Peter Huber Kältemaschinenbau Unichiller 600Tw-H
Peter Huber Kältemaschinenbau Unichiller 600Tw-H — Recirculating chillers.
The Unichiller 600Tw-H is a water-cooled recirculating chiller with a 2 kW heater and a D3 pressure pump delivering up to 234 l/min, covering a working range of -20 to 100 °C. It is the high-capacity modification in the Unichiller series, distinguished by its 600-size refrigeration circuit and water-cooled condenser. It serves as a central temperature control unit for externally closed process loops requiring stable cooling and heating.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Pump type | Safety class | Temperature range, °C | Power supply | Expansion volume, m³ | Minimum fill volume, m³ |
|---|---|---|---|---|---|---|
| Unichiller 600Tw-H Current | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 130T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 160T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.01 | — |
| Unichiller 180-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 180Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 180w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180w-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 200T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 200Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 210T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 230-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w-H | E1 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller 250T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 260T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 260Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 300T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 300T-H | D3 | III / FL | -20–100 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 300Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 300Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 350T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 350Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 500Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 700Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 800Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller GL 1000w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.05 | — |
| Unichiller GL 200w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 250w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 300w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 400w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 750w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.05 | — |
| Unichiller P180 | E2 | I / NFL | -20–40 | — | 0.048 | 0.018 |
| Unichiller P180-H | E2 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w | E2 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230 | E2 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
Full specifications
| Expansion volume | 7 l |
| Heating capacity | 2 kW |
| Maximum delivery flow rate | 234 l/min (pressure pump max.) |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | D3 |
| External sensor connection | Pt100 |
| Temperature range | -20...100 °C |
| Weight | 634 kg |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1 1/4 male |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
The Unichiller 600Tw-H is a free-standing, water-cooled recirculating chiller. Its model code indicates the 600-size refrigeration circuit and the 'Tw-H' configuration, which includes a 2 kW heater and a pressure pump (type D3) with a maximum delivery flow rate of 234 l/min. The pump speed is not adjustable. The unit is designed for externally closed applications only, as specified in the operation manual. Compared to smaller siblings such as the Unichiller 110T and 110Tw, this model provides a larger cooling capacity and higher flow rate, making it suitable for larger process heat loads. The 'w' in the model name denotes water cooling, so the unit requires a cooling water supply and does not reject heat to ambient air.
Working range and media
- Temperature range: -20 to 100 °C.
- Heating capacity: 2 kW.
- Cooling water connection: G1 1/4 male; pump connection: G1 1/4 AG male.
- Cooling water pressure: minimum differential pressure 1.5 bar, maximum 6 bar.
- Cooling water inlet temperature: 5 to 20 °C; if below the condensation point, internal cooling water pipes must be insulated.
- Cooling capacity decreases as cooling water temperature rises, and the required cooling water flow rate may increase.
- Wetted materials in the cooling water circuit include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM; only compatible cooling water may be used.
- Refrigerant: R-449A (A1, H280), charge 6 kg, CO2 equivalent 8.4 t (GWP 1397).
Control and operation
The unit is equipped with the Pilot ONE controller, providing Ethernet, USB (Host and Device) and RS232 interfaces. An external Pt100 sensor can be connected. The power-failure response is configurable via the Pilot ONE as OFF/Standby, ON, or Power Failure Auto. The 'Process Safety' mode releases maximum cooling capacity when the overtemperature trigger value is reached, instead of switching off all poles. The unit is rated IP20 and safety class III/FL. Electrical supply is 400 V 3~ 50 Hz; voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed. Three-phase devices with current consumption less than 63 A are supplied with a cable but without a plug.
Installation and environment
The unit must be installed indoors at an illuminance of at least 300 lx and up to 2000 m altitude. Ambient temperature must be between 5 and 40 °C. A wall clearance of 50 to 200 cm per side is required for maintenance on free-standing models. Leveling feet must be unscrewed/activated and parking brakes applied before operation. Electrical lines directly below the unit must be protected against liquid ingress. The unit must not be installed in an ATEX zone. Refrigerant charge and room size must be checked to avoid exceeding the refrigerant limit per m³ of ambient air. Minimum filling capacity of the system is 12.7 liters; expansion volume is 7 liters.
Integration & Operating Notes
The chiller requires a cooling water differential pressure of at least 1.5 bar and no more than 6 bar, with inlet temperature between 5 and 20 °C. Operation outside these limits may reduce cooling capacity or require additional measures such as pipe insulation when below the condensation point.
The 400 V 3~ 50 Hz supply must stay within ±5% voltage and ±2% frequency simultaneously; certain combinations are not allowed. Exceeding these limits risks compressor and electronics damage. The unit is delivered with a cable but without a plug for currents below 63 A.
A wall clearance of 50 to 200 cm per side is required for maintenance access. Leveling feet must be unscrewed and parking brakes applied to prevent movement. Electrical lines below the unit must be protected against condensate drip.
The unit is designed exclusively for externally closed applications. Open baths or systems where the heat transfer fluid contacts ambient air are not allowed, as this can cause evaporation, contamination, or unsafe operation.
When Process Safety is active, reaching the overtemperature trigger releases maximum cooling capacity instead of switching off all poles. This is critical for exothermic processes where loss of cooling could cause runaway temperatures. The mode must be deliberately configured via the Pilot ONE.
Any work on the refrigerant circuit must be performed by approved refrigeration/air-conditioning system contractors. The refrigerant R-449A is classified A1, H280 (pressurized gas), and the charge of 6 kg corresponds to 8.4 t CO2 equivalent, relevant for F-gas reporting.
The cooling water supply must be shut off during overnight or other short downtimes to prevent water damage and corrosion inside the unit. This is a mandatory operational step, not a recommendation.
Selection & Integration Q&A
Choose the 600Tw-H for higher heat loads and flow rates up to 234 l/min, and when a 2 kW heater and water cooling are required. The 110T/110Tw models offer lower capacity and are air-cooled (no 'w'), suitable for smaller loads where a cooling water supply is unavailable.
Provide cooling water at 5–20 °C with a differential pressure of 1.5–6 bar. Connections are G1 1/4 male. If water temperature is below the condensation point, insulate internal pipes. Use only water compatible with copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.
Install indoors at 5–40 °C ambient, minimum 300 lx illuminance, up to 2000 m altitude, with 50–200 cm side clearance. Do not install in ATEX zones. Ensure room volume is sufficient for the 6 kg R-449A charge. Secure leveling feet and parking brakes before start-up.
Refrigerant circuit work must be by certified contractors. The unit is delivered with a cable but without a plug. No consumables or accessories are included beyond the base unit; the external Pt100 sensor and interface cables must be ordered separately if needed. Regular cooling water shutoff during downtimes is mandatory.