Peter Huber Kältemaschinenbau Unichiller 210T-H
Peter Huber Kältemaschinenbau Unichiller 210T-H — Recirculating chillers.
The Unichiller 210T-H is an air-cooled, externally closed recirculating chiller with a D3 pressure pump (G1 1/4 male, 210 l/min at zero head) and Pilot ONE controller, rated -20 to 100 °C. It delivers 21 kW at 0 °C and 15 kW at 15 °C, with 2 kW heating and 39 l expansion volume. It serves closed process loops requiring three-phase 400 V supply and IP20 indoor installation.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Pump type | Safety class | Temperature range, °C | Power supply | Expansion volume, m³ | Minimum fill volume, m³ |
|---|---|---|---|---|---|---|
| Unichiller 210T-H Current | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 110T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 130T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 160T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.01 | — |
| Unichiller 180-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 180Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 180w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180w-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 200T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 200Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 210T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 230-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w-H | E1 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller 250T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 260T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 260Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 300T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 300T-H | D3 | III / FL | -20–100 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 300Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 300Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 350T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 350Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 500Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 700Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 800Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller GL 1000w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.05 | — |
| Unichiller GL 200w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 250w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 300w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 400w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 750w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.05 | — |
| Unichiller P180 | E2 | I / NFL | -20–40 | — | 0.048 | 0.018 |
| Unichiller P180-H | E2 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w | E2 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230 | E2 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
Full specifications
| Expansion volume | 39 l |
| Heating capacity | 2 kW |
| Minimum fill volume | 14 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | D3 |
| External sensor connection | Pt100 |
| Temperature range | -20...100 °C |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling capacity | 15 kW (at 15 °C) at 0 °C: 21 kW |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Air-cooled, CFC/HCFC-free refrigeration circuit using R-449A (GHS H280); IP20 enclosure; overall dimensions 905x2172x1900 mm (WxDxH).
- Fixed-speed D3 pressure pump with G1 1/4 male connection; pump speed is not adjustable.
- Within the Unichiller series, this 210T-H is the larger-capacity modification; sibling 110T/110T-H and 110Tw/110Tw-H models are smaller and differ in noise level (74 dB(A) for 110T/110T-H, 65 dB(A) for 110Tw/110Tw-H). The 210T-H is distinguished by its 21 kW at 0 °C cooling capacity and 39 l expansion volume.
- Intended only for externally closed applications; models with 'P' in the model name are the ones suited to high pressure-drop loops, so the 210T-H should be matched to low-to-moderate pressure-drop circuits.
Behavior under load and media
- Cooling capacity is strongly temperature-dependent: 21 kW at 0 °C, 15 kW at 15 °C, and 7.5 kW at -20 °C (datasheet). Low-temperature duty therefore requires derating or a larger unit.
- Pump curve: 210 l/min at zero head, falling to 31 l/min at 4.5 bar; the external circuit pressure drop must be checked against this curve to avoid flow collapse.
- Hydraulic volume: minimum filling capacity 14 l, expansion volume 39 l; the external loop volume must be considered for proper operation.
- Heating capacity 2 kW; operating temperature range -20 to 100 °C; ambient temperature limit 5 to 40 °C, with cooling capacity possibly falling above 20 °C room temperature.
Installation, control and operation
- Power supply 400 V 3~ 50 Hz; three-phase devices below 63 A are delivered with cable but without a plug, so the customer provides the final connection. Voltage tolerance ±5% and frequency tolerance ±2% must not be combined in opposite directions.
- Pilot ONE controller with Ethernet, USB (Host and Device) and RS232 interfaces; external Pt100 sensor connection; adjustable overtemperature protection per DIN 12876, safety class III/FL.
- Power-failure response configurable as OFF/Standby, ON, or Power Failure Auto; Process Safety mode releases maximum cooling capacity when the OT trigger value is reached instead of switching off all poles.
- Installation: indoors only, illuminance at least 300 lx, up to 2000 m altitude; wall clearance 50 to 200 cm per side for maintenance; leveling feet/parking brakes must be activated before operation; pump connections must be counterheld and maximum tightening torques not exceeded.
- Refrigerant charge and room size must be checked to avoid exceeding the refrigerant limit per m³ of ambient air; EMC Class A Group 1 (EN55011) may cause radio interference in residential areas.
Integration & Operating Notes
The datasheet rates 21 kW at 0 °C but only 7.5 kW at -20 °C. Processes running near the cold end of the -20 to 100 °C range must be sized against the derated capacity, not the 0 °C figure, or the unit will be undersized.
The D3 pump delivers 210 l/min at zero head but only 31 l/min at 4.5 bar. The external circuit resistance must be calculated and compared with the pump curve; a high-pressure-drop loop will starve flow. Models with 'P' in the name are the ones intended for high pressure drops.
Voltage tolerance is ±5% and frequency tolerance ±2%, but the manual states they must not be combined in opposite directions. Site power quality must be verified before connection; the unit is delivered with cable but without a plug for three-phase devices below 63 A.
The unit is IP20 and must be installed indoors, protected from water and dust, at illuminance of at least 300 lx and up to 2000 m altitude. A wall clearance of 50 to 200 cm per side is required for maintenance; leveling feet or parking brakes must be activated before startup.
In Process Safety mode the controller releases maximum cooling capacity when the adjustable overtemperature trigger is reached, instead of switching off all poles. For exothermic processes this prevents runaway; for other processes the standard cutoff behavior may be preferred. The mode is selected via the Pilot ONE.
Any work on the refrigerant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The unit uses R-449A (GHS H280), and refrigerant charge versus room size must be checked at installation to stay within the limit per m³ of ambient air.
The equipment is classified EMC Class A Group 1 to EN55011, so it may cause radio interference in residential areas. Placement near sensitive measurement or control electronics should account for this, and appropriate installation measures may be needed.
Selection & Integration Q&A
Pick the 210T-H when the process needs up to 21 kW at 0 °C and a 39 l expansion volume in a closed loop. The 110T/110T-H and 110Tw/110Tw-H are smaller-capacity units; the Tw variants are quieter (65 dB(A) vs 74 dB(A)) but do not match the 210T-H cooling capacity.
Provide a 400 V 3~ 50 Hz supply with voltage/frequency tolerances not combined adversely; three-phase units below 63 A come with cable but no plug. The external loop must mate with the G1 1/4 male pump connection, hold at least 14 l fill volume, and its pressure drop must be checked against the pump curve (210 l/min at 0 bar, 31 l/min at 4.5 bar).
Install indoors only, IP20, at 5–40 °C ambient, illuminance ≥300 lx, altitude ≤2000 m, with 50–200 cm wall clearance per side. Not permitted in ATEX zones. Leveling feet/parking brakes must be set before operation. Cooling capacity may fall above 20 °C room temperature.
The base delivery includes cable but no plug for three-phase units below 63 A. Refrigerant circuit work must be done by approved refrigeration contractors; R-449A charge and room size must be checked. The Pilot ONE offers Ethernet, USB and RS232; an external Pt100 sensor can be connected. No consumables or spare parts are listed in the provided documents.