Peter Huber Kältemaschinenbau Unichiller P180w
Peter Huber Kältemaschinenbau Unichiller P180w — Recirculating chillers.
The Unichiller P180w is a water-cooled recirculating chiller with a non-variable E2 pressure pump (130 l/min, 5.7 bar) and Pilot ONE controller, covering -20 to 40 °C. Its 'P' designation targets externally closed loops with high pressure drop, distinguishing it from standard-flow siblings. It serves as a central cooling source for process equipment requiring stable temperature control.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Pump type | Safety class | Temperature range, °C | Power supply | Expansion volume, m³ | Minimum fill volume, m³ |
|---|---|---|---|---|---|---|
| Unichiller P180w Current | E2 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 110T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 130T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 160T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.01 | — |
| Unichiller 180-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 180Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 180w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180w-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 200T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 200Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 210T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 230-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w-H | E1 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller 250T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 260T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 260Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 300T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 300T-H | D3 | III / FL | -20–100 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 300Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 300Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 350T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 350Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 500Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 700Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 800Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller GL 1000w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.05 | — |
| Unichiller GL 200w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 250w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 300w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 400w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 750w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.05 | — |
| Unichiller P180 | E2 | I / NFL | -20–40 | — | 0.048 | 0.018 |
| Unichiller P180-H | E2 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230 | E2 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
Full specifications
| Expansion volume | 48 l |
| Minimum fill volume | 18 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | E2 |
| External sensor connection | Pt100 |
| Temperature range | -20...40 °C |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
- Water-cooled refrigeration circuit using R-449A (A1, H280; 1.49 kg, CO2 equivalent 2.1 t) with copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM wetted materials.
- Non-variable E2 pressure pump: maximum delivery 130 l/min and maximum delivery pressure 5.7 bar; flow decreases as external pressure drop rises.
- Pump connection G1 1/4 male; separate cooling-water connection G3/4 male.
- The 'P' in the model name marks this unit as particularly suitable for applications requiring high pressure drops, unlike the 110T/110Tw siblings whose published data focuses on noise level (74 dB(A) for 110T, 65 dB(A) for 110Tw).
Working range and behavior
- Temperature range -20 to 40 °C; ambient temperature 5 to 40 °C; IP20 protection for dry indoor siting.
- Cooling water intake must be 5 to 20 °C. As cooling-water temperature rises, expect reduced cooling power and possibly increased cooling-water flow rate.
- Minimum filling capacity 18 l and expansion volume 48 l must be considered when sizing the external loop.
- Intended only for externally closed applications; not explosion-protected and must not be installed in an ATEX zone.
Control, installation and operation
- Pilot ONE controller with Ethernet, USB (Host and Device) and RS232 interfaces; external Pt100 sensor connection.
- Power-failure response configurable as OFF/Standby, ON, or Power Failure Auto; Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles.
- Three-phase 400 V 3~ 50 Hz supply; units below 63 A are delivered with cable but without plug, units above 63 A without cable and without plug. Voltage tolerance ±5% only with simultaneous frequency tolerance ±2%.
- Requires 50–200 cm wall clearance per side for maintenance, indoor illuminance at least 300 lx, and installation altitude up to 2000 m.
- Cooling water supply must be shut off even for short downtimes such as overnight; electrical lines directly below the unit must be protected against liquid ingress.
Integration & Operating Notes
The datasheet specifies cooling water intake temperature between 5 °C and 20 °C. Higher cooling-water temperatures reduce cooling capacity and may require increased cooling-water flow rate, so the site utility must be sized for the worst-case summer condition.
The manual requires a wall clearance of 50 to 200 cm on either side of free-standing models for maintenance. Planning the installation footprint without this clearance blocks access to the refrigerant circuit and pump connections.
The Pilot ONE allows power-failure response OFF/Standby, ON, or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles, preventing runaway temperatures.
Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The unit contains 1.49 kg R-449A (GWP 1397, CO2 equivalent 2.1 t), so leak-check and service obligations under refrigerant regulations apply.
The unit is designed for externally closed applications and is not explosion-protected. It must not be installed or started up inside an ATEX zone, and the refrigerant charge versus room volume must be checked to avoid exceeding the limit per m³ of ambient air.
Selection & Integration Q&A
Choose the P180w when the external loop has high pressure drop: the 'P' model is specifically suited to such applications and its E2 pump delivers up to 5.7 bar. The 110T/110Tw siblings are positioned by noise level (74 vs 65 dB(A)); select them when acoustic limits dominate and pressure drop is moderate.
Provide cooling water at 5–20 °C to the G3/4 male connection. Use water compatible with copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Insulate internal cooling-water pipes if water temperature can fall below the condensation point, and protect electrical lines below the unit against liquid ingress.
Install indoors only, at 5–40 °C ambient, illuminance at least 300 lx, altitude up to 2000 m, with 50–200 cm side clearance. Shut off cooling water even for overnight downtime. Activate leveling feet or parking brakes before start-up. The unit is IP20 and not ATEX-rated.
Three-phase units below 63 A ship with cable but no plug; above 63 A without cable or plug. Matching hoses/fittings for G1 1/4 male pump and G3/4 male cooling-water connections are not supplied. Refrigerant circuit service must be performed by approved contractors; plan side clearance for maintenance access.