Peter Huber Kältemaschinenbau Unichiller 300T
Peter Huber Kältemaschinenbau Unichiller 300T — Recirculating chillers.
The Unichiller 300T is an air-cooled recirculating chiller with a fixed-speed D3 pressure pump and G1 1/4 male connections, rated for -20 to 40 °C. Its cooling capacity falls from 30 kW at 15 °C to 6 kW at -20 °C, so sizing must follow the actual process temperature. It serves externally closed industrial cooling loops on a 400 V three-phase supply.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Pump type | Safety class | Temperature range, °C | Power supply | Expansion volume, m³ | Minimum fill volume, m³ |
|---|---|---|---|---|---|---|
| Unichiller 300T Current | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 110T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 130T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 160T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.01 | — |
| Unichiller 180-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 180Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 180w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180w-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 200T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 200Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 210T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 230-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w-H | E1 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller 250T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 260T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 260Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 300T-H | D3 | III / FL | -20–100 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 300Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 300Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 350T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 350Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 500Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 700Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 800Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller GL 1000w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.05 | — |
| Unichiller GL 200w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 250w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 300w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 400w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 750w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.05 | — |
| Unichiller P180 | E2 | I / NFL | -20–40 | — | 0.048 | 0.018 |
| Unichiller P180-H | E2 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w | E2 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230 | E2 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
Full specifications
| Expansion volume | 39 l |
| Minimum fill volume | 14 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | D3 |
| External sensor connection | Pt100 |
| Temperature range | -20...40 °C |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
The 300T is the largest of the Unichiller T-series models listed here; siblings 110T, 110T-H, 110Tw and 110Tw-H are smaller-capacity units (their published noise figures are 74 dB(A) for 110T/110T-H and 65 dB(A) for the Tw variants). The 300T uses a Pilot ONE controller, a D3 pressure pump with non-variable speed, and G1 1/4 male pump connections. The refrigerant is R-449A (A1, H280) with GWP 1397. The unit is rated IP20 and EMC Class A Group 1, i.e. industrial-environment equipment.
Thermal behavior
- Operating temperature range: -20 to 40 °C.
- Cooling capacity: 30 kW at 15 °C, 18 kW at 0 °C, 11 kW at -10 °C, 6 kW at -20 °C.
- Output data are valid at 20 °C room temperature; capacity may drop as ambient temperature rises.
- Pump: max. delivery 210 l/min, max. delivery pressure 4.7 bar, 202 l/min at 1.1 bar.
- Hydraulic volume: minimum filling capacity 14 l, expansion volume 39 l.
Installation and control
Commissioning requires a 400 V 3~ 50 Hz feed; three-phase devices below 63 A are delivered with cable but without a plug. Supply tolerance is ±5% voltage with simultaneous ±2% frequency, and some combinations are not permitted. The unit is air-cooled and needs the manufacturer's space requirements respected; a wall clearance of 50 to 200 cm per side is required for maintenance on free-standing models. Leveling feet must be activated and parking brakes applied before start-up. The Pilot ONE provides Ethernet, USB (host and device) and RS232 interfaces, a Pt100 external sensor connection, and a configurable power-failure response (OFF/Standby, ON, Power Failure Auto). Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached.
Integration & Operating Notes
Capacity falls steeply as the setpoint drops (30 kW at 15 °C versus 6 kW at -20 °C), so a loop sized on the warm-end figure will be undersized for low-temperature duty. The published data are valid at 20 °C room temperature and may derate in warmer plant rooms.
The 400 V 3~ 50 Hz unit below 63 A arrives with a cable only, so the customer must provide the final connection. Voltage tolerance is ±5% only with simultaneous ±2% frequency tolerance, and some combinations are not allowed, so supply quality should be verified before commissioning.
The IP20 enclosure must be installed indoors, protected from water and dust, at 5 to 40 °C ambient and at least 300 lx illuminance. A wall clearance of 50 to 200 cm per side is required for maintenance on free-standing models, and the manufacturer's space requirements must be respected to avoid reduced cooling performance.
The unit is designed for externally closed applications and must not be installed or started up inside an ATEX zone because it is not explosion-protected. Refrigerant charge and room size must be checked at installation to stay within the refrigerant limit per m³ of ambient air.
Any work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The refrigerant is R-449A with GWP 1397, so handling and greenhouse-gas reporting obligations apply; the rating plate and datasheet give the charge.
Selection & Integration Q&A
Pick the 300T when the required cooling capacity at the actual process temperature exceeds what the 110T family delivers; the 300T provides 30 kW at 15 °C down to 6 kW at -20 °C. The 110Tw variants are the quieter option (65 dB(A) versus 74 dB(A) for 110T/110T-H), so choose a Tw sibling where noise is the deciding constraint.
The pump connection is G1 1/4 male, so hoses and fittings must match that thread and be counterheld at the maximum tightening torque. The D3 pump is not speed-variable: it delivers up to 210 l/min and 4.7 bar, with 202 l/min at 1.1 bar, so the external pressure drop must be checked against that curve. Loop volume must be at least 14 l with 39 l expansion volume.
Three-phase units below 63 A ship with a cable but without a plug, so the final connection is customer-supplied. Refrigerant-circuit work must be done by approved refrigeration contractors; R-449A has GWP 1397, so charge and room size must be checked and reporting obligations apply. Shut off cooling water even overnight on water-cooled installations.