Peter Huber Kältemaschinenbau Unichiller P180-H
Peter Huber Kältemaschinenbau Unichiller P180-H — Recirculating chillers.
The Unichiller P180-H is a recirculating chiller with a high-discharge-pressure E2 pump, built for externally closed loops with high pressure drop. Its 100 kW cooling capacity at 20 °C and -20…100 °C range distinguish it from lower-capacity siblings. It serves as a central cooling source for industrial processes requiring stable temperature control.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Pump type | Safety class | Temperature range, °C | Power supply | Expansion volume, m³ | Minimum fill volume, m³ |
|---|---|---|---|---|---|---|
| Unichiller P180-H Current | E2 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 110T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 130T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 160T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.01 | — |
| Unichiller 180-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 180Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 180w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180w-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 200T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 200Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 210T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 230-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w-H | E1 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller 250T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 260T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 260Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 300T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 300T-H | D3 | III / FL | -20–100 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 300Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 300Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 350T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 350Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 500Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 700Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 800Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller GL 1000w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.05 | — |
| Unichiller GL 200w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 250w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 300w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 400w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 750w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.05 | — |
| Unichiller P180 | E2 | I / NFL | -20–40 | — | 0.048 | 0.018 |
| Unichiller P180w | E2 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230 | E2 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
Full specifications
| Expansion volume | 48 l |
| Heating capacity | 2 kW |
| Minimum fill volume | 18 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | E2 |
| External sensor connection | Pt100 |
| Temperature range | -20...100 °C |
| Weight | 492 kg |
| Current consumption | 19 A |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling capacity | 100 kW (at 20 °C) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The P180-H belongs to the Unichiller series; the 'P' designation indicates a high-discharge-pressure circulation pump (type E2) for applications with high pressure drops, as stated in the manual and datasheet.
- Unlike sibling models such as the Unichiller 110T (74 dB(A)) or 110Tw (65 dB(A)), the P180-H is rated at 100 kW cooling capacity at 20 °C, placing it at the higher end of the series' capacity range.
- The unit is designed exclusively for externally closed applications; open baths or unsealed loops are not permitted.
- It features an atmospheric open expansion tank with 48 l expansion volume and a minimum fill volume of 18 l, plus a digital level indicator.
Behavior under load and media
- Cooling capacity is specified at 20 °C room temperature; higher ambient temperatures reduce available cooling, so derating is necessary in warm plant rooms.
- The temperature range spans -20 to 100 °C, with a 2 kW heating capacity for precise temperature control near setpoint.
- The refrigerant is R-449A (safety class A1, GHS hazard H280), requiring ventilation and leak precautions.
- Overtemperature protection is adjustable per DIN 12876; in Process Safety mode, maximum cooling is released when the OT trigger value is reached, rather than shutting down all poles.
Installation, connection and control
- Electrical supply is 400 V 3~ 50 Hz with maximum current 19 A and 3x20 A fusing; the unit is delivered with cable but without a plug, so site termination is required.
- Voltage and frequency tolerances must be met simultaneously; a combination of -5% voltage and +2% frequency is not allowed.
- The Pilot ONE controller provides Ethernet, USB (host and device) and RS232 interfaces, and supports configurable power-failure response (OFF/Standby, ON, or Power Failure Auto).
- Pump connections are G1 1/4 AG male; maximum tightening torques must be observed and connections counterheld during hose installation.
- Installation requires 50–200 cm wall clearance per side for maintenance, indoor use only with illuminance ≥300 lx and altitude up to 2000 m, and ambient temperature between 5 and 40 °C.
- The unit is IP20 and EMC Class A Group 1 (industrial environments); it must not be installed in ATEX zones.
Integration & Operating Notes
Free-standing models require a wall clearance of 50 to 200 cm on either side to allow service access and adequate airflow. Install indoors only, at an illuminance of at least 300 lx and at altitudes up to 2000 m. Ambient temperature must remain between 5 and 40 °C.
The chiller must not be installed or started up inside an ATEX zone. Additionally, the refrigerant R-449A (A1, H280) requires that the charge and room size be checked to avoid exceeding the refrigerant limit per m³ of ambient air. Ensure adequate ventilation and take leak precautions.
In Process Safety mode, the controller releases maximum cooling capacity when the overtemperature protection trigger value is reached, instead of switching off all poles. This is essential for exothermic processes where cutting cooling could cause a temperature runaway. Set the OT trigger value according to DIN 12876 for the application.
For water-cooled units, the cooling water supply must be shut off during downtimes, including overnight, to prevent water damage and corrosion. Also protect electrical lines directly below the unit against liquid ingress from condensation.
Before switching on, unscrew or activate the leveling feet and apply parking brakes at the casters (if any) to prevent the unit from moving. This avoids crushing injuries and ensures stable operation.
Selection & Integration Q&A
Choose the P180-H when your process loop has high pressure drop and requires up to 100 kW cooling at 20 °C. The 'P' pump delivers high discharge pressure. Siblings like the 110T offer lower capacity and different noise levels; select them for lower loads or where noise is critical.
The external circuit must be closed and pressure-tight. Pump connections are G1 1/4 AG male; observe maximum tightening torques and counterhold fittings. Electrical supply is 400 V 3~ 50 Hz, 19 A max, 3x20 A fusing; the unit comes with cable but no plug, so site termination is needed.
Install indoors only, with 50–200 cm wall clearance per side, ambient temperature 5–40 °C, illuminance ≥300 lx, and altitude ≤2000 m. Do not install in ATEX zones. Ensure room volume is sufficient for the R-449A charge. Activate leveling feet and parking brakes before operation.
Refrigerant circuit work must be by approved contractors. Shut off cooling water during downtimes. The base delivery includes cable but no plug; mating connectors, hoses, and fittings for the G1 1/4 connections are not included. No specific maintenance interval is stated in the provided documents.