Peter Huber Kältemaschinenbau Unichiller GL 750w
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller GL 750w

Peter Huber Kältemaschinenbau Unichiller GL 750w — Recirculating chillers.

The Unichiller GL 750w is a water-cooled recirculating chiller delivering 75 kW at 20 °C, operating from -20 to 40 °C with a fixed-speed pressure pump. It is the large-capacity GL variant in the Unichiller series, distinguished by its 75 kW rating and G1 1/4 process connections. It serves as a central cooling source for externally closed industrial process loops.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Pump typeSafety classTemperature range, °CPower supplyExpansion volume, m³Minimum fill volume, m³
Unichiller GL 750w Current Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.05—
Unichiller 110T C3I / NFL-10–40400V 3~ 50Hz0.007—
Unichiller 110T-H C3III / FL-10–100400V 3~ 50Hz0.007—
Unichiller 110Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 130T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 130T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 130Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 160T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 160T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 160Tw C3I / NFL-20–40400V 3~ 50Hz0.01—
Unichiller 180-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 180T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 180T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 180Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 180Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 180w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 180w-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 200T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 200Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 210T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 210T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 210Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 230-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 230w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 230w-H E1III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller 250T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 250T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 250Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 260T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 260T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 260Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 260Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 300T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 300T-H D3III / FL-20–100460V 3~ 60Hz0.0390.014
Unichiller 300Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 300Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 350T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 350Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 500Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw-H D3III / FL-20–100400V 3~ 50Hz0.007—
Unichiller 700Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 800Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller GL 1000w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.05—
Unichiller GL 200w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 250w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 300w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 400w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller P180 E2I / NFL-20–40—0.0480.018
Unichiller P180-H E2III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller P180w E2I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller P180w-H E2III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller P230 E2I / NFL-20–40460V 3~ 60Hz0.0480.018
Unichiller P230-H E2III / FL-20–100460V 3~ 60Hz0.0480.018

Full specifications

Core Performance & Functional Specifications
Expansion volume 50 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type Pressure pump, speed not variable
External sensor connection Pt100
Temperature range -20...40 °C
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling capacity 75 kW (at 20 °C)
Cooling water connection G1 1/4 male
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

The GL 750w is a water-cooled (not air-cooled) chiller with a 75 kW cooling capacity at 20 °C and a -20…40 °C temperature range. Its process connections are G1 1/4 male on both pump and cooling-water sides, and it carries a 50 l expansion volume with a 17.5 l minimum filling capacity. Within the Unichiller series, the sibling models listed (110T, 110T-H, 110Tw, 110Tw-H) are air-cooled variants with published noise levels; the GL 750w is water-cooled and therefore has no fan-noise figure. The fixed-speed pressure pump delivers 320 l/min maximum and 110 l/min at 4.0 bar, so the external circuit must be matched to this pump curve rather than to a variable-speed pump.

Behavior under load and media

Rated output is valid only at 20 °C room temperature, 15 °C cooling-water inlet and 1 bar differential pressure; at other site conditions the capacity will differ. The cooling-water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used. Cooling-water inlet must be between 5 °C and 20 °C; below the condensation point the internal cooling-water pipes must be insulated. The refrigerant is R-744 (CO2), GWP 1, safety class A1/H280.

Installation, control and operation

The Pilot ONE controller provides Ethernet, USB (host and device) and RS232 interfaces, plus a Pt100 external sensor connection. Overtemperature protection per DIN 12876 is adjustable, and the Process Safety mode releases maximum cooling capacity when the OT trigger is reached instead of switching off. Power-failure response is configurable as OFF/Standby, ON or Power Failure Auto. The unit is IP20, EMC Class A Group 1, and must be installed indoors with at least 300 lx illuminance and up to 2000 m altitude. Three-phase devices below 63 A are delivered with a cable but without a plug.

Integration & Operating Notes

Process Limits
Cooling water must be 5–20 °C and 1–6 bar differential

Insufficient differential pressure prevents rated cooling and may trip the unit; exceeding 6 bar risks damage. Below the condensation point, internal cooling-water pipes must be insulated to prevent condensation.

Datasheet, Specifications
Mounting
50–200 cm wall clearance per side and leveling feet before start

Free-standing models require 50 to 200 cm clearance on either side for maintenance. Parking brakes must be activated and leveling feet unscrewed before operation. Electrical lines directly below a water-cooled unit must be protected against liquid ingress.

Operation Manual, Installation
Process Limits
Externally closed applications only; no ATEX zone installation

The unit is designed for externally closed circuits and is not explosion-protected, so it must not be installed or started inside an ATEX zone. Refrigerant charge and room size must be checked against the limit per m³ of ambient air.

Operation Manual, Function Description

Selection & Integration Q&A

When should I choose the GL 750w over the Unichiller 110T or 110Tw siblings?

Choose the GL 750w for 75 kW water-cooled duty with G1 1/4 connections and no fan noise; the 110T/110Tw are air-cooled with published noise levels (74/65 dB(A)) and suit sites without cooling water.

What pump and external circuit must be matched to the GL 750w?

The fixed-speed pressure pump delivers 320 l/min max and 110 l/min at 4.0 bar; size the external loop so its pressure drop at required flow stays within this curve, and respect the 17.5 l minimum filling and 50 l expansion volume.

What installation environment and duty limits apply from the manual?

Indoors only, 5–40 °C ambient, at least 300 lx, up to 2000 m altitude, 50–200 cm side clearance, leveling feet and parking brakes set before start; not permitted in ATEX zones.

What is not included in base delivery and what service is restricted?

Three-phase units below 63 A ship with cable but no plug; the customer provides the plug and connection. Refrigerant-circuit work is restricted to approved refrigeration contractors; cooling water must be shut off even overnight.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller GL 750w
Datasheet ·30,6 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF