Peter Huber Kältemaschinenbau Unichiller 600Tw-H
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 600Tw-H

Peter Huber Kältemaschinenbau Unichiller 600Tw-H — Recirculating chillers.

The Unichiller 600Tw-H is a water-cooled recirculating chiller with a 2 kW heater and a D3 pressure pump delivering up to 234 l/min, covering a working range of -20 to 100 °C. It is the high-capacity modification in the Unichiller series, distinguished by its 600-size refrigeration circuit and water-cooled condenser. It serves as a central temperature control unit for externally closed process loops requiring stable cooling and heating.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Pump typeSafety classTemperature range, °CPower supplyExpansion volume, m³Minimum fill volume, m³
Unichiller 600Tw-H Current D3III / FL-20–100400V 3~ 50Hz0.007—
Unichiller 110T C3I / NFL-10–40400V 3~ 50Hz0.007—
Unichiller 110T-H C3III / FL-10–100400V 3~ 50Hz0.007—
Unichiller 110Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 130T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 130T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 130Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 160T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 160T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 160Tw C3I / NFL-20–40400V 3~ 50Hz0.01—
Unichiller 180-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 180T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 180T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 180Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 180Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 180w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 180w-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 200T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 200Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 210T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 210T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 210Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 230-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 230w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 230w-H E1III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller 250T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 250T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 250Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 260T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 260T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 260Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 260Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 300T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 300T-H D3III / FL-20–100460V 3~ 60Hz0.0390.014
Unichiller 300Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 300Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 350T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 350Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 500Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 700Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 800Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller GL 1000w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.05—
Unichiller GL 200w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 250w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 300w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 400w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 750w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.05—
Unichiller P180 E2I / NFL-20–40—0.0480.018
Unichiller P180-H E2III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller P180w E2I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller P180w-H E2III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller P230 E2I / NFL-20–40460V 3~ 60Hz0.0480.018
Unichiller P230-H E2III / FL-20–100460V 3~ 60Hz0.0480.018

Full specifications

Core Performance & Functional Specifications
Expansion volume 7 l
Heating capacity 2 kW
Maximum delivery flow rate 234 l/min (pressure pump max.)
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type D3
External sensor connection Pt100
Temperature range -20...100 °C
Dimensions, Weight & Mechanical Design
Weight 634 kg
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1 1/4 male
Cooling Water Pressure 6 bar (max.)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

The Unichiller 600Tw-H is a free-standing, water-cooled recirculating chiller. Its model code indicates the 600-size refrigeration circuit and the 'Tw-H' configuration, which includes a 2 kW heater and a pressure pump (type D3) with a maximum delivery flow rate of 234 l/min. The pump speed is not adjustable. The unit is designed for externally closed applications only, as specified in the operation manual. Compared to smaller siblings such as the Unichiller 110T and 110Tw, this model provides a larger cooling capacity and higher flow rate, making it suitable for larger process heat loads. The 'w' in the model name denotes water cooling, so the unit requires a cooling water supply and does not reject heat to ambient air.

Working range and media

Control and operation

The unit is equipped with the Pilot ONE controller, providing Ethernet, USB (Host and Device) and RS232 interfaces. An external Pt100 sensor can be connected. The power-failure response is configurable via the Pilot ONE as OFF/Standby, ON, or Power Failure Auto. The 'Process Safety' mode releases maximum cooling capacity when the overtemperature trigger value is reached, instead of switching off all poles. The unit is rated IP20 and safety class III/FL. Electrical supply is 400 V 3~ 50 Hz; voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed. Three-phase devices with current consumption less than 63 A are supplied with a cable but without a plug.

Installation and environment

The unit must be installed indoors at an illuminance of at least 300 lx and up to 2000 m altitude. Ambient temperature must be between 5 and 40 °C. A wall clearance of 50 to 200 cm per side is required for maintenance on free-standing models. Leveling feet must be unscrewed/activated and parking brakes applied before operation. Electrical lines directly below the unit must be protected against liquid ingress. The unit must not be installed in an ATEX zone. Refrigerant charge and room size must be checked to avoid exceeding the refrigerant limit per m³ of ambient air. Minimum filling capacity of the system is 12.7 liters; expansion volume is 7 liters.

