Peter Huber Kältemaschinenbau Unichiller 180-H
Peter Huber Kältemaschinenbau Unichiller 180-H — Recirculating chillers.
The Unichiller 180-H is a 476 kg, 400 V three-phase recirculating chiller with a -20 to 100 °C working range and 18 kW cooling at 20 °C. It is the high-capacity modification in the Unichiller series, distinguished by its 48 l expansion volume and 2 kW heating. It serves externally closed process loops requiring industrial-grade temperature control.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Pump type | Safety class | Temperature range, °C | Power supply | Expansion volume, m³ | Minimum fill volume, m³ |
|---|---|---|---|---|---|---|
| Unichiller 180-H Current | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 110T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 130T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 160T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.01 | — |
| Unichiller 180T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 180Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 180w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180w-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 200T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 200Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 210T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 230-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w-H | E1 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller 250T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 260T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 260Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 300T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 300T-H | D3 | III / FL | -20–100 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 300Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 300Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 350T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 350Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 500Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 700Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 800Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller GL 1000w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.05 | — |
| Unichiller GL 200w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 250w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 300w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 400w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 750w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.05 | — |
| Unichiller P180 | E2 | I / NFL | -20–40 | — | 0.048 | 0.018 |
| Unichiller P180-H | E2 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w | E2 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230 | E2 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
Full specifications
| Expansion volume | 48 l |
| Heating capacity | 2 kW |
| Minimum fill volume | 18 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | E1 |
| External sensor connection | Pt100 |
| Temperature range | -20...100 °C |
| Weight | 476 kg |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
The Unichiller 180-H is a free-standing, IP20-rated recirculating chiller built for indoor industrial environments. Its overall dimensions of 938×1288×2003 mm and 476 kg weight require verified floor loading and access routes. Within the Unichiller series, this model is the high-capacity variant: sibling models such as the 110T and 110Tw are smaller and differ in noise level (74 dB(A) and 65 dB(A) respectively), while the 180-H is specified for larger cooling loads. The unit uses refrigerant R-449A (1.7 kg, CO2 equivalent 2.4 t) and is classified for Class A, Group 1 EMC disturbances per EN55011, restricting use to industrial environments.
Thermal behavior and process limits
The operating temperature range is -20 to 100 °C, but cooling capacity is strongly temperature-dependent: 18 kW at 20 °C falls to 3.5 kW at -20 °C. Sizing must therefore be based on the coldest required set point. The unit includes a 2 kW heater for precise temperature control above ambient. It is designed exclusively for externally closed applications; open-loop or tank operation is not supported. The hydraulic circuit has a minimum fill volume of 18 l and an expansion volume of 48 l, so external loop volume must be considered to avoid insufficient fill or overflow. The pump connection is G1 1/4 male, and the pump is non-adjustable (type E1).
Control, installation and operation
The Pilot ONE controller provides Ethernet, USB (Host and Device) and RS232 interfaces, plus a Pt100 external sensor connection. Power-failure response is configurable (OFF/Standby, ON, or Power Failure Auto). Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached, rather than switching off all poles. Installation requires a 400 V 3~ 50 Hz supply with 3×16 A fusing; three-phase devices below 63 A are delivered with cable but without a plug. Voltage and frequency tolerances are interdependent (-5% voltage with +2% frequency is not allowed). Ambient temperature must stay within 5–40 °C, and cooling capacity may drop above 20 °C. The unit must not be installed in ATEX zones, requires indoor use with at least 300 lx illuminance, and needs 50–200 cm wall clearance per side for maintenance. Leveling feet or parking brakes must be engaged before operation.
Integration & Operating Notes
The unit requires a three-phase 400 V 50 Hz supply protected by 3×16 A fuses. Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide a suitable plug. Voltage and frequency tolerances are interdependent: -5% voltage with +2% frequency is not allowed, so site supply quality must be verified before connection.
The unit is IP20 and must be installed indoors in a dry location with at least 300 lx illuminance and up to 2000 m altitude. A wall clearance of 50 to 200 cm per side is required for maintenance access. Leveling feet must be unscrewed or parking brakes activated before operation to prevent movement. Electrical lines directly below a water-cooled unit must be protected against liquid ingress from condensation.
The chiller is designed exclusively for externally closed applications; open-loop or tank operation is not supported. The minimum filling capacity is 18 l and expansion volume is 48 l, so the external loop volume must be sized to stay within these limits. The pump connection is G1 1/4 male, and maximum tightening torques must not be exceeded during hose installation.
The Pilot ONE controller allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles. These settings must be configured according to the process risk assessment before commissioning.
Work on the refrigerant circuit must only be performed by approved refrigeration/air-conditioning system contractors. The refrigerant charge is 1.7 kg R-449A with a CO2 equivalent of 2.4 t, which triggers leak-check and F-gas record-keeping obligations. The refrigerant charge and room size must be checked at installation to avoid exceeding the refrigerant limit per m³ of ambient air.
The unit is not explosion-protected and must not be installed or started up inside an ATEX zone. It emits Class A, Group 1 disturbances per EN55011, so it is intended for industrial environments and may cause interference in residential areas. Ambient temperature must stay within 5–40 °C; cooling capacity may drop above 20 °C.
Selection & Integration Q&A
Choose the 180-H when your process requires up to 18 kW cooling at 20 °C and a -20 to 100 °C range with 2 kW heating. The 110T and 110Tw are smaller-capacity models; select them only if their lower cooling output and different noise levels (74 dB(A) and 65 dB(A)) suit your duty. The 180-H also provides 48 l expansion volume for larger external loops.
The unit requires a 400 V 3~ 50 Hz supply with 3×16 A fusing. Because it is a three-phase device below 63 A, it is delivered with a cable but without a plug, so you must provide a suitable plug. Verify that site voltage and frequency tolerances do not combine -5% voltage with +2% frequency, as this is not allowed.
Install indoors only, with at least 300 lx illuminance and a wall clearance of 50–200 cm per side for maintenance. Ambient temperature must be 5–40 °C. The unit is IP20 and not ATEX-rated, so it must not be placed in hazardous zones. Engage leveling feet or parking brakes before operation.
Refrigerant circuit work must be performed by approved refrigeration contractors. The 1.7 kg R-449A charge (2.4 t CO2e) requires leak checks and F-gas records. The unit is delivered with a cable but no plug; mating hoses and fittings for the G1 1/4 male pump connection are not included. No consumables or accessories are specified in the base delivery.