Peter Huber Kältemaschinenbau Unichiller 700Tw
Peter Huber Kältemaschinenbau Unichiller 700Tw — Recirculating chillers.
The Unichiller 700Tw is a water-cooled recirculating chiller in the Unichiller series, rated for a -20…40 °C working range and requiring 400 V 3~ 50 Hz supply at up to 45 A. Its G1 1/4 cooling-water connections and 3780 l/h consumption at 15 °C distinguish this larger modification from the 110T/110Tw siblings. It serves externally closed process loops where heat is rejected to a site cooling-water circuit.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Pump type | Safety class | Temperature range, °C | Power supply | Expansion volume, m³ | Minimum fill volume, m³ |
|---|---|---|---|---|---|---|
| Unichiller 700Tw Current | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 130T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 160T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.01 | — |
| Unichiller 180-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 180Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 180w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180w-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 200T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 200Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 210T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 230-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w-H | E1 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller 250T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 260T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 260Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 300T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 300T-H | D3 | III / FL | -20–100 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 300Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 300Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 350T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 350Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 500Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 800Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller GL 1000w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.05 | — |
| Unichiller GL 200w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 250w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 300w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 400w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 750w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.05 | — |
| Unichiller P180 | E2 | I / NFL | -20–40 | — | 0.048 | 0.018 |
| Unichiller P180-H | E2 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w | E2 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230 | E2 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
Full specifications
| Expansion volume | 7 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | D3 |
| External sensor connection | Pt100 |
| Temperature range | -20...40 °C |
| Weight | 600 kg |
| Current consumption | 45 A (max) |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1 1/4 male |
| Cooling water consumption | 1980 л/час (at 15°C, flow -10°C) at 15°C, flow 0°C: 2940 l/h · at 15°C, flow 15°C: 3780 l/h |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
- Water-cooled recirculating chiller with refrigerant R-449A (A1, H280; GWP 1397); wetted cooling-water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.
- Pump type D3 with fixed speed (pump-speed-adjustable: no); pump connection G1 1/4 AG male; cooling-water connection G1 1/4 male.
- 7 l expansion volume; 600 kg; overall dimensions 1000x1100x1635 mm.
- Within the series, the 700Tw is the larger water-cooled modification; the 110T/110Tw siblings are smaller units (the 110Tw listed at 65 dB(A)). Selection between them is driven by required cooling capacity and site water availability, not by the model number.
Behavior and process limits
- Temperature range -20…40 °C; ambient 5…40 °C.
- Cooling-water consumption depends on the flow temperature: 3780 l/h at 15 °C flow, 2940 l/h at 0 °C flow, 1980 l/h at -10 °C flow (all at 15 °C water).
- Rated output is valid only at 20 °C room temperature, 15 °C cooling-water inlet and 1.5 bar differential pressure; deviating site conditions reduce capacity.
- Cooling-water differential pressure minimum 1.5 bar; maximum cooling-water pressure 6 bar.
- Intended for externally closed applications only; not explosion-protected and must not be installed in an ATEX zone.
Control, connection and operation
- Pilot ONE controller; interfaces Ethernet, USB (Host and Device), RS232; external Pt100 sensor connection.
- Power-failure response configurable as OFF/Standby, ON or Power Failure Auto.
- Process Safety mode releases maximum cooling capacity when the overtemperature trigger value is reached, rather than switching off all poles.
- Three-phase devices below 63 A are supplied with cable but without a plug; voltage tolerance ±5% with simultaneous frequency tolerance ±2%, but combinations such as -5% voltage with +2% frequency are not permitted.
- Protection IP20; EMC classification Class A, Group 1 to EN55011 (may cause radio interference in residential areas).
Integration & Operating Notes
The unit is designed for cooling-water intake of 5…20 °C. If the water temperature falls below the condensation point, the cooling-water pipes inside the unit must be insulated to prevent condensation and electrical hazard.
The minimum cooling-water differential pressure is 1.5 bar and the maximum cooling-water pressure is 6 bar. The site supply must fall within these limits; rated capacity is quoted at 1.5 bar differential, so lower differential pressure degrades performance.
The cooling-water supply must be shut off during idle periods, including short downtimes such as overnight, to prevent water damage and corrosion.
Free-standing models require a wall clearance of 50 to 200 cm on either side for maintenance access. The 600 kg weight and 1000x1100x1635 mm footprint must be accommodated with adequate floor loading and doorway clearance.
The unit may only be used indoors with illuminance of at least 300 lx and at altitudes up to 2000 m. IP20 protection means it must be kept dry and protected from water ingress.
The unit is not explosion-protected and must not be installed or started up inside an ATEX zone. Refrigerant charge and room size must be checked at installation to avoid exceeding the refrigerant limit per m³ of ambient air.
Three-phase devices with current consumption less than 63 A are supplied with cable but without a plug; the installer must provide a suitable plug or hard-wired connection. The 700Tw draws up to 45 A at 400 V 3~ 50 Hz.
Voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%, but combinations such as -5% voltage with +2% frequency are not allowed. Out-of-tolerance supply may cause malfunction.
Electrical lines directly below a water-cooled unit must be protected against liquid ingress from condensation.
The Pilot ONE allows the power-failure response to be set to OFF/Standby, ON or Power Failure Auto. The choice determines whether the process restarts automatically after a power outage.
In Process Safety mode, maximum cooling capacity is released when the overtemperature protection trigger value is reached, instead of switching off all poles. This is relevant for exothermic processes where cutting cooling would cause runaway temperatures.
Work on the refrigerant circuit must be carried out only by approved refrigeration/air-conditioning system contractors. The refrigerant is R-449A (A1, H280; GWP 1397), so leak detection and ventilation requirements apply.
Before the unit is put into operation, parking brakes must be activated at the casters (if any) and/or leveling feet unscrewed/activated to prevent movement and crushing injuries.
Selection & Integration Q&A
Choose the 700Tw when the process requires the larger cooling capacity implied by its 3780 l/h cooling-water consumption at 15 °C flow and its -20…40 °C range. The 110Tw is a smaller unit (listed at 65 dB(A)); select it only when its capacity matches the load. The 700Tw is water-cooled and needs a site water circuit delivering ≥1.5 bar differential.
A 400 V 3~ 50 Hz supply capable of 45 A (cable supplied without plug below 63 A), a cooling-water circuit with G1 1/4 connections delivering 1.5–6 bar differential and up to 3780 l/h at 15 °C, and a dry indoor location with ≥300 lx and 50–200 cm side clearance. The process loop must be externally closed.
Install indoors only, not in an ATEX zone, at 5…40 °C ambient and below 2000 m altitude. Allow 50–200 cm wall clearance per side, verify floor loading for 600 kg, and activate leveling feet/parking brakes before start-up. Shut off cooling water even for overnight downtime.
The base delivery includes cable but no plug for three-phase units below 63 A, so a plug or hard-wired connection must be provided. Refrigerant circuit work requires approved refrigeration contractors. Cooling-water pipes may need insulation if water temperature falls below the condensation point. No consumables or spare parts are listed in the supplied documents.