Integration & Operating Notes

Cooling & Media
Cooling water supply must meet pressure and temperature window

The chiller requires a cooling water differential pressure of at least 1.5 bar and no more than 6 bar, with inlet temperature between 5 and 20 °C. Operation outside these limits may reduce cooling capacity or require additional measures such as pipe insulation when below the condensation point.

Unichiller 600Tw-H Technical Datasheet, page 1-2
Electrical
Supply voltage and frequency tolerances are tightly coupled

The 400 V 3~ 50 Hz supply must stay within ±5% voltage and ±2% frequency simultaneously; certain combinations are not allowed. Exceeding these limits risks compressor and electronics damage. The unit is delivered with a cable but without a plug for currents below 63 A.

Unichiller 600Tw-H Technical Datasheet, page 2
Mounting
Maintain side clearance and secure the unit before start-up

A wall clearance of 50 to 200 cm per side is required for maintenance access. Leveling feet must be unscrewed and parking brakes applied to prevent movement. Electrical lines below the unit must be protected against condensate drip.

Unichiller Operation Manual, pages 25, 28, 31
Process Limits
Only externally closed loops are permitted

The unit is designed exclusively for externally closed applications. Open baths or systems where the heat transfer fluid contacts ambient air are not allowed, as this can cause evaporation, contamination, or unsafe operation.

Unichiller Operation Manual, page 34
Safety
Process Safety mode overrides overtemperature shutdown

When Process Safety is active, reaching the overtemperature trigger releases maximum cooling capacity instead of switching off all poles. This is critical for exothermic processes where loss of cooling could cause runaway temperatures. The mode must be deliberately configured via the Pilot ONE.

Unichiller Operation Manual, page 18
Maintenance
Refrigerant circuit work requires certified contractors

Any work on the refrigerant circuit must be performed by approved refrigeration/air-conditioning system contractors. The refrigerant R-449A is classified A1, H280 (pressurized gas), and the charge of 6 kg corresponds to 8.4 t CO2 equivalent, relevant for F-gas reporting.

Unichiller Operation Manual, page 17; Datasheet page 1
Operation
Shut off cooling water during even short downtimes

The cooling water supply must be shut off during overnight or other short downtimes to prevent water damage and corrosion inside the unit. This is a mandatory operational step, not a recommendation.

Unichiller Operation Manual, page 28

Selection & Integration Q&A

When should I choose the Unichiller 600Tw-H over the smaller 110T or 110Tw models?

Choose the 600Tw-H for higher heat loads and flow rates up to 234 l/min, and when a 2 kW heater and water cooling are required. The 110T/110Tw models offer lower capacity and are air-cooled (no 'w'), suitable for smaller loads where a cooling water supply is unavailable.

What are the requirements for the cooling water circuit and adjacent piping?

Provide cooling water at 5–20 °C with a differential pressure of 1.5–6 bar. Connections are G1 1/4 male. If water temperature is below the condensation point, insulate internal pipes. Use only water compatible with copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.

What ambient and installation conditions must be met for reliable operation?

Install indoors at 5–40 °C ambient, minimum 300 lx illuminance, up to 2000 m altitude, with 50–200 cm side clearance. Do not install in ATEX zones. Ensure room volume is sufficient for the 6 kg R-449A charge. Secure leveling feet and parking brakes before start-up.

What maintenance and service items are required, and what is not included in the base delivery?

Refrigerant circuit work must be by certified contractors. The unit is delivered with a cable but without a plug. No consumables or accessories are included beyond the base unit; the external Pt100 sensor and interface cables must be ordered separately if needed. Regular cooling water shutoff during downtimes is mandatory.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller 600Tw-H Technical Datasheet
Datasheet ·94,0 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF