Peter Huber Kältemaschinenbau Unichiller 800Tw
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 800Tw

Peter Huber Kältemaschinenbau Unichiller 800Tw — Recirculating chillers.

The Unichiller 800Tw is a water-cooled recirculating chiller for externally closed circuits, operating from -20 to 40 °C with a fixed-speed D3 pump and Pilot ONE controller. It is distinguished within the series by its 3480 l/h cooling-water demand and 400 V three-phase supply, positioning it as a high-capacity central cooling source rather than a compact point-of-use unit.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Pump typeSafety classTemperature range, °CPower supplyExpansion volume, m³Minimum fill volume, m³
Unichiller 800Tw Current D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 110T C3I / NFL-10–40400V 3~ 50Hz0.007—
Unichiller 110T-H C3III / FL-10–100400V 3~ 50Hz0.007—
Unichiller 110Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 130T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 130T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 130Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 160T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 160T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 160Tw C3I / NFL-20–40400V 3~ 50Hz0.01—
Unichiller 180-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 180T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 180T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 180Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 180Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 180w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 180w-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 200T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 200Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 210T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 210T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 210Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 230-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 230w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 230w-H E1III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller 250T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 250T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 250Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 260T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 260T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 260Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 260Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 300T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 300T-H D3III / FL-20–100460V 3~ 60Hz0.0390.014
Unichiller 300Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 300Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 350T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 350Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 500Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw-H D3III / FL-20–100400V 3~ 50Hz0.007—
Unichiller 700Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller GL 1000w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.05—
Unichiller GL 200w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 250w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 300w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 400w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 750w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.05—
Unichiller P180 E2I / NFL-20–40—0.0480.018
Unichiller P180-H E2III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller P180w E2I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller P180w-H E2III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller P230 E2I / NFL-20–40460V 3~ 60Hz0.0480.018
Unichiller P230-H E2III / FL-20–100460V 3~ 60Hz0.0480.018

Full specifications

Core Performance & Functional Specifications
Expansion volume 7 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type D3
External sensor connection Pt100
Temperature range -20...40 °C
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1 1/4 male
Cooling water consumption 3480 l/h (at 15°C, flow 15°C)
Cooling Water Pressure 6 bar (max.)
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

Behavior and process limits

Installation, control and operation

Integration & Operating Notes

Process Limits
Cooling-water inlet must remain within 5–20 °C

Published performance is only valid at 15 °C cooling-water inlet; operation outside 5–20 °C invalidates the data and risks condensation. Insulate pipes wherever the water can drop below the condensation point.

Unichiller 800Tw Technical Data, page 2
Electrical
Unit arrives without cable and without plug

Because the three-phase 400 V draw exceeds 63 A, the customer must supply the feeder cable and a suitable disconnect/protection device. Plan the electrical room and protection coordination accordingly before delivery.

Unichiller 800Tw Technical Data, page 2
Electrical
Voltage and frequency tolerances are coupled

±5% voltage is permitted only together with ±2% frequency; opposite-sign deviations (e.g. -5% voltage with +2% frequency) are not allowed. Check site power quality against this combined envelope, not each parameter separately.

Unichiller 800Tw Technical Data, page 2
Mounting
Allow 50–200 cm wall clearance per side for maintenance

Free-standing installation requires this clearance on either side for service access; the unit itself occupies 1000×1600×1620 mm. Leveling feet must be extended and caster brakes applied before startup to prevent movement.

Unichiller Operation Manual, page 25 and page 31
Mounting
Indoor, dry, non-ATEX location only

IP20 protection and the absence of explosion protection rule out washdown areas, outdoors and ATEX zones. The room must provide at least 300 lx illuminance, stay at 5–40 °C ambient, and be checked against the refrigerant charge per m³ limit.

Unichiller Operation Manual, page 13 and page 25
Cooling & Media
Cooling-water circuit materials restrict water chemistry

Copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM are wetted, so only compatible cooling water may be used. Aggressive or fouling water will corrode or block the circuit and degrade capacity.

Unichiller 800Tw Technical Data, page 2
Maintenance
Refrigerant work restricted to approved contractors

Any work on the refrigerant circuit must be carried out by approved refrigeration/air-conditioning contractors. Plan service contracts accordingly; R-449A is classified A1 with GHS hazard H280, so ventilation and handling rules apply.

Unichiller Operation Manual, page 17
Control & I/O
Configure power-failure response before connecting the process

The Pilot ONE can be set to OFF/Standby, ON or Power Failure Auto. Choose the mode that matches the process risk: automatic restart may be unsafe for attended processes, while manual restart may be unacceptable for unattended cooling.

Unichiller Operation Manual, page 19
Operation
Shut off cooling water even for overnight downtime

Leaving the cooling-water supply open during idle periods risks water damage and corrosion. Include a shut-off valve in the facility connection and make closing it part of the daily shutdown procedure.

Unichiller Operation Manual, page 28

Selection & Integration Q&A

When should I choose the 800Tw instead of the smaller 110T/110Tw siblings?

Pick the 800Tw when the process needs a high-capacity water-cooled source and the site can supply 3480 l/h at 15 °C with 2.5–6 bar differential. The 110T/110Tw siblings suit lower loads; the 800Tw is not a drop-in for sites without that cooling-water flow.

What must the facility provide for the cooling-water and electrical connections?

A G1 1/4 male cooling-water connection delivering 3480 l/h at 15 °C within 2.5–6 bar differential, plus a 400 V 3~ 50 Hz feeder with disconnect/protection — the unit ships without cable and without plug because draw exceeds 63 A.

Where can the unit be installed and how much space does it need?

Indoors only, dry, at least 300 lx, 5–40 °C ambient, never in an ATEX zone (IP20, not explosion-protected). Allow 50–200 cm wall clearance per side for maintenance around the 1000×1600×1620 mm footprint, and check refrigerant charge against room volume.

What is not included in base delivery and what service regime applies?

No power cable and no plug are supplied; the customer provides the feeder and protection device. Refrigerant-circuit work is reserved for approved refrigeration contractors, and the cooling-water supply must be shut off even overnight. Minimum filling capacity is 12.7 l with 7 l expansion volume.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller 800Tw Technical Data
Datasheet ·148,6 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF