Peter Huber Kältemaschinenbau Unichiller 130T-H
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 130T-H

Peter Huber Kältemaschinenbau Unichiller 130T-H — Recirculating chillers with Noise level 74 dB.

The Unichiller 130T-H is an air-cooled, externally closed recirculating chiller with a fixed-speed C3 pump and Pilot ONE controller, rated for -10 to 100 °C. It distinguishes itself within the series by its 2 kW heating capacity and 39 l expansion volume, serving as a temperature control unit for closed process loops requiring moderate heating and cooling.

Noise level
74 dB

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Pump typeSafety classTemperature range, °CPower supplyExpansion volume, m³Minimum fill volume, m³
Unichiller 130T-H Current C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 110T C3I / NFL-10–40400V 3~ 50Hz0.007—
Unichiller 110T-H C3III / FL-10–100400V 3~ 50Hz0.007—
Unichiller 110Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 130T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 130Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 160T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 160T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 160Tw C3I / NFL-20–40400V 3~ 50Hz0.01—
Unichiller 180-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 180T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 180T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 180Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 180Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 180w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 180w-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 200T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 200Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 210T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 210T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 210Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 230-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 230w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 230w-H E1III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller 250T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 250T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 250Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 260T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 260T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 260Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 260Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 300T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 300T-H D3III / FL-20–100460V 3~ 60Hz0.0390.014
Unichiller 300Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 300Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 350T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 350Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 500Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw-H D3III / FL-20–100400V 3~ 50Hz0.007—
Unichiller 700Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 800Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller GL 1000w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.05—
Unichiller GL 200w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 250w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 300w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 400w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 750w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.05—
Unichiller P180 E2I / NFL-20–40—0.0480.018
Unichiller P180-H E2III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller P180w E2I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller P180w-H E2III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller P230 E2I / NFL-20–40460V 3~ 60Hz0.0480.018
Unichiller P230-H E2III / FL-20–100460V 3~ 60Hz0.0480.018

Full specifications

Core Performance & Functional Specifications
Noise level 74 dB(A)
Expansion volume 39 l
Heating capacity 2 kW
Minimum fill volume 14 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type C3
External sensor connection Pt100
Temperature range -10...100 °C
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and Series Positioning

The Unichiller 130T-H is a free-standing, air-cooled recirculating chiller intended exclusively for externally closed applications. Within the Unichiller series, it is positioned as a heating-capable variant (suffix -H) with a 2 kW heater, distinguishing it from non-heating siblings such as the Unichiller 110T. The unit uses a fixed-speed C3 pressure pump with G1 1/4 male connections and is controlled by the Pilot ONE controller. It is not explosion-protected and must not be installed in ATEX zones.

Operating Behavior and Media

Installation, Control and Operation

Integration & Operating Notes

Mounting
Maintain 50–200 cm side clearance for service access

Free-standing models require a wall clearance of 50 to 200 cm on either side to allow maintenance and ensure adequate airflow for the air-cooled condenser. Insufficient clearance can reduce cooling performance and hinder service access.

Unichiller Operation Manual, page 25
Electrical
Three-phase supply with cable but no plug provided

The unit requires a 400 V 3~ 50 Hz supply. Devices with current consumption below 63 A are delivered with a cable but without a plug, so the customer must provide and fit a suitable plug or hard-wired connection according to local regulations.

Unichiller 130T-H Technical Data Sheet, page 2
Process Limits
External circuit must be closed and match G1 1/4 male threads

The chiller is intended only for externally closed applications. The pump connection is G1 1/4 male, so hoses or adapters must match this thread. Maximum tightening torques for pump connections must not be exceeded and connections must be counterheld during installation.

Unichiller Operation Manual, page 28 and 34
Cooling & Media
Cooling capacity derates above 20 °C ambient

Cooling capacity is rated at 20 °C room temperature and may decrease at higher ambient temperatures. The ambient temperature must stay between 5 °C and 40 °C for valid operation; exceeding this range risks reduced performance or unit protection trips.

Unichiller 130T-H Technical Data Sheet, page 2
Safety
Process Safety mode releases max cooling on OT trip

In Process Safety mode, the controller releases maximum cooling capacity when the overtemperature trigger value is reached, instead of switching off all poles. This is critical for exothermic processes where loss of cooling could cause runaway temperatures.

Unichiller Operation Manual, page 18
Maintenance
Refrigerant circuit work requires certified contractor

Any work on the refrigerant circuit must be performed only by approved refrigeration/air-conditioning system contractors. The refrigerant is R-449A (A1, H280), and refrigerant charge versus room volume must be checked at installation to avoid exceeding the limit per m³ of ambient air.

Unichiller Operation Manual, page 15 and 17

Selection & Integration Q&A

When should I choose the Unichiller 130T-H over the 110T or 110Tw variants?

Choose the 130T-H when you need active heating (2 kW) and a larger expansion volume (39 l) for a closed loop. The 110T lacks heating; the 110Tw offers lower noise (65 dB(A)) but may have different capacity. Select based on heating demand and acoustic constraints.

What are the requirements for the external piping and pump connections?

The pump connection is G1 1/4 male, so hoses or adapters must match this thread. Use counterheld connections and do not exceed maximum tightening torques. The external circuit must be closed; open baths are not permitted. Minimum fill volume is 14 l.

What are the ambient and installation limits for this chiller?

Install indoors only, at 5–40 °C ambient, with illuminance ≥300 lx and altitude ≤2000 m. Maintain 50–200 cm side clearance for maintenance. IP20 requires a dry location. Cooling capacity is rated at 20 °C and derates at higher ambient.

What maintenance and accessories are required that are not included in base delivery?

The unit is delivered with a cable but without a plug for three-phase supply; the customer must provide a suitable plug. Refrigerant circuit work requires certified contractors. Regular maintenance includes checking refrigerant charge and room volume, and shutting off cooling water during downtimes (if water-cooled).