Peter Huber Kältemaschinenbau Unichiller 350Tw-H
Peter Huber Kältemaschinenbau Unichiller 350Tw-H — Recirculating chillers.
The Unichiller 350Tw-H is a water-cooled recirculating chiller with a 400 V 3~ 50 Hz supply, a D3 pump delivering up to 210 l/min at 4.7 bar, and a -20...100 °C working range. It distinguishes itself within the series by its high-capacity 350 size and water-cooled condenser, requiring a 1–6 bar cooling-water differential. It is intended for externally closed process loops where heat loads up to 25 kW at 0 °C must be rejected.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Pump type | Safety class | Temperature range, °C | Power supply | Expansion volume, m³ | Minimum fill volume, m³ |
|---|---|---|---|---|---|---|
| Unichiller 350Tw-H Current | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 110T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 130T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 160T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.01 | — |
| Unichiller 180-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 180Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 180w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180w-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 200T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 200Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 210T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 230-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w-H | E1 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller 250T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 260T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 260Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 300T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 300T-H | D3 | III / FL | -20–100 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 300Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 300Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 350T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 500Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 700Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 800Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller GL 1000w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.05 | — |
| Unichiller GL 200w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 250w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 300w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 400w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 750w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.05 | — |
| Unichiller P180 | E2 | I / NFL | -20–40 | — | 0.048 | 0.018 |
| Unichiller P180-H | E2 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w | E2 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230 | E2 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
Full specifications
| Expansion volume | 22 l |
| Heating capacity | 2 kW |
| Minimum fill volume | 15 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | D3 |
| External sensor connection | Pt100 |
| Temperature range | -20...100 °C |
| Weight | 387 kg |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling capacity | 15 kW (at 15 °C) at 0 °C: 25 kW · at -10 °C: 16 kW · at -20 °C: 10 kW |
| Cooling water connection | G3/4 male |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The 350Tw-H is the water-cooled variant of the 350 size; sibling models 110T/110T-H and 110Tw/110Tw-H are smaller-capacity air- and water-cooled units. The 'w' denotes water cooling, and the 'H' denotes the heating function (2 kW).
- The hydraulic circuit uses a fixed-speed D3 pump (no speed adjustment) with G1 1/4 AG male process connections and G3/4 male cooling-water connections. The expansion tank is atmospheric and open, with a 22 l expansion volume and a 15 l minimum fill.
- The refrigerant is R-449A (A1, H280), charge 3 kg, GWP 1397 (4.2 t CO2e). The unit is rated IP20 and safety class III/FL.
Behavior under load and media
- Cooling capacity is rated at 25 kW at 0 °C, 16 kW at -10 °C, 10 kW at -20 °C, and 15 kW at 15 °C, valid for 20 °C room and 15 °C cooling-water inlet. Capacity falls as process temperature decreases, so the -20 °C end of the range is limited to lower heat loads.
- The pump curve is not flat: maximum delivery is 210 l/min at 4.7 bar, falling to 140 l/min at 3.0 bar and 77 l/min at 4.0 bar. External circuit pressure drop must be checked against this curve.
- The unit is designed for externally closed applications only; the open expansion tank means the thermal fluid is exposed to atmosphere and must be compatible with the circuit materials (copper, stainless steel 1.4401, MS, PA, PPE, PTFE, EPDM).
Installation, control and operation
- The Pilot ONE controller provides Ethernet, USB (host and device) and RS232 interfaces, plus a Pt100 external sensor connection. Power-failure response is configurable (OFF/Standby, ON, Power Failure Auto), and Process Safety mode releases maximum cooling on overtemperature trigger.
- Electrical supply is 400 V 3~ 50 Hz with max current 19.5 A and 3×20 A fuses; three-phase devices under 63 A are supplied with cable but without plug. Voltage and frequency tolerances are ±5% and ±2%, but simultaneous opposite-sign deviations are not allowed.
- Installation requires indoor use only, 5–40 °C ambient, at least 300 lx illuminance, up to 2000 m altitude, and 50–200 cm wall clearance per side for maintenance. Cooling-water inlet must be 5–20 °C; pipes below dew point must be insulated. Electrical lines directly below the unit must be protected against condensate.
Integration & Operating Notes
Operating outside this window invalidates the published cooling capacities and can cause low-temperature or high-temperature faults. If the inlet is below the ambient dew point, insulate all cooling-water pipes to prevent condensation on the circuit and on electrical components.
The unit accepts ±5% voltage and ±2% frequency, but a simultaneous -5% voltage with +2% frequency (or vice versa) is not permitted. Verify site power quality before connection; a 3×20 A fuse and a suitable plug (not supplied) are required on the 400 V 3~ 50 Hz feed.
Free-standing installation requires this wall clearance to allow maintenance access to the refrigerant circuit and pump. The unit must be indoors, at 5–40 °C ambient, with at least 300 lx illuminance and at a site no higher than 2000 m altitude. Leveling feet must be extended and parking brakes applied before start-up.
The chiller is designed only for externally closed applications; an open loop will not work with the atmospheric expansion tank. Size the external circuit so its pressure drop at the required flow intersects the D3 pump curve (210 l/min at 4.7 bar, 140 l/min at 3.0 bar, 77 l/min at 4.0 bar). Minimum fill is 15 l.
The Pilot ONE can be set to OFF/Standby, ON, or Power Failure Auto after a mains interruption. For exothermic processes, Process Safety mode releases maximum cooling when the overtemperature trigger is reached instead of switching all poles off. Choose the mode that matches the process hazard before commissioning.
Any intervention on the refrigerant circuit (R-449A, 3 kg, GWP 1397) must be performed by approved refrigeration/air-conditioning system contractors. Check room volume against the refrigerant charge limit per m³ at installation. Shut off cooling water even for overnight downtime to prevent condensate and corrosion.
Water-cooled units can produce condensation; electrical lines routed directly below the chiller must be protected against liquid ingress. The unit is IP20 and Class A, Group 1 per EN55011, so maintain separation from sensitive equipment or add filtering if required.
Selection & Integration Q&A
Choose the 350Tw-H when the heat load at 0 °C approaches 25 kW and a water-cooled condenser is acceptable. The 110Tw-H is smaller and water-cooled; the 110T-H is air-cooled and needs no cooling-water circuit but has a higher noise level (74 dB(A) vs 65 dB(A) for the 110Tw).
Provide cooling water at 5–20 °C with 1 bar minimum differential pressure and 6 bar maximum, connected via G3/4 male fittings. Electrical supply is 400 V 3~ 50 Hz, max 19.5 A, 3×20 A fuses; a cable is supplied but no plug. Process connections are G1 1/4 AG male.
Indoors only, 5–40 °C ambient, at least 300 lx illuminance, max 2000 m altitude, and 50–200 cm clearance per side. The unit is not ATEX-rated and must not be installed in a hazardous zone. Leveling feet and parking brakes must be set before operation.
The base delivery includes a cable but no plug; the customer must supply a suitable 3-phase plug and fusing. Refrigerant circuit service is restricted to approved contractors. No consumables or spare parts are listed in the provided documents; consult the manual's maintenance section for intervals.