Peter Huber Kältemaschinenbau Unichiller 230-H
Peter Huber Kältemaschinenbau Unichiller 230-H — Recirculating chillers.
The Unichiller 230-H is an air-cooled, externally closed recirculating chiller with a 54 l/min E1 pump, G1 1/4 male connections and a -20…100 °C working range. Within the Unichiller series it is the 230-size, high-capacity modification, distinguished by 23 kW cooling at 20 °C falling to 5.5 kW at -20 °C. It serves closed process loops requiring stable temperature control with Pilot ONE and adjustable DIN 12876 overtemperature protection.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Pump type | Safety class | Temperature range, °C | Power supply | Expansion volume, m³ | Minimum fill volume, m³ |
|---|---|---|---|---|---|---|
| Unichiller 230-H Current | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 110T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 130T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 160T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.01 | — |
| Unichiller 180-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 180Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 180w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180w-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 200T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 200Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 210T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 230w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w-H | E1 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller 250T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 260T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 260Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 300T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 300T-H | D3 | III / FL | -20–100 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 300Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 300Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 350T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 350Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 500Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 700Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 800Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller GL 1000w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.05 | — |
| Unichiller GL 200w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 250w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 300w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 400w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 750w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.05 | — |
| Unichiller P180 | E2 | I / NFL | -20–40 | — | 0.048 | 0.018 |
| Unichiller P180-H | E2 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w | E2 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230 | E2 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
Full specifications
| Expansion volume | 48 l |
| Heating capacity | 2 kW |
| Minimum fill volume | 18 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | E1 |
| External sensor connection | Pt100 |
| Temperature range | -20...100 °C |
| Weight | 477 kg |
| Current consumption | 14 A |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Air-cooled refrigeration circuit using R-449A (A1, H280), 1.49 kg charge, 2.1 t CO2 equivalent; 2 kW electric heating integrated for bidirectional temperature control.
- Safety class III / FL with adjustable overtemperature protection to DIN 12876; Pilot ONE controller with Ethernet, USB host/device and RS232.
- Hydraulic side: E1 circulation pump, non-adjustable speed, G1 1/4 AG male connections, 18 l minimum fill and 48 l expansion volume.
- Compared with the Unichiller 110T/110Tw siblings, this 230-H is the larger-capacity unit; the 110Tw variants are the quieter (65 dB(A)) alternatives where acoustic limits dominate.
Behavior under load and media
- Cooling capacity is strongly temperature-dependent: 23 kW at 20 °C but only 5.5 kW at -20 °C, so the load must be checked at the lowest process temperature, not at the 20 °C rating point.
- Pump curve is steep: 54 l/min at zero head, 11 l/min at 3.0 bar; external circuit pressure drop must be matched to this curve to preserve flow.
- Published output data are valid at 20 °C room temperature; capacity drops as ambient rises toward the 40 °C limit.
- Intended exclusively for externally closed applications; open baths or unsealed loops are not a valid configuration.
Installation, control and operation
- Requires 400 V 3~ 50 Hz, 14 A, 16 A fuse per phase; three-phase units below 63 A are delivered with cable but without plug, so the site plug must be provided.
- IP20 enclosure, indoor use only, 5…40 °C ambient, illuminance at least 300 lx, altitude up to 2000 m; air-cooled so free space around the unit is mandatory.
- Pilot ONE offers configurable power-failure response (OFF/Standby, ON, Power Failure Auto) and a Process Safety mode that releases maximum cooling at the OT trigger instead of switching off.
- Leveling feet/parking brakes must be set before start-up; pump connections must be counterheld and torque limits respected.
Integration & Operating Notes
The 14 A three-phase load requires a 16 A fuse per phase and a site-fitted plug, since units below 63 A ship with cable only. Voltage and frequency tolerances are coupled: ±5 % voltage only with ±2 % frequency, and -5 % voltage with +2 % frequency is not allowed.
Cooling capacity falls from 23 kW at 20 °C to 5.5 kW at -20 °C, and the datasheet ratings assume 20 °C room temperature. A process sized on the 20 °C figure will be undercooled at low temperature or in a warm room approaching the 40 °C ambient limit.
The E1 pump delivers 54 l/min at zero head but only 11 l/min at 3.0 bar, and pump speed is not adjustable. The external circuit must be designed so its pressure drop leaves sufficient flow; models with 'P' in the name are the alternative for high-pressure-drop applications.
Free-standing units require 50 to 200 cm clearance on either side, and the air-cooled condenser needs sufficient free space to avoid derating. The unit is IP20 and must be installed indoors, dry, at 5–40 °C ambient and at least 300 lx illuminance, up to 2000 m altitude.
The chiller is not explosion-protected and must not be installed or started up within an ATEX zone. Refrigerant charge (1.49 kg R-449A) must also be checked against room volume to stay within the refrigerant limit per m³ of ambient air.
The unit is designed for externally closed applications, so the hydraulic circuit must be sealed. Minimum filling capacity is 18 l with 48 l expansion volume, which sets the fluid charge and expansion allowance for the loop.
The Pilot ONE allows OFF/Standby, ON or Power Failure Auto after a mains failure, and Process Safety mode releases maximum cooling at the OT trigger instead of switching off all poles. For exothermic processes the OT trigger and mode must be set deliberately, not left at default.
Any work on the coolant/refrigerant circuit must be carried out by approved refrigeration and air-conditioning system contractors. R-449A is A1 with GHS H280, so leak checks, labelling and F-gas reporting obligations apply to the 1.49 kg charge.
The unit is classified EN55011 Class A Group 1, which is the industrial disturbance class. Installation near residential or light-industrial sensitive equipment may require additional EMC measures, and electrical lines below the unit must be protected against condensate ingress.
Selection & Integration Q&A
Pick the 230-H when the process needs the larger 230-size capacity (23 kW at 20 °C, 5.5 kW at -20 °C) and a 54 l/min pump. Choose a 110Tw variant when low noise (65 dB(A)) matters more than capacity; the 110T is 74 dB(A).
A sealed external loop with G1 1/4 male mating fittings, at least 18 l fill and pressure drop compatible with 54 l/min at zero head / 11 l/min at 3.0 bar. Supply is 400 V 3~ 50 Hz, 14 A, 16 A fuse per phase; the plug is not included.
Indoors only, 5–40 °C ambient, at least 300 lx, up to 2000 m altitude, IP20, no ATEX zone, 50–200 cm wall clearance per side. Cooling capacity is rated at 20 °C room temperature and drops as ambient rises.
Refrigerant work only by approved contractors; R-449A charge 1.49 kg subject to leak checks and F-gas reporting. The unit ships with cable but no plug, so the site plug is a required extra; leveling feet/parking brakes must be set before start-up.