l/min at 4.7 bar, 140 l/min at 3.0 bar, 77 l/min at 4.0 bar). Minimum fill is 15 l.

Unichiller 350Tw-H Technical Datasheet, page 1; Unichiller Operation Manual, page 34
Control & I/O
Configure power-failure response for the process risk

The Pilot ONE can be set to OFF/Standby, ON, or Power Failure Auto after a mains interruption. For exothermic processes, Process Safety mode releases maximum cooling when the overtemperature trigger is reached instead of switching all poles off. Choose the mode that matches the process hazard before commissioning.

Unichiller Operation Manual, page 18 and page 19
Maintenance
Refrigerant work only by approved contractors

Any intervention on the refrigerant circuit (R-449A, 3 kg, GWP 1397) must be performed by approved refrigeration/air-conditioning system contractors. Check room volume against the refrigerant charge limit per m³ at installation. Shut off cooling water even for overnight downtime to prevent condensate and corrosion.

Unichiller Operation Manual, page 17 and page 28
Electrical
Protect cables below the unit from condensate

Water-cooled units can produce condensation; electrical lines routed directly below the chiller must be protected against liquid ingress. The unit is IP20 and Class A, Group 1 per EN55011, so maintain separation from sensitive equipment or add filtering if required.

Unichiller Operation Manual, page 28; Unichiller 350Tw-H Technical Datasheet, page 2

Selection & Integration Q&A

When should I choose the 350Tw-H over the smaller 110Tw-H or the air-cooled 110T-H?

Choose the 350Tw-H when the heat load at 0 °C approaches 25 kW and a water-cooled condenser is acceptable. The 110Tw-H is smaller and water-cooled; the 110T-H is air-cooled and needs no cooling-water circuit but has a higher noise level (74 dB(A) vs 65 dB(A) for the 110Tw).

What cooling-water supply and electrical infrastructure must be provided for this chiller?

Provide cooling water at 5–20 °C with 1 bar minimum differential pressure and 6 bar maximum, connected via G3/4 male fittings. Electrical supply is 400 V 3~ 50 Hz, max 19.5 A, 3×20 A fuses; a cable is supplied but no plug. Process connections are G1 1/4 AG male.

What are the installation environment and duty-cycle limits from the manual?

Indoors only, 5–40 °C ambient, at least 300 lx illuminance, max 2000 m altitude, and 50–200 cm clearance per side. The unit is not ATEX-rated and must not be installed in a hazardous zone. Leveling feet and parking brakes must be set before operation.

What maintenance and accessories are required beyond the base delivery?

The base delivery includes a cable but no plug; the customer must supply a suitable 3-phase plug and fusing. Refrigerant circuit service is restricted to approved contractors. No consumables or spare parts are listed in the provided documents; consult the manual's maintenance section for intervals.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller 350Tw-H Technical Datasheet
Datasheet ·18,3 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF