Peter Huber Kältemaschinenbau Unichiller 500Tw
Peter Huber Kältemaschinenbau Unichiller 500Tw — Recirculating chillers.
The Unichiller 500Tw is a water-cooled recirculating chiller with a Pilot ONE controller, covering -20 to 40 °C and requiring 400 V 3~ 50 Hz. Its D3 pump delivers up to 234 l/min at 4.9 bar through G1 1/4 male connections, with cooling water at 5–20 °C and 1.5–6 bar differential. It serves externally closed process loops needing high flow and pressure.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Pump type | Safety class | Temperature range, °C | Power supply | Expansion volume, m³ | Minimum fill volume, m³ |
|---|---|---|---|---|---|---|
| Unichiller 500Tw Current | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 130T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 160T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.01 | — |
| Unichiller 180-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 180Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 180w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180w-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 200T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 200Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 210T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 230-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w-H | E1 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller 250T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 260T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 260Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 300T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 300T-H | D3 | III / FL | -20–100 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 300Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 300Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 350T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 350Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 600Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 700Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 800Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller GL 1000w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.05 | — |
| Unichiller GL 200w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 250w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 300w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 400w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 750w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.05 | — |
| Unichiller P180 | E2 | I / NFL | -20–40 | — | 0.048 | 0.018 |
| Unichiller P180-H | E2 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w | E2 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230 | E2 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
Full specifications
| Expansion volume | 7 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | D3 |
| External sensor connection | Pt100 |
| Temperature range | -20...40 °C |
| Weight | 634 kg |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1 1/4 male |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
- Water-cooled (not air-cooled) recirculating chiller; the 'w' in 500Tw denotes the water-cooled condenser, distinguishing it from air-cooled siblings such as the Unichiller 110T.
- Fixed-speed D3 pump (pump-speed-adjustable: no) with G1 1/4 AG male pump connection and G1 1/4 male cooling-water connection; hose couplings for both are included.
- 7 l expansion volume; minimum filling capacity 12.7 l; 634 kg; IP20; safety class I / NFL.
- Controller: Pilot ONE with Ethernet, USB (Host and Device) and RS232; external Pt100 sensor input.
Behavior under load and media
- Temperature range -20...40 °C; ambient 5...40 °C.
- Pump curve: maximum delivery 234 l/min at 4.9 bar, falling to 40 l/min at 4.5 bar — the external circuit pressure drop must be checked against this curve.
- Cooling capacity is referenced to 15 °C cooling-water inlet; warmer cooling water reduces cooling power and may require higher cooling-water flow.
- Cooling-water circuit materials: copper, stainless steel 1.4401, MS, PA, PPE, PTFE, EPDM — only compatible water may be used.
- Refrigerant R-449A (A1, H280), GWP 1397; charge and room volume must be checked against the refrigerant limit per m³.
Installation, connection and operation
- 400 V 3~ 50 Hz supply; three-phase units below 63 A are delivered with cable but without plug — a suitable plug must be provided on site.
- Voltage tolerance ±5% only with frequency tolerance ±2%; -5% voltage with +2% frequency is not allowed.
- Cooling water: inlet 5–20 °C, differential pressure min. 1.5 bar / max. 6 bar; shut off cooling water even for overnight downtimes.
- Insulate internal cooling-water pipes when water temperature falls below the condensation point; protect electrical lines below the unit against condensate.
- Leveling feet/parking brakes must be set before start-up; wall clearance 50–200 cm per side for maintenance; indoor use only, illuminance ≥300 lx, altitude up to 2000 m.
- Power-failure response configurable via Pilot ONE: OFF/Standby, ON, or Power Failure Auto; Process Safety mode releases maximum cooling at the OT trigger value.
- Not explosion-protected: must not be installed or started in an ATEX zone; intended for externally closed applications only.
Integration & Operating Notes
Verify the site cooling-water supply before installation: below 1.5 bar differential the unit cannot operate correctly, above 6 bar it may be damaged. Inlet temperature above 20 °C reduces cooling capacity and may demand higher flow; below the condensation point the internal cooling-water pipes must be insulated.
Units drawing less than 63 A are shipped with a cable only; the installer must provide a matching plug and verify the 400 V 3~ 50 Hz supply. Voltage tolerance is ±5% only with ±2% frequency tolerance, and the combination of -5% voltage with +2% frequency is explicitly not allowed.
Free-standing installation requires 50 to 200 cm clearance on either side for maintenance access. The unit is for indoor use only at illuminance of at least 300 lx and up to 2000 m altitude; leveling feet or parking brakes must be set before switching on. Electrical lines directly below the water-cooled unit must be protected against condensate ingress.
The unit is designed for externally closed applications. The D3 pump delivers 234 l/min at 4.9 bar but only 40 l/min at 4.5 bar, so the external circuit pressure drop must be matched to the pump curve. Minimum filling capacity is 12.7 l with a 7 l expansion volume; pump connections are G1 1/4 male and must be counterheld at the maximum tightening torque.
The Pilot ONE allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto — the choice determines whether the process restarts automatically after an outage. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger value is reached instead of switching off all poles.
Any work on the refrigerant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The refrigerant is R-449A (A1, H280, GWP 1397); charge and room size must be checked at installation to stay within the refrigerant limit per m³ of ambient air. Shut off cooling water even for overnight downtimes.
The unit is not explosion-protected and must not be installed or started within an ATEX zone. It is classified for electromagnetic disturbance as Class A, Group 1 per EN55011, which may require additional EMC precautions in residential or light-industrial environments. Ambient temperature must stay within 5–40 °C.
Selection & Integration Q&A
Pick the 500Tw when a cooling-water circuit at 5–20 °C and 1.5–6 bar is available and high pump flow (234 l/min at 4.9 bar) is needed; air-cooled siblings such as the 110T reject heat to room air and avoid the cooling-water connection but differ in capacity and noise.
An externally closed loop with at least 12.7 l filling volume, G1 1/4 male mating fittings (hose couplings included), pressure drop matched to the D3 pump curve, cooling water at 5–20 °C and 1.5–6 bar differential, and a 400 V 3~ 50 Hz supply with a site-provided plug.
Indoor use only, 5–40 °C ambient, ≥300 lx illuminance, up to 2000 m altitude, 50–200 cm wall clearance per side, no ATEX zones, leveling feet/parking brakes set before start-up, and cooling water shut off even for overnight downtimes.
The three-phase cable is supplied without a plug, so a matching plug is a site item; refrigerant-circuit service must be performed by approved contractors. Hose couplings for pump and cooling-water connections are included, but external hoses, insulation for sub-dew-point cooling water, and condensate protection for cables below the unit are installer scope.