Peter Huber Kältemaschinenbau Unichiller 180w
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 180w

Peter Huber Kältemaschinenbau Unichiller 180w — Recirculating chillers.

Water-cooled recirculating chiller with a non-variable E1 pressure pump, rated for -20 to 40 °C in externally closed loops. It is the 180w modification of the Unichiller series, distinguished by its 54 l/min pump and 48 l expansion volume. It serves as a central cooling source for closed process circuits requiring a 400 V three-phase supply and G3/4 cooling water.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Pump typeSafety classTemperature range, °CPower supplyExpansion volume, m³Minimum fill volume, m³
Unichiller 180w Current E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 110T C3I / NFL-10–40400V 3~ 50Hz0.007—
Unichiller 110T-H C3III / FL-10–100400V 3~ 50Hz0.007—
Unichiller 110Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 130T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 130T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 130Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 160T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 160T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 160Tw C3I / NFL-20–40400V 3~ 50Hz0.01—
Unichiller 180-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 180T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 180T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 180Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 180Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 180w-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 200T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 200Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 210T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 210T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 210Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 230-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 230w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 230w-H E1III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller 250T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 250T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 250Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 260T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 260T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 260Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 260Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 300T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 300T-H D3III / FL-20–100460V 3~ 60Hz0.0390.014
Unichiller 300Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 300Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 350T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 350Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 500Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw-H D3III / FL-20–100400V 3~ 50Hz0.007—
Unichiller 700Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 800Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller GL 1000w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.05—
Unichiller GL 200w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 250w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 300w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 400w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 750w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.05—
Unichiller P180 E2I / NFL-20–40—0.0480.018
Unichiller P180-H E2III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller P180w E2I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller P180w-H E2III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller P230 E2I / NFL-20–40460V 3~ 60Hz0.0480.018
Unichiller P230-H E2III / FL-20–100460V 3~ 60Hz0.0480.018

Full specifications

Core Performance & Functional Specifications
Expansion volume 48 l
Maximum delivery flow rate 54 l/min (pressure pump max; also 3, 5 bar)
Minimum fill volume 18 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type E1
External sensor connection Pt100
Temperature range -20...40 °C
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

Hydraulic and thermal behavior

Installation, connection and control

Integration & Operating Notes

Process Limits
Cooling water inlet must stay between 5 °C and 20 °C

The chiller is designed to operate with a cooling water intake temperature between 5 °C and 20 °C. Higher inlet temperatures reduce cooling capacity and may require an increased cooling water flow rate, so site water conditions must be checked before sizing the unit for the process load.

Unichiller 180w Technical Data, page 2
Electrical
Three-phase cable supplied without a plug

Three-phase devices with current consumption below 63 A are delivered with a cable but without a plug. The customer must provide and fit the site plug, and the electrical installation must be sized and protected for a 400 V 3~ 50 Hz supply.

Unichiller 180w Technical Data, page 2
Electrical
Voltage and frequency tolerances are coupled

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed. The site power quality must be checked against this combined tolerance before connection.

Unichiller 180w Technical Data, page 2
Mounting
Keep 50–200 cm wall clearance per side

A wall clearance of 50 to 200 cm per side is required for maintenance on free-standing models. The unit may only be used indoors at an illuminance of at least 300 lx and up to 2000 m altitude, and must not be installed or started up inside an ATEX zone.

Unichiller Operation Manual, page 25
Mounting
Protect cables below the unit from condensate

Electrical lines directly below a water-cooled unit must be protected against liquid ingress from condensation. The unit has IP20 protection and must be installed in a dry indoor location protected from water ingress.

Unichiller Operation Manual, page 28
Process Limits
External circuit must be closed and sealed

The unit is intended only for externally closed applications. Open baths or unsealed loops are not permitted, which determines the permissible hydraulic circuit configuration and the minimum filling capacity of 18 l.

Unichiller Operation Manual, page 34
Maintenance
Refrigerant work requires certified contractors

Work on the refrigerant circuit must only be performed by approved refrigeration/air-conditioning system contractors. The refrigerant is R-449A with safety group A1 and GHS hazard H280, and the charge of 1.49 kg corresponds to 2.1 t CO2 equivalent, which affects leak detection and F-gas reporting obligations.

Unichiller Operation Manual, page 17
Operation
Configure power-failure response before commissioning

The power-failure response can be configured as OFF/Standby, ON, or Power Failure Auto via the Pilot ONE. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger value is reached instead of switching off all poles.

Unichiller Operation Manual, pages 18–19
Operation
Shut off cooling water even for overnight stops

The cooling water supply must be shut off even for shorter downtimes such as overnight to prevent water damage and corrosion during idle periods. Leveling feet must be activated and parking brakes applied before switching on the unit.

Unichiller Operation Manual, pages 28 and 31

Selection & Integration Q&A

When should I choose the Unichiller 180w instead of a sibling model such as the 110T or 110Tw?

Choose the 180w when you need its specific pump and hydraulic configuration (E1 pressure pump, 54 l/min, non-adjustable speed) and 48 l expansion volume. Sibling models differ in noise level (74 dB(A) for 110T, 65 dB(A) for 110Tw) and are not substitutes for this model's pump and circuit design.

What cooling water supply and piping must be provided for this chiller?

Provide a cooling water supply and return with G3/4 male connections, with inlet temperature between 5 °C and 20 °C. The circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used. Insulate pipes where condensation can occur.

What are the installation environment and duty-cycle limits for the Unichiller 180w?

Install indoors only, at 5–40 °C ambient, at least 300 lx illuminance, up to 2000 m altitude, with 50–200 cm wall clearance per side. IP20 requires a dry location. Do not install in an ATEX zone. Shut off cooling water even for overnight stops.

What is not included in the base delivery and what maintenance access is required?

Three-phase devices below 63 A are delivered with a cable but without a plug, so the site plug must be provided. Refrigerant circuit work must be performed by approved refrigeration/air-conditioning contractors. A wall clearance of 50 to 200 cm per side is required for maintenance on free-standing models.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller 180w Technical Data
Datasheet ·18,1 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF