Peter Huber Kältemaschinenbau Unichiller 130T-H
Peter Huber Kältemaschinenbau Unichiller 130T-H — Recirculating chillers with Noise level 74 dB.
The Unichiller 130T-H is an air-cooled, externally closed recirculating chiller with a fixed-speed C3 pump and Pilot ONE controller, rated for -10 to 100 °C. It distinguishes itself within the series by its 2 kW heating capacity and 39 l expansion volume, serving as a temperature control unit for closed process loops requiring moderate heating and cooling.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Pump type | Safety class | Temperature range, °C | Power supply | Expansion volume, m³ | Minimum fill volume, m³ |
|---|---|---|---|---|---|---|
| Unichiller 130T-H Current | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 110T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 130T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 160T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.01 | — |
| Unichiller 180-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 180Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 180w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180w-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 200T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 200Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 210T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 230-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w-H | E1 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller 250T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 260T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 260Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 300T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 300T-H | D3 | III / FL | -20–100 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 300Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 300Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 350T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 350Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 500Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 700Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 800Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller GL 1000w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.05 | — |
| Unichiller GL 200w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 250w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 300w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 400w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 750w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.05 | — |
| Unichiller P180 | E2 | I / NFL | -20–40 | — | 0.048 | 0.018 |
| Unichiller P180-H | E2 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w | E2 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230 | E2 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
Full specifications
| Noise level | 74 dB(A) |
| Expansion volume | 39 l |
| Heating capacity | 2 kW |
| Minimum fill volume | 14 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | C3 |
| External sensor connection | Pt100 |
| Temperature range | -10...100 °C |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and Series Positioning
The Unichiller 130T-H is a free-standing, air-cooled recirculating chiller intended exclusively for externally closed applications. Within the Unichiller series, it is positioned as a heating-capable variant (suffix -H) with a 2 kW heater, distinguishing it from non-heating siblings such as the Unichiller 110T. The unit uses a fixed-speed C3 pressure pump with G1 1/4 male connections and is controlled by the Pilot ONE controller. It is not explosion-protected and must not be installed in ATEX zones.
Operating Behavior and Media
- Temperature range: -10 to 100 °C, with a minimum fill volume of 14 l and an expansion volume of 39 l.
- Cooling capacity is rated at 20 °C room temperature and may decrease at higher ambient temperatures; ambient temperature must remain between 5 and 40 °C.
- The refrigerant is R-449A (A1, H280), requiring ventilation and handling precautions.
- The unit is designed for closed-loop operation; open baths or unsealed circuits are not permitted.
Installation, Control and Operation
- Electrical supply: 400 V 3~ 50 Hz, with voltage tolerance ±5% and frequency tolerance ±2% (not all combinations allowed). Three-phase devices below 63 A are delivered with cable but without a plug.
- Protection rating IP20 requires a dry indoor location; electrical lines below a water-cooled unit (if applicable) must be protected against condensate.
- The Pilot ONE controller offers Ethernet, USB (host and device) and RS232 interfaces, and allows configuration of power-failure response (OFF/Standby, ON, or Power Failure Auto).
- Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached, rather than switching off all poles.
- Installation requires wall clearance of 50–200 cm per side for maintenance, leveling feet or parking brakes engaged before start-up, and observance of refrigerant charge versus room volume.
Integration & Operating Notes
Free-standing models require a wall clearance of 50 to 200 cm on either side to allow maintenance and ensure adequate airflow for the air-cooled condenser. Insufficient clearance can reduce cooling performance and hinder service access.
The unit requires a 400 V 3~ 50 Hz supply. Devices with current consumption below 63 A are delivered with a cable but without a plug, so the customer must provide and fit a suitable plug or hard-wired connection according to local regulations.
The chiller is intended only for externally closed applications. The pump connection is G1 1/4 male, so hoses or adapters must match this thread. Maximum tightening torques for pump connections must not be exceeded and connections must be counterheld during installation.
Cooling capacity is rated at 20 °C room temperature and may decrease at higher ambient temperatures. The ambient temperature must stay between 5 °C and 40 °C for valid operation; exceeding this range risks reduced performance or unit protection trips.
In Process Safety mode, the controller releases maximum cooling capacity when the overtemperature trigger value is reached, instead of switching off all poles. This is critical for exothermic processes where loss of cooling could cause runaway temperatures.
Any work on the refrigerant circuit must be performed only by approved refrigeration/air-conditioning system contractors. The refrigerant is R-449A (A1, H280), and refrigerant charge versus room volume must be checked at installation to avoid exceeding the limit per m³ of ambient air.
Selection & Integration Q&A
Choose the 130T-H when you need active heating (2 kW) and a larger expansion volume (39 l) for a closed loop. The 110T lacks heating; the 110Tw offers lower noise (65 dB(A)) but may have different capacity. Select based on heating demand and acoustic constraints.
The pump connection is G1 1/4 male, so hoses or adapters must match this thread. Use counterheld connections and do not exceed maximum tightening torques. The external circuit must be closed; open baths are not permitted. Minimum fill volume is 14 l.
Install indoors only, at 5–40 °C ambient, with illuminance ≥300 lx and altitude ≤2000 m. Maintain 50–200 cm side clearance for maintenance. IP20 requires a dry location. Cooling capacity is rated at 20 °C and derates at higher ambient.
The unit is delivered with a cable but without a plug for three-phase supply; the customer must provide a suitable plug. Refrigerant circuit work requires certified contractors. Regular maintenance includes checking refrigerant charge and room volume, and shutting off cooling water during downtimes (if water-cooled).