Peter Huber Kältemaschinenbau Unichiller 250Tw-H
Peter Huber Kältemaschinenbau Unichiller 250Tw-H — Recirculating chillers.
The Unichiller 250Tw-H is a water-cooled recirculating chiller with a 2 kW heater and a fixed-speed D3 pressure pump, covering -20 to 100 °C. It delivers 18 kW at 0 °C and 6 kW at -20 °C, positioning it as the high-capacity modification in the Unichiller series. It serves externally closed process loops requiring precise temperature control with Pilot ONE electronics.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Pump type | Safety class | Temperature range, °C | Power supply | Expansion volume, m³ | Minimum fill volume, m³ |
|---|---|---|---|---|---|---|
| Unichiller 250Tw-H Current | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 110T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 130T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 160T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.01 | — |
| Unichiller 180-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 180Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 180w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180w-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 200T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 200Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 210T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 230-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w-H | E1 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller 250T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 260T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 260T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 260Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 300T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 300T-H | D3 | III / FL | -20–100 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 300Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 300Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 350T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 350Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 500Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 700Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 800Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller GL 1000w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.05 | — |
| Unichiller GL 200w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 250w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 300w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 400w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 750w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.05 | — |
| Unichiller P180 | E2 | I / NFL | -20–40 | — | 0.048 | 0.018 |
| Unichiller P180-H | E2 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w | E2 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230 | E2 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
Full specifications
| Expansion volume | 22 l |
| Heating capacity | 2 kW |
| Minimum fill volume | 15 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | D3 |
| External sensor connection | Pt100 |
| Temperature range | -20...100 °C |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling capacity | 15 kW (at 25 °C) at 0 °C: 18 kW · at -10 °C: 11 kW · at -20 °C: 6 kW |
| Cooling water connection | G3/4 male |
| Cooling water consumption | 438 л/час (at 15°C, flow -20°C) at 15°C, flow -10°C: 594 л/час · at 15°C, flow 0°C: 786 l/h · at 15°C, flow 15°C: 852 l/h |
| Cooling water differential pressure | 1 bar (min.) |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and Series Positioning
- The 250Tw-H is a water-cooled chiller with a 2 kW heater and a non-variable D3 pressure pump (max 210 l/min, 4.7 bar).
- It is the high-capacity variant among the listed siblings (110T, 110T-H, 110Tw, 110Tw-H), which are air-cooled and have lower noise levels (65–74 dB(A)).
- The 'w' in the model code denotes water cooling; the 'H' indicates an integrated heater. The unit uses R-449A refrigerant (GWP 1397) and is classified IP20.
Performance and Behavior
- Cooling capacity is strongly temperature-dependent: 18 kW at 0 °C, 15 kW at 25 °C, 11 kW at -10 °C, and 6 kW at -20 °C.
- Cooling water consumption varies with flow temperature: 852 l/h at 15 °C, 786 l/h at 0 °C, 594 l/h at -10 °C, and 438 l/h at -20 °C, based on 15 °C cooling water inlet.
- The unit requires a minimum cooling water differential pressure of 1 bar and can withstand up to 6 bar. Cooling water inlet must be between 5 °C and 20 °C.
- The pump is not speed-adjustable; flow depends on the external circuit resistance.
Installation, Control and Operation
- Electrical supply is 400 V 3~ 50 Hz, delivered with cable but without a plug. Voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%.
- The Pilot ONE controller provides Ethernet, USB (host and device), and RS232 interfaces. A Pt100 external sensor connection is available.
- Safety features include adjustable overtemperature protection per DIN 12876, safety class III/FL, and a Process Safety mode that releases maximum cooling capacity when the OT trigger is reached.
- The unit is EMC Class A Group 1 and must be installed indoors with at least 300 lx illuminance, up to 2000 m altitude, and 50–200 cm wall clearance per side for maintenance.
Integration & Operating Notes
The chiller is delivered with a cable but without a plug, so the site must provide a suitable 400 V 3~ 50 Hz connection. Voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%, but not all combinations are allowed. Verify the supply quality before commissioning.
The unit is designed for cooling water intake between 5 °C and 20 °C. If water temperature can drop below the dew point, pipes must be insulated to prevent condensation. A minimum differential pressure of 1 bar is required, and pressure must not exceed 6 bar.
Free-standing models require a wall clearance of 50 to 200 cm on either side. The unit must be installed indoors with illuminance of at least 300 lx and up to 2000 m altitude. Leveling feet must be activated and parking brakes applied before operation.
In Process Safety mode, the controller releases maximum cooling capacity when the overtemperature protection trigger value is reached, instead of switching off all poles. This is critical for exothermic processes where cutting cooling could cause runaway temperatures.
The cooling water supply must be shut off during shorter downtimes such as overnight to prevent water damage and corrosion. Electrical lines directly below the unit must be protected against liquid ingress from condensation.
Selection & Integration Q&A
Choose the 250Tw-H for higher cooling capacity (18 kW at 0 °C vs lower capacities of 110 series) and when water cooling is available. The 110T-H is air-cooled and suited for lower capacity or where cooling water is not available. The 250Tw-H also includes a 2 kW heater.
The cooling water inlet must be between 5 °C and 20 °C, with a minimum differential pressure of 1 bar and maximum pressure of 6 bar. Flow rates vary with load: up to 852 l/h at 15 °C flow. Insulate pipes if below dew point.
The unit is IP20 and must be installed indoors in a dry location protected from water ingress. Ambient temperature must stay between 5 °C and 40 °C. It is not explosion-protected and must not be placed in an ATEX zone. Ensure at least 300 lx illuminance.
The unit is delivered without a plug; a suitable 400 V 3-phase plug must be sourced. Refrigerant circuit work requires certified contractors. Regular maintenance includes checking cooling water connections for leaks and shutting off water during downtimes. No specific maintenance interval is given in the digests.