Peter Huber Kältemaschinenbau Unichiller 180w
Peter Huber Kältemaschinenbau Unichiller 180w — Recirculating chillers.
Water-cooled recirculating chiller with a non-variable E1 pressure pump, rated for -20 to 40 °C in externally closed loops. It is the 180w modification of the Unichiller series, distinguished by its 54 l/min pump and 48 l expansion volume. It serves as a central cooling source for closed process circuits requiring a 400 V three-phase supply and G3/4 cooling water.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Pump type | Safety class | Temperature range, °C | Power supply | Expansion volume, m³ | Minimum fill volume, m³ |
|---|---|---|---|---|---|---|
| Unichiller 180w Current | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 110T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 130T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 160T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.01 | — |
| Unichiller 180-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 180Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 180w-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 200T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 200Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 210T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 230-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w-H | E1 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller 250T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 260T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 260Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 300T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 300T-H | D3 | III / FL | -20–100 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 300Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 300Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 350T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 350Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 500Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 700Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 800Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller GL 1000w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.05 | — |
| Unichiller GL 200w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 250w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 300w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 400w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 750w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.05 | — |
| Unichiller P180 | E2 | I / NFL | -20–40 | — | 0.048 | 0.018 |
| Unichiller P180-H | E2 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w | E2 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230 | E2 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
Full specifications
| Expansion volume | 48 l |
| Maximum delivery flow rate | 54 l/min (pressure pump max; also 3, 5 bar) |
| Minimum fill volume | 18 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | E1 |
| External sensor connection | Pt100 |
| Temperature range | -20...40 °C |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
- The Unichiller 180w is a water-cooled recirculating chiller built for externally closed applications only; open baths or unsealed loops are not permitted.
- Within the series, this modification uses the E1 pressure pump with a maximum delivery flow rate of 54 l/min and a non-adjustable speed. Sibling models such as the Unichiller 110T and 110Tw differ in noise level (74 dB(A) and 65 dB(A) respectively) and are not interchangeable with this model's pump and hydraulic configuration.
- The refrigerant circuit contains R-449A (safety group A1, GHS hazard H280) with a charge of 1.49 kg, corresponding to 2.1 t CO2 equivalent.
Hydraulic and thermal behavior
- The working temperature range is -20 to 40 °C. The minimum filling capacity is 18 l and the expansion volume is 48 l, which set the minimum system volume and expansion allowance for the external loop.
- Cooling capacity depends on the cooling water inlet temperature, which must be kept between 5 °C and 20 °C. As the cooling water temperature rises, a drop in cooling power and a possible increase in required cooling water flow rate must be expected.
- The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM; only suitable cooling water may be used to avoid corrosion and fouling.
Installation, connection and control
- The unit requires a 400 V 3~ 50 Hz supply. Three-phase devices below 63 A are delivered with a cable but without a plug, so the site plug must be provided and fitted.
- Cooling water connections are G3/4 male; pump connections are G1 1/4 AG male. Maximum tightening torques for pump connections must not be exceeded and connections must be counterheld during hose installation.
- The Pilot ONE controller provides Ethernet, USB (Host and Device) and RS232 interfaces, and accepts an external Pt100 sensor. The power-failure response can be configured as OFF/Standby, ON, or Power Failure Auto.
- Electrical lines directly below the water-cooled unit must be protected against liquid ingress from condensation. The cooling water supply must be shut off even for short downtimes such as overnight.
- The unit has IP20 protection and must be installed in a dry indoor location at an illuminance of at least 300 lx and up to 2000 m altitude. A wall clearance of 50 to 200 cm per side is required for maintenance on free-standing models. Leveling feet must be activated and parking brakes applied before switching on.
Integration & Operating Notes
The chiller is designed to operate with a cooling water intake temperature between 5 °C and 20 °C. Higher inlet temperatures reduce cooling capacity and may require an increased cooling water flow rate, so site water conditions must be checked before sizing the unit for the process load.
Three-phase devices with current consumption below 63 A are delivered with a cable but without a plug. The customer must provide and fit the site plug, and the electrical installation must be sized and protected for a 400 V 3~ 50 Hz supply.
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed. The site power quality must be checked against this combined tolerance before connection.
A wall clearance of 50 to 200 cm per side is required for maintenance on free-standing models. The unit may only be used indoors at an illuminance of at least 300 lx and up to 2000 m altitude, and must not be installed or started up inside an ATEX zone.
Electrical lines directly below a water-cooled unit must be protected against liquid ingress from condensation. The unit has IP20 protection and must be installed in a dry indoor location protected from water ingress.
The unit is intended only for externally closed applications. Open baths or unsealed loops are not permitted, which determines the permissible hydraulic circuit configuration and the minimum filling capacity of 18 l.
Work on the refrigerant circuit must only be performed by approved refrigeration/air-conditioning system contractors. The refrigerant is R-449A with safety group A1 and GHS hazard H280, and the charge of 1.49 kg corresponds to 2.1 t CO2 equivalent, which affects leak detection and F-gas reporting obligations.
The power-failure response can be configured as OFF/Standby, ON, or Power Failure Auto via the Pilot ONE. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger value is reached instead of switching off all poles.
The cooling water supply must be shut off even for shorter downtimes such as overnight to prevent water damage and corrosion during idle periods. Leveling feet must be activated and parking brakes applied before switching on the unit.
Selection & Integration Q&A
Choose the 180w when you need its specific pump and hydraulic configuration (E1 pressure pump, 54 l/min, non-adjustable speed) and 48 l expansion volume. Sibling models differ in noise level (74 dB(A) for 110T, 65 dB(A) for 110Tw) and are not substitutes for this model's pump and circuit design.
Provide a cooling water supply and return with G3/4 male connections, with inlet temperature between 5 °C and 20 °C. The circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used. Insulate pipes where condensation can occur.
Install indoors only, at 5–40 °C ambient, at least 300 lx illuminance, up to 2000 m altitude, with 50–200 cm wall clearance per side. IP20 requires a dry location. Do not install in an ATEX zone. Shut off cooling water even for overnight stops.
Three-phase devices below 63 A are delivered with a cable but without a plug, so the site plug must be provided. Refrigerant circuit work must be performed by approved refrigeration/air-conditioning contractors. A wall clearance of 50 to 200 cm per side is required for maintenance on free-standing models.