Peter Huber Kältemaschinenbau Unichiller 130Tw
Peter Huber Kältemaschinenbau Unichiller 130Tw — Recirculating chillers with Noise level 65 dB.
The Unichiller 130Tw is a water-cooled recirculating chiller for externally closed circuits, rated -20 to 40 °C with a C3 fixed-speed pump (96 l/min, 5.6 bar). Its 'w' variant is the low-noise (65 dB(A)) water-cooled member of the 110/130 series, distinguished from air-cooled siblings by a G3/4 cooling-water connection and 150–500 l/h water demand.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Pump type | Safety class | Temperature range, °C | Power supply | Expansion volume, m³ | Minimum fill volume, m³ |
|---|---|---|---|---|---|---|
| Unichiller 130Tw Current | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 110T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 130T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.01 | — |
| Unichiller 180-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 180Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 180w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180w-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 200T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 200Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 210T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 230-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w-H | E1 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller 250T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 260T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 260Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 300T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 300T-H | D3 | III / FL | -20–100 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 300Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 300Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 350T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 350Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 500Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 700Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 800Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller GL 1000w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.05 | — |
| Unichiller GL 200w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 250w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 300w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 400w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 750w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.05 | — |
| Unichiller P180 | E2 | I / NFL | -20–40 | — | 0.048 | 0.018 |
| Unichiller P180-H | E2 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w | E2 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230 | E2 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
Full specifications
| Noise level | 65 dB(A) |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | C3 |
| External sensor connection | Pt100 |
| Temperature range | -20...40 °C |
| Weight | 229 kg |
| Current consumption | 12 A (max.) |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water consumption | 150 л/час (at 15°C, flow -10°C) at 15°C, flow 0°C: 280 l/h · at 15°C, flow 15°C: 500 l/h |
| Cooling water differential pressure | 1 bar (min.) |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
- Water-cooled condensing via a G3/4 male cooling-water connection; no air-side condenser, so heat is rejected to the facility water loop rather than the room.
- The 'w' suffix marks the water-cooled, 65 dB(A) variant; air-cooled siblings Unichiller 110T and 110T-H are rated 74 dB(A), and the 110Tw/110Tw-H share the 65 dB(A) water-cooled platform at lower capacity.
- Refrigerant R-449A (GWP 1397, CO2 equivalent 2.9 t); cooling-water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.
- IP20 enclosure, 400 V 3~ 50 Hz supply, 12 A max current, 229 kg.
Behavior and media
- Working range -20 to 40 °C; intended only for externally closed applications.
- Cooling-water inlet must be 5–20 °C; as water temperature rises, cooling power falls and required flow rises — consumption is 150 l/h at -10 °C, 280 l/h at 0 °C and 500 l/h at 15 °C (all at 15 °C water).
- Cooling-water differential pressure min. 1 bar, max. pressure 6 bar.
- C3 pump is not speed-adjustable; max delivery 96 l/min at 5.6 bar, so the external loop pressure drop must be checked against this fixed curve.
Control and operation
- Pilot ONE controller with Ethernet, USB (host and device) and RS232; external Pt100 sensor input.
- Power-failure response configurable as OFF/Standby, ON or Power Failure Auto.
- Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles — relevant for exothermic loads.
- Safety class I / NFL per DIN 12876.
Integration & Operating Notes
Below 1 bar differential the chiller cannot reject its rated load; above 6 bar the water circuit is at risk. Cooling power drops and flow demand rises as inlet water warms, so size the supply for the worst-case 20 °C condition, not the nominal 15 °C.
Leaving the water supply open during idle periods invites condensation, corrosion and water damage inside the unit. Isolate the supply whenever the chiller is not actively cooling, including short overnight stops.
Free-standing installation requires this access envelope on either side for maintenance. Also protect any electrical lines routed directly below the unit against liquid ingress from condensation, and keep the unit out of ATEX zones — it is not explosion-protected.
Three-phase devices drawing under 63 A ship with a cable only; the mating plug is a site responsibility. Supply tolerance is ±5% voltage with simultaneous ±2% frequency, and only permitted combinations apply. The unit is EN 55011 Class A, Group 1, so EMC measures may be needed in sensitive environments.
The chiller is designed exclusively for externally closed applications. An open bath or vented tank will not match the hydraulic design and risks pump and temperature-control problems. Verify the external loop pressure drop against the fixed C3 pump curve (96 l/min, 5.6 bar).
Pilot ONE offers OFF/Standby, ON or Power Failure Auto after mains loss. For processes that generate heat, Process Safety mode should be configured so maximum cooling is released at the overtemperature trigger rather than all poles switching off.
Any intervention on the refrigerant circuit must be performed by approved refrigeration/air-conditioning contractors. The R-449A charge (GWP 1397, 2.9 t CO2e) also triggers leak-check and reporting obligations; confirm room volume against the charge limit at installation.
Selection & Integration Q&A
Pick the 130Tw when heat must be rejected to facility water and low noise (65 dB(A)) matters; air-cooled 110T/110T-H are 74 dB(A) and dump heat into the room. The 110Tw shares the water-cooled 65 dB(A) platform but at lower capacity — choose it only if its cooling output covers the load.
A G3/4 male connection fed with 5–20 °C water at 1–6 bar differential; flow demand is 150–500 l/h depending on setpoint. Use only water compatible with copper, 1.4401 stainless, MS, PA, PPE, PTFE and EPDM, and insulate pipes where condensation can form.
Indoor use only, 5–40 °C ambient, ≥300 lx lighting, up to 2000 m altitude, 50–200 cm side clearance, leveling feet/parking brakes set before start-up, and no ATEX zone. IP20 means a dry location protected from water ingress.
The three-phase cable is supplied without a plug, so the mating plug is on site. Refrigerant-circuit service is limited to approved contractors; the R-449A charge (2.9 t CO2e) carries leak-check and reporting duties, and cooling water must be isolated during downtime.