Peter Huber Kältemaschinenbau Unichiller 110T
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 110T

Peter Huber Kältemaschinenbau Unichiller 110T — Recirculating chillers with Noise level 74 dB.

The Unichiller 110T is a floor-standing, air-cooled recirculating chiller with a C3 fixed-speed pump and G1 1/4 male connections, rated for -10 to 40 °C in externally closed loops. Within the series it is the base 110 frame; the -H and Tw siblings add high-pressure-drop or water-cooled capability. It serves as the primary cooling source for closed process circuits.

Noise level
74 dB

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Pump typeSafety classTemperature range, °CPower supplyExpansion volume, m³Minimum fill volume, m³
Unichiller 110T Current C3I / NFL-10–40400V 3~ 50Hz0.007—
Unichiller 110T-H C3III / FL-10–100400V 3~ 50Hz0.007—
Unichiller 110Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 130T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 130T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 130Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 160T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 160T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 160Tw C3I / NFL-20–40400V 3~ 50Hz0.01—
Unichiller 180-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 180T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 180T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 180Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 180Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 180w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 180w-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 200T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 200Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 210T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 210T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 210Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 230-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 230w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 230w-H E1III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller 250T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 250T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 250Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 260T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 260T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 260Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 260Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 300T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 300T-H D3III / FL-20–100460V 3~ 60Hz0.0390.014
Unichiller 300Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 300Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 350T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 350Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 500Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw-H D3III / FL-20–100400V 3~ 50Hz0.007—
Unichiller 700Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 800Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller GL 1000w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.05—
Unichiller GL 200w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 250w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 300w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 400w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 750w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.05—
Unichiller P180 E2I / NFL-20–40—0.0480.018
Unichiller P180-H E2III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller P180w E2I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller P180w-H E2III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller P230 E2I / NFL-20–40460V 3~ 60Hz0.0480.018
Unichiller P230-H E2III / FL-20–100460V 3~ 60Hz0.0480.018

Full specifications

Core Performance & Functional Specifications
Noise level 74 dB(A)
Expansion volume 7 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type C3
External sensor connection Pt100
Temperature range -10...40 °C
Dimensions, Weight & Mechanical Design
Weight 230 kg
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

Working range and behavior

Control, installation and operation

Integration & Operating Notes

Mounting
Reserve 50–200 cm clearance per side for service access

Free-standing installation requires a wall clearance of 50 to 200 cm on either side so maintenance can be performed. Plan the footprint around the 600 × 790 × 1614 mm cabinet and 230 kg weight, and verify floor loading at the final location.

Unichiller Operation Manual, Installation
Electrical
Provide a three-phase plug; unit ships with cable only

Three-phase devices with current consumption below 63 A are delivered with a cable but without a plug, so the customer must supply and install a suitable plug. Confirm site supply is 400 V 3~ 50 Hz with voltage tolerance ±5% only when frequency tolerance is ±2%.

Unichiller 110T Datasheet, Technical Specifications
Process Limits
Process loop must be externally closed to atmosphere

The chiller is designed only for externally closed applications. An open bath or vented circuit is not permitted; the hydraulic circuit must be designed as a closed loop with the G1 1/4 male pump connections matched by suitable hoses or fittings.

Unichiller 110T Datasheet, Product Overview
Process Limits
Cooling capacity falls when ambient exceeds 20 °C

Output data are valid at 20 °C room temperature; if ambient temperature rises, cooling capacity may drop. Size the chiller against the actual site ambient (permitted 5–40 °C), not the nominal rating, and keep the installation indoors at ≥300 lx and ≤2000 m altitude.

Unichiller 110T Datasheet, Technical Specifications
Control & I/O
Configure power-failure response before connecting the process

The Pilot ONE allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling capacity at the overtemperature trigger instead of switching off all poles; select the mode that matches the process risk.

Unichiller Operation Manual, Normal Operation / Safety
Maintenance
Refrigerant circuit work requires certified contractors

Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The unit uses R-449A (A1, H280, GWP 1397); observe refrigerant charge and room size at installation to stay within the limit per m³ of ambient air.

Unichiller Operation Manual, Maintenance / Installation
Environment
Do not install in an ATEX zone; unit is not explosion-protected

The chiller must not be installed or started up inside an ATEX zone. It is EMC Class A, Group 1 to EN55011, so additional measures may be needed near sensitive or residential equipment, and IP20 protection means it must be kept from water and dust ingress.

Unichiller Operation Manual, Foreword / Datasheet

Selection & Integration Q&A

When should I choose the 110T instead of the 110T-H or 110Tw?

Pick the 110T for standard pressure-drop, air-cooled duty at 74 dB(A). Choose the 110T-H if the process loop has high pressure drop (the 'P'/'H' variant is suited to that). Choose the 110Tw or 110Tw-H when water cooling is available and lower noise (65 dB(A)) matters.

What supply and mating hardware must be provided on site for the 110T?

A 400 V 3~ 50 Hz three-phase supply with ±5% voltage / ±2% frequency tolerance; the unit ships with cable but no plug below 63 A, so the customer installs the plug. Process hoses/fittings must match the G1 1/4 male pump connections, and the loop must be externally closed.

What installation and duty-cycle limits does the manual impose on this chiller?

Indoors only, 5–40 °C ambient, ≥300 lx, ≤2000 m altitude, no ATEX zone. Keep 50–200 cm clearance per side, set leveling feet/parking brakes before start-up, and shut off cooling water even for overnight downtime. Cooling capacity is rated at 20 °C room temperature.

What maintenance and accessories are not included in the base delivery?

The three-phase plug is not included. Refrigerant-circuit work must be done by approved refrigeration contractors; the R-449A charge and room size must be checked at installation. The expansion tank is atmospheric open with a standard cover, which must be handled correctly during filling and venting.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF
PDF
Unichiller 110T Datasheet
Datasheet ·165,2 КБ ·2 pp.
View PDF