Peter Huber Kältemaschinenbau Unichiller 300Tw-H
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 300Tw-H

Peter Huber Kältemaschinenbau Unichiller 300Tw-H — Recirculating chillers.

The Unichiller 300Tw-H is a water-cooled recirculating chiller with a 400 V three-phase feed, a Pilot ONE controller and a non-variable D3 pressure pump, covering -20 to 100 °C. Within the Unichiller series it is the 300-size Tw-H modification, distinguished by its higher cooling capacity and water-cooled condenser. It serves externally closed process loops that need stable temperature control with heat removal down to -20 °C.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Pump typeSafety classTemperature range, °CPower supplyExpansion volume, m³Minimum fill volume, m³
Unichiller 300Tw-H Current D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 110T C3I / NFL-10–40400V 3~ 50Hz0.007—
Unichiller 110T-H C3III / FL-10–100400V 3~ 50Hz0.007—
Unichiller 110Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 130T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 130T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 130Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 160T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 160T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 160Tw C3I / NFL-20–40400V 3~ 50Hz0.01—
Unichiller 180-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 180T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 180T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 180Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 180Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 180w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 180w-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 200T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 200Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 210T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 210T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 210Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 230-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 230w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 230w-H E1III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller 250T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 250T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 250Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 260T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 260T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 260Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 260Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 300T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 300T-H D3III / FL-20–100460V 3~ 60Hz0.0390.014
Unichiller 300Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 350T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 350Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 500Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw-H D3III / FL-20–100400V 3~ 50Hz0.007—
Unichiller 700Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 800Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller GL 1000w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.05—
Unichiller GL 200w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 250w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 300w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 400w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 750w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.05—
Unichiller P180 E2I / NFL-20–40—0.0480.018
Unichiller P180-H E2III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller P180w E2I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller P180w-H E2III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller P230 E2I / NFL-20–40460V 3~ 60Hz0.0480.018
Unichiller P230-H E2III / FL-20–100460V 3~ 60Hz0.0480.018

Full specifications

Core Performance & Functional Specifications
Expansion volume 22 l
Heating capacity 2 kW
Minimum fill volume 15 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type D3
External sensor connection Pt100
Temperature range -20...100 °C
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling capacity 15 kW (at 0 °C)
at -10 °C: 18 kW · at -20 °C: 13 kW
Cooling water connection G3/4 male
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

The 300Tw-H is a water-cooled (Tw) recirculating chiller with heating capability (H suffix). It is built around a refrigerant circuit using R-449A (GWP 1397) and a cooling-water heat exchanger whose wetted materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. The unit is rated IP20 and safety class III/FL. Compared with the sibling Unichiller 110T/110Tw models, the 300Tw-H is the larger-capacity frame in this data set; the 110-series siblings are air-cooled (T) or water-cooled (Tw) 110-size units with lower noise figures (65–74 dB(A)), whereas the 300Tw-H is specified for higher cooling duty. Selection between them should follow the required cooling capacity at the target temperature and the available cooling-water supply, not the model number alone.

Thermal behavior and operating envelope

Cooling capacity is strongly temperature-dependent: 18 kW at -10 °C, 13 kW at -20 °C and 15 kW at 0 °C (spec table). The heating capacity is 2 kW. The working temperature range is -20 to 100 °C. The circulation pump is a D3 pressure pump with fixed speed (not adjustable), delivering up to 210 l/min, with flow declining as external pressure drop rises (202 l/min at 1.1 bar). Because pump speed cannot be trimmed, the process loop must be designed so that its pressure drop at the required flow stays within the pump curve.

Installation, connection and control

The unit requires a 400 V 3~ 50 Hz supply with voltage tolerance ±5% and simultaneous frequency tolerance ±2%; three-phase devices below 63 A are delivered with a cable but without a plug, so a suitable plug must be provided on site. Pump connection is G1 1/4 male and cooling-water connection is G3/4 male; matching fittings and hoses are required. Cooling water must be supplied at 5–20 °C inlet with a differential pressure of at least 1 bar and a maximum pressure of 6 bar; pipes must be insulated if below the condensation point. The Pilot ONE controller provides Ethernet, USB (host and device) and RS232 interfaces, an external Pt100 sensor input, and configurable power-failure response (OFF/Standby, ON, Power Failure Auto). Adjustable overtemperature protection to DIN 12876 is provided, with a Process Safety mode that releases maximum cooling capacity when the OT trigger is reached.

Integration & Operating Notes

Process Limits
Cooling water must be 5–20 °C at 1–6 bar differential pressure

The chiller is designed for cooling-water intake between 5 °C and 20 °C. Differential pressure must be at least 1 bar and must never exceed 6 bar. Insulate cooling-water pipes if their surface temperature can fall below the condensation point. Only water compatible with the copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM materials of the cooling-water circuit should be used.

Unichiller 300Tw-H Datasheet, Installation and Maintenance
Environment
Do not install in an ATEX zone; ambient must stay 5–40 °C

The unit is not explosion-protected and must not be installed or started up inside an ATEX zone. Ambient temperature must remain between 5 °C and 40 °C for reliable operation. The IP20 rating means the enclosure must be kept in a dry indoor location protected from water and dust ingress.

Unichiller Operation Manual, Foreword; Unichiller 300Tw-H Datasheet, Installation

Selection & Integration Q&A

When should I choose the Unichiller 300Tw-H instead of a sibling such as the 110Tw-H?

Choose the 300Tw-H when the required cooling capacity at your setpoint exceeds what the 110-size frame can deliver; the 300Tw-H provides 18 kW at -10 °C, 13 kW at -20 °C and 15 kW at 0 °C. The 110Tw-H is a smaller water-cooled unit with a 65 dB(A) noise figure. Both need a cooling-water supply; confirm capacity at your actual temperature before deciding.

What cooling-water supply and fittings must be provided for the 300Tw-H?

Provide cooling water at 5–20 °C inlet with a differential pressure of at least 1 bar and a maximum of 6 bar. The cooling-water connection is G3/4 male and the pump connection is G1 1/4 male, so matching hoses and adapters are needed. Insulate cooling-water pipes if they can drop below the condensation point.

What maintenance and accessories are required beyond the base delivery?

The base delivery includes a cable but no plug for the 400 V three-phase feed, so a suitable plug must be supplied on site. Cooling-water hoses and G3/4 and G1 1/4 fittings are not included. Shut off cooling water even overnight. Refrigerant-circuit service must be performed by approved contractors; no consumables or spare-part intervals are stated in the provided documents.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller 300Tw-H Datasheet
Datasheet ·172,8 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF