Peter Huber Kältemaschinenbau Unichiller 230w
Peter Huber Kältemaschinenbau Unichiller 230w — Recirculating chillers.
The Unichiller 230w is a water-cooled recirculating chiller with a Pilot ONE controller, rated 23 kW at 20 °C and 5.5 kW at -20 °C over a -20…40 °C range. Its E1 pressure pump delivers 54 l/min at 3.5 bar, distinguishing it from air-cooled siblings. It serves externally closed process loops requiring stable temperature control.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Pump type | Safety class | Temperature range, °C | Power supply | Expansion volume, m³ | Minimum fill volume, m³ |
|---|---|---|---|---|---|---|
| Unichiller 230w Current | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 110T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 130T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 160T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.01 | — |
| Unichiller 180-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 180Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 180w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180w-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 200T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 200Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 210T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 230-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w-H | E1 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller 250T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 260T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 260Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 300T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 300T-H | D3 | III / FL | -20–100 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 300Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 300Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 350T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 350Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 500Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 700Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 800Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller GL 1000w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.05 | — |
| Unichiller GL 200w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 250w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 300w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 400w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 750w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.05 | — |
| Unichiller P180 | E2 | I / NFL | -20–40 | — | 0.048 | 0.018 |
| Unichiller P180-H | E2 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w | E2 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230 | E2 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
Full specifications
| Expansion volume | 48 l |
| Minimum fill volume | 18 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | E1 |
| External sensor connection | Pt100 |
| Temperature range | -20...40 °C |
| Temperature stability | 0.5 K |
| Dimensions (W x D x H) mfr | 940 x 1290 x 1130 mm |
| Delivery capacity pressure pump max. mfr | 54 l/min; 3,5 bar |
| Pump characteristics mfr | Pressure pump, speed not variable |
| Weight | 339 kg |
| Current consumption | 14 A (max) |
| Power supply | 400V 3~ 50Hz |
| Fuse mfr | 3x16A |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling capacity | 23 kW (at 20 °C) at 0 °C: 13.5 kW · at -10 °C: 9 kW · at -20 °C: 5.5 kW |
| Cooling water connection | G3/4 male |
| Cooling water consumption | 324 l/h (at 15 °C, flow -20 °C) at 15 °C, flow -10 °C: 414 l/h · at 15 °C, flow 0 °C: 540 l/h |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
- Water-cooled refrigeration circuit using R-449A (GHS H280, GWP 1397); cooling water connections G3/4 male, pump connections G1 1/4 AG male.
- E1 pressure pump, speed not variable, delivering 54 l/min at 3.5 bar; 18 l minimum fill, 48 l expansion volume.
- Pilot ONE controller with Ethernet, USB (Host and Device) and RS232; Pt100 external sensor input.
- Cabinet 940 x 1290 x 1130 mm, 339 kg, IP20, safety class I / NFL.
- Sibling models 110T/110T-H/110Tw/110Tw-H are air-cooled variants (noise 74 or 65 dB(A)); the 230w is the water-cooled modification, so it needs a cooling water supply instead of fan-based heat rejection.
Behavior under load and media
- Cooling capacity is strongly temperature-dependent: 23 kW at 20 °C, 13.5 kW at 0 °C, 9 kW at -10 °C, 5.5 kW at -20 °C.
- Cooling water consumption at 15 °C intake: 540 l/h at 0 °C flow, 414 l/h at -10 °C, 324 l/h at -20 °C; higher cooling water temperature reduces cooling power and may require increased flow.
- Temperature stability 0.5 K; ambient operating range 5…40 °C.
- Cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM; only compatible water quality may be used.
Installation, connection and control
- Requires 400 V 3~ 50 Hz supply, max 14 A, fuse 3x16A; three-phase devices below 63 A are supplied with cable but without plug.
- Voltage tolerance ±5% with simultaneous frequency tolerance ±2%, with certain combinations not allowed.
- Cooling water supply: minimum differential pressure 1 bar, maximum pressure 6 bar; intake temperature 5…20 °C; insulate pipes if condensation occurs.
- Must be installed indoors, not in an ATEX zone, at illuminance ≥300 lx and altitude up to 2000 m; wall clearance 50–200 cm per side for maintenance.
- Leveling feet must be extended and parking brakes applied before start-up; electrical lines directly below the unit must be protected against liquid ingress.
- Power-failure response configurable as OFF/Standby, ON, or Power Failure Auto via Pilot ONE; Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached.
Integration & Operating Notes
The chiller is designed for cooling water intake temperatures of 5…20 °C. Higher intake temperatures reduce cooling power and may require increased cooling water flow, so the supply circuit must be sized for the worst-case intake temperature.
Free-standing installation requires a wall clearance of 50 to 200 cm on either side. The unit must be placed indoors at illuminance of at least 300 lx and at altitudes up to 2000 m; installation inside an ATEX zone is prohibited because the unit is not explosion-protected.
The 400 V 3~ 50 Hz unit draws max 14 A and requires a 3x16A fuse. Because current is below 63 A, it is delivered with a cable but without a plug, so the site must provide a suitable connector. Voltage tolerance is ±5% with ±2% frequency tolerance, and certain combinations are not allowed.
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Only cooling water compatible with these materials may be used to avoid corrosion or damage. Pipes must be insulated if condensation occurs.
The cooling water supply must be shut off even for short downtimes such as overnight to prevent water damage and corrosion. Work on the refrigerant circuit may only be performed by approved refrigeration/air-conditioning contractors.
The Pilot ONE allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles.
Selection & Integration Q&A
Choose the 230w when a cooling water supply (1–6 bar differential, 5–20 °C intake) is available and higher capacity is needed: 23 kW at 20 °C and 5.5 kW at -20 °C. The 110T/110Tw are air-cooled and rely on fan heat rejection instead.
Provide cooling water at 5–20 °C intake with minimum differential pressure 1 bar and maximum pressure 6 bar. Connections are G3/4 male. Consumption is 324–540 l/h depending on flow temperature. Use water compatible with copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.
Install indoors only, not in an ATEX zone, at 5–40 °C ambient, illuminance ≥300 lx, altitude ≤2000 m, with 50–200 cm side clearance. Extend leveling feet and apply parking brakes before start-up. Shut off cooling water even for overnight downtime.
The unit ships with cable but no plug for the 400 V 3~ supply; a suitable connector and 3x16A fusing must be site-provided. Refrigerant circuit work requires approved contractors. Cooling water piping and insulation (if condensation occurs) are not included.