Peter Huber Kältemaschinenbau Unichiller 180-H
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 180-H

Peter Huber Kältemaschinenbau Unichiller 180-H — Recirculating chillers.

The Unichiller 180-H is a 476 kg, 400 V three-phase recirculating chiller with a -20 to 100 °C working range and 18 kW cooling at 20 °C. It is the high-capacity modification in the Unichiller series, distinguished by its 48 l expansion volume and 2 kW heating. It serves externally closed process loops requiring industrial-grade temperature control.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Pump typeSafety classTemperature range, °CPower supplyExpansion volume, m³Minimum fill volume, m³
Unichiller 180-H Current E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 110T C3I / NFL-10–40400V 3~ 50Hz0.007—
Unichiller 110T-H C3III / FL-10–100400V 3~ 50Hz0.007—
Unichiller 110Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 130T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 130T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 130Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 160T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 160T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 160Tw C3I / NFL-20–40400V 3~ 50Hz0.01—
Unichiller 180T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 180T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 180Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 180Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 180w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 180w-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 200T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 200Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 210T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 210T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 210Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 230-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 230w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 230w-H E1III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller 250T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 250T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 250Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 260T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 260T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 260Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 260Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 300T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 300T-H D3III / FL-20–100460V 3~ 60Hz0.0390.014
Unichiller 300Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 300Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 350T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 350Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 500Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw-H D3III / FL-20–100400V 3~ 50Hz0.007—
Unichiller 700Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 800Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller GL 1000w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.05—
Unichiller GL 200w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 250w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 300w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 400w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 750w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.05—
Unichiller P180 E2I / NFL-20–40—0.0480.018
Unichiller P180-H E2III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller P180w E2I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller P180w-H E2III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller P230 E2I / NFL-20–40460V 3~ 60Hz0.0480.018
Unichiller P230-H E2III / FL-20–100460V 3~ 60Hz0.0480.018

Full specifications

Core Performance & Functional Specifications
Expansion volume 48 l
Heating capacity 2 kW
Minimum fill volume 18 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type E1
External sensor connection Pt100
Temperature range -20...100 °C
Dimensions, Weight & Mechanical Design
Weight 476 kg
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

The Unichiller 180-H is a free-standing, IP20-rated recirculating chiller built for indoor industrial environments. Its overall dimensions of 938×1288×2003 mm and 476 kg weight require verified floor loading and access routes. Within the Unichiller series, this model is the high-capacity variant: sibling models such as the 110T and 110Tw are smaller and differ in noise level (74 dB(A) and 65 dB(A) respectively), while the 180-H is specified for larger cooling loads. The unit uses refrigerant R-449A (1.7 kg, CO2 equivalent 2.4 t) and is classified for Class A, Group 1 EMC disturbances per EN55011, restricting use to industrial environments.

Thermal behavior and process limits

The operating temperature range is -20 to 100 °C, but cooling capacity is strongly temperature-dependent: 18 kW at 20 °C falls to 3.5 kW at -20 °C. Sizing must therefore be based on the coldest required set point. The unit includes a 2 kW heater for precise temperature control above ambient. It is designed exclusively for externally closed applications; open-loop or tank operation is not supported. The hydraulic circuit has a minimum fill volume of 18 l and an expansion volume of 48 l, so external loop volume must be considered to avoid insufficient fill or overflow. The pump connection is G1 1/4 male, and the pump is non-adjustable (type E1).

Control, installation and operation

The Pilot ONE controller provides Ethernet, USB (Host and Device) and RS232 interfaces, plus a Pt100 external sensor connection. Power-failure response is configurable (OFF/Standby, ON, or Power Failure Auto). Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached, rather than switching off all poles. Installation requires a 400 V 3~ 50 Hz supply with 3×16 A fusing; three-phase devices below 63 A are delivered with cable but without a plug. Voltage and frequency tolerances are interdependent (-5% voltage with +2% frequency is not allowed). Ambient temperature must stay within 5–40 °C, and cooling capacity may drop above 20 °C. The unit must not be installed in ATEX zones, requires indoor use with at least 300 lx illuminance, and needs 50–200 cm wall clearance per side for maintenance. Leveling feet or parking brakes must be engaged before operation.

Integration & Operating Notes

Electrical
Supply must match 400 V 3~ 50 Hz with 3×16 A fusing

The unit requires a three-phase 400 V 50 Hz supply protected by 3×16 A fuses. Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide a suitable plug. Voltage and frequency tolerances are interdependent: -5% voltage with +2% frequency is not allowed, so site supply quality must be verified before connection.

Unichiller 180-H Datasheet, page 2
Mounting
Indoor installation only, with 50–200 cm side clearance

The unit is IP20 and must be installed indoors in a dry location with at least 300 lx illuminance and up to 2000 m altitude. A wall clearance of 50 to 200 cm per side is required for maintenance access. Leveling feet must be unscrewed or parking brakes activated before operation to prevent movement. Electrical lines directly below a water-cooled unit must be protected against liquid ingress from condensation.

Unichiller Operation Manual, page 25 and page 28
Process Limits
Externally closed circuits only; minimum fill 18 l

The chiller is designed exclusively for externally closed applications; open-loop or tank operation is not supported. The minimum filling capacity is 18 l and expansion volume is 48 l, so the external loop volume must be sized to stay within these limits. The pump connection is G1 1/4 male, and maximum tightening torques must not be exceeded during hose installation.

Unichiller 180-H Datasheet, page 1; Unichiller Operation Manual, page