Peter Huber Kältemaschinenbau Unichiller 210T
Peter Huber Kältemaschinenbau Unichiller 210T — Recirculating chillers.
The Unichiller 210T is an air-cooled recirculating chiller with a 15 kW cooling capacity at 15 °C and a -20 to 40 °C working range, positioned as the larger-capacity variant in the Unichiller series. Its non-variable D3 pressure pump delivers up to 210 l/min at zero head, making it suited to closed-loop process cooling where high flow and moderate pressure are required.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Pump type | Safety class | Temperature range, °C | Power supply | Expansion volume, m³ | Minimum fill volume, m³ |
|---|---|---|---|---|---|---|
| Unichiller 210T Current | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 110T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 130T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 160T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.01 | — |
| Unichiller 180-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 180Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 180w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180w-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 200T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 200Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 210T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 230-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w-H | E1 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller 250T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 260T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 260Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 300T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 300T-H | D3 | III / FL | -20–100 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 300Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 300Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 350T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 350Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 500Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 700Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 800Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller GL 1000w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.05 | — |
| Unichiller GL 200w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 250w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 300w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 400w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 750w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.05 | — |
| Unichiller P180 | E2 | I / NFL | -20–40 | — | 0.048 | 0.018 |
| Unichiller P180-H | E2 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w | E2 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230 | E2 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
Full specifications
| Expansion volume | 39 l |
| Minimum fill volume | 14 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | D3 |
| External sensor connection | Pt100 |
| Temperature range | -20...40 °C |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling capacity | 15 kW (at 15 °C) at 0 °C: 21 kW |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
The Unichiller 210T is a free-standing, air-cooled recirculating chiller using R-449A refrigerant (GWP 1397) and a non-variable D3 pressure pump. Within the Unichiller series, it is the higher-capacity sibling to the 110T variants; the 110T models are distinguished by lower noise levels (74 dB(A) for 110T/110T-H, 65 dB(A) for 110Tw/110Tw-H), whereas the 210T is specified for greater cooling duty. The unit is intended exclusively for externally closed applications and is not explosion-protected, so it must not be installed in an ATEX zone.
Performance and load behavior
Cooling capacity is 21 kW at 0 °C and 15 kW at 15 °C, dropping to 7.5 kW at -20 °C; low-temperature duty therefore requires either a larger unit or reduced load. The pump curve shows maximum delivery of 210 l/min at zero head falling to 31 l/min at 4.5 bar, so the system pressure drop must be checked against this curve to ensure adequate flow. Ambient temperature must remain between 5 and 40 °C, and capacity may fall if ambient rises.
Installation, control and operation
The Pilot ONE controller provides Ethernet, USB (Host and Device) and RS232 interfaces, plus a Pt100 external sensor connection. Power-failure response is configurable as OFF/Standby, ON, or Power Failure Auto. Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached, rather than switching off all poles. The unit requires a 400 V three-phase 50 Hz supply; three-phase devices below 63 A are delivered with cable but without a plug. The pump connection is G1 1/4 male, and minimum filling capacity is 14 l with 39 l expansion volume. IP20 protection and EMC Class A Group 1 mean the unit must be installed indoors in a dry location and may cause radio interference in residential areas.
Integration & Operating Notes
Free-standing models require a wall clearance of 50 to 200 cm on either side to allow maintenance and ensure adequate air flow for heat rejection. The unit must be installed indoors at an illuminance of at least 300 lx and at altitudes up to 2000 m.
The unit requires a 400 V three-phase 50 Hz supply. Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%; combinations such as -5% voltage with +2% frequency are not allowed. Three-phase devices below 63 A are delivered with cable but without a plug, so the site plug must be provided.
In Process Safety mode, the controller releases maximum cooling capacity when the overtemperature trigger value is reached, instead of switching off all poles. This is critical for exothermic processes where loss of cooling could cause runaway temperatures. The power-failure response can be set to OFF/Standby, ON, or Power Failure Auto via the Pilot ONE.
Before putting the unit into operation, the parking brakes must be activated at the casters (if any) and/or the leveling feet must be unscrewed/activated to prevent movement and crushing injuries. Electrical lines directly below the unit must be protected against liquid ingress from condensation.
Selection & Integration Q&A
Choose the 210T when higher cooling capacity is needed: 21 kW at 0 °C and 15 kW at 15 °C, versus the 110T series which is positioned at lower capacity. The 110T variants offer lower noise levels (65–74 dB(A)), so if acoustic limits are tight and capacity demand is lower, a 110T model may be preferable.
The pump connection is G1 1/4 male, so mating hoses and fittings must match this thread. The external loop must be closed and its pressure drop checked against the pump curve (210 l/min at 0 bar to 31 l/min at 4.5 bar). Minimum filling capacity is 14 l and expansion volume is 39 l; ensure the loop volume and fluid selection are compatible.
Install indoors only, at 5–40 °C ambient, with 50–200 cm side clearance for maintenance, at least 300 lx illuminance, and up to 2000 m altitude. IP20 requires a dry, protected location. The unit is not ATEX-rated and must not be used in hazardous zones. Cooling capacity may drop if ambient temperature rises.
The base delivery includes a cable but no plug for the 400 V three-phase supply; the site must provide a suitable plug. Refrigerant circuit work must be performed by approved refrigeration contractors. Regular maintenance includes checking refrigerant charge and room size per m³, and shutting off cooling water (if applicable) during downtimes. No specific maintenance interval is stated in the provided documents.