Peter Huber Kältemaschinenbau Unichiller P180-H
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller P180-H

Peter Huber Kältemaschinenbau Unichiller P180-H — Recirculating chillers.

The Unichiller P180-H is a recirculating chiller with a high-discharge-pressure E2 pump, built for externally closed loops with high pressure drop. Its 100 kW cooling capacity at 20 °C and -20…100 °C range distinguish it from lower-capacity siblings. It serves as a central cooling source for industrial processes requiring stable temperature control.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Pump typeSafety classTemperature range, °CPower supplyExpansion volume, m³Minimum fill volume, m³
Unichiller P180-H Current E2III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 110T C3I / NFL-10–40400V 3~ 50Hz0.007—
Unichiller 110T-H C3III / FL-10–100400V 3~ 50Hz0.007—
Unichiller 110Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 130T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 130T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 130Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 160T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 160T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 160Tw C3I / NFL-20–40400V 3~ 50Hz0.01—
Unichiller 180-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 180T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 180T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 180Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 180Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 180w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 180w-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 200T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 200Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 210T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 210T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 210Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 230-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 230w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 230w-H E1III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller 250T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 250T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 250Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 260T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 260T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 260Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 260Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 300T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 300T-H D3III / FL-20–100460V 3~ 60Hz0.0390.014
Unichiller 300Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 300Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 350T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 350Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 500Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw-H D3III / FL-20–100400V 3~ 50Hz0.007—
Unichiller 700Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 800Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller GL 1000w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.05—
Unichiller GL 200w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 250w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 300w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 400w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 750w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.05—
Unichiller P180 E2I / NFL-20–40—0.0480.018
Unichiller P180w E2I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller P180w-H E2III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller P230 E2I / NFL-20–40460V 3~ 60Hz0.0480.018
Unichiller P230-H E2III / FL-20–100460V 3~ 60Hz0.0480.018

Full specifications

Core Performance & Functional Specifications
Expansion volume 48 l
Heating capacity 2 kW
Minimum fill volume 18 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type E2
External sensor connection Pt100
Temperature range -20...100 °C
Dimensions, Weight & Mechanical Design
Weight 492 kg
Electrical, Power & Drive Specifications
Current consumption 19 A
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling capacity 100 kW (at 20 °C)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, connection and control

Integration & Operating Notes

Mounting
Maintain 50–200 cm wall clearance per side for maintenance access

Free-standing models require a wall clearance of 50 to 200 cm on either side to allow service access and adequate airflow. Install indoors only, at an illuminance of at least 300 lx and at altitudes up to 2000 m. Ambient temperature must remain between 5 and 40 °C.

Unichiller Operation Manual, page 25
Mounting
Do not install in ATEX zones; unit is not explosion-protected

The chiller must not be installed or started up inside an ATEX zone. Additionally, the refrigerant R-449A (A1, H280) requires that the charge and room size be checked to avoid exceeding the refrigerant limit per m³ of ambient air. Ensure adequate ventilation and take leak precautions.

Unichiller Operation Manual, page 13; Datasheet page 1
Safety
Use Process Safety mode for exothermic reactions to avoid runaway

In Process Safety mode, the controller releases maximum cooling capacity when the overtemperature protection trigger value is reached, instead of switching off all poles. This is essential for exothermic processes where cutting cooling could cause a temperature runaway. Set the OT trigger value according to DIN 12876 for the application.

Unichiller Operation Manual, page 18; Datasheet page 1
Operation
Shut off cooling water supply even for short downtimes

For water-cooled units, the cooling water supply must be shut off during downtimes, including overnight, to prevent water damage and corrosion. Also protect electrical lines directly below the unit against liquid ingress from condensation.

Unichiller Operation Manual, page 28
Mounting
Activate leveling feet and parking brakes before operation

Before switching on, unscrew or activate the leveling feet and apply parking brakes at the casters (if any) to prevent the unit from moving. This avoids crushing injuries and ensures stable operation.

Unichiller Operation Manual, page 31

Selection & Integration Q&A

When should I choose the Unichiller P180-H instead of a sibling model like the Unichiller 110T or 110Tw?

Choose the P180-H when your process loop has high pressure drop and requires up to 100 kW cooling at 20 °C. The 'P' pump delivers high discharge pressure. Siblings like the 110T offer lower capacity and different noise levels; select them for lower loads or where noise is critical.

What are the requirements for connecting the P180-H to my external circuit and electrical supply?

The external circuit must be closed and pressure-tight. Pump connections are G1 1/4 AG male; observe maximum tightening torques and counterhold fittings. Electrical supply is 400 V 3~ 50 Hz, 19 A max, 3x20 A fusing; the unit comes with cable but no plug, so site termination is needed.

What mounting and environmental limits apply to the P180-H according to the manual?

Install indoors only, with 50–200 cm wall clearance per side, ambient temperature 5–40 °C, illuminance ≥300 lx, and altitude ≤2000 m. Do not install in ATEX zones. Ensure room volume is sufficient for the R-449A charge. Activate leveling feet and parking brakes before operation.

What maintenance and service requirements should I plan for, and what is not included in the base delivery?

Refrigerant circuit work must be by approved contractors. Shut off cooling water during downtimes. The base delivery includes cable but no plug; mating connectors, hoses, and fittings for the G1 1/4 connections are not included. No specific maintenance interval is stated in the provided documents.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller P180-H Technical Data
Datasheet ·121,6 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF