Peter Huber Kältemaschinenbau Unichiller 800Tw
Peter Huber Kältemaschinenbau Unichiller 800Tw — Recirculating chillers.
The Unichiller 800Tw is a water-cooled recirculating chiller for externally closed circuits, operating from -20 to 40 °C with a fixed-speed D3 pump and Pilot ONE controller. It is distinguished within the series by its 3480 l/h cooling-water demand and 400 V three-phase supply, positioning it as a high-capacity central cooling source rather than a compact point-of-use unit.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Pump type | Safety class | Temperature range, °C | Power supply | Expansion volume, m³ | Minimum fill volume, m³ |
|---|---|---|---|---|---|---|
| Unichiller 800Tw Current | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 130T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 160T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.01 | — |
| Unichiller 180-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 180Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 180w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180w-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 200T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 200Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 210T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 230-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w-H | E1 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller 250T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 260T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 260Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 300T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 300T-H | D3 | III / FL | -20–100 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 300Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 300Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 350T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 350Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 500Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 700Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller GL 1000w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.05 | — |
| Unichiller GL 200w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 250w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 300w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 400w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 750w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.05 | — |
| Unichiller P180 | E2 | I / NFL | -20–40 | — | 0.048 | 0.018 |
| Unichiller P180-H | E2 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w | E2 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230 | E2 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
Full specifications
| Expansion volume | 7 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | D3 |
| External sensor connection | Pt100 |
| Temperature range | -20...40 °C |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1 1/4 male |
| Cooling water consumption | 3480 l/h (at 15°C, flow 15°C) |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
- Water-cooled refrigeration circuit using R-449A (A1, H280); no air-cooled condenser, so heat is rejected to the facility cooling-water loop.
- Hydraulic connections: pump side G1 1/4 AG male, cooling-water side G1 1/4 male.
- Fixed-speed pump type D3 (pump speed not adjustable), 7 l expansion volume, minimum filling capacity 12.7 l.
- Sibling models in the series (110T, 110T-H, 110Tw, 110Tw-H) are smaller air/water variants; the 800Tw is the high-capacity water-cooled member, so selection hinges on available cooling-water flow rather than on footprint alone.
Behavior and process limits
- Working temperature range -20...40 °C; published cooling capacity is valid only at 20 °C room temperature, 15 °C cooling-water inlet and 2.5 bar differential pressure.
- Cooling-water inlet must stay between 5 °C and 20 °C; below the condensation point, pipes must be insulated.
- Intended exclusively for externally closed applications; open-loop or unpressurized external circuits are not supported.
- Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles — relevant for exothermic loads.
Installation, control and operation
- Pilot ONE controller with Ethernet, USB (host and device) and RS232; external Pt100 sensor connection.
- Power-failure response configurable as OFF/Standby, ON or Power Failure Auto.
- Three-phase 400 V 50 Hz unit with current above 63 A: supplied without cable and without plug; customer must provide cabling and a disconnect/protection device.
- IP20, indoor dry location only; not explosion-protected, so no ATEX zone installation.
Integration & Operating Notes
Published performance is only valid at 15 °C cooling-water inlet; operation outside 5–20 °C invalidates the data and risks condensation. Insulate pipes wherever the water can drop below the condensation point.
Because the three-phase 400 V draw exceeds 63 A, the customer must supply the feeder cable and a suitable disconnect/protection device. Plan the electrical room and protection coordination accordingly before delivery.
±5% voltage is permitted only together with ±2% frequency; opposite-sign deviations (e.g. -5% voltage with +2% frequency) are not allowed. Check site power quality against this combined envelope, not each parameter separately.
Free-standing installation requires this clearance on either side for service access; the unit itself occupies 1000×1600×1620 mm. Leveling feet must be extended and caster brakes applied before startup to prevent movement.
IP20 protection and the absence of explosion protection rule out washdown areas, outdoors and ATEX zones. The room must provide at least 300 lx illuminance, stay at 5–40 °C ambient, and be checked against the refrigerant charge per m³ limit.
Copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM are wetted, so only compatible cooling water may be used. Aggressive or fouling water will corrode or block the circuit and degrade capacity.
Any work on the refrigerant circuit must be carried out by approved refrigeration/air-conditioning contractors. Plan service contracts accordingly; R-449A is classified A1 with GHS hazard H280, so ventilation and handling rules apply.
The Pilot ONE can be set to OFF/Standby, ON or Power Failure Auto. Choose the mode that matches the process risk: automatic restart may be unsafe for attended processes, while manual restart may be unacceptable for unattended cooling.
Leaving the cooling-water supply open during idle periods risks water damage and corrosion. Include a shut-off valve in the facility connection and make closing it part of the daily shutdown procedure.
Selection & Integration Q&A
Pick the 800Tw when the process needs a high-capacity water-cooled source and the site can supply 3480 l/h at 15 °C with 2.5–6 bar differential. The 110T/110Tw siblings suit lower loads; the 800Tw is not a drop-in for sites without that cooling-water flow.
A G1 1/4 male cooling-water connection delivering 3480 l/h at 15 °C within 2.5–6 bar differential, plus a 400 V 3~ 50 Hz feeder with disconnect/protection — the unit ships without cable and without plug because draw exceeds 63 A.
Indoors only, dry, at least 300 lx, 5–40 °C ambient, never in an ATEX zone (IP20, not explosion-protected). Allow 50–200 cm wall clearance per side for maintenance around the 1000×1600×1620 mm footprint, and check refrigerant charge against room volume.
No power cable and no plug are supplied; the customer provides the feeder and protection device. Refrigerant-circuit work is reserved for approved refrigeration contractors, and the cooling-water supply must be shut off even overnight. Minimum filling capacity is 12.7 l with 7 l expansion volume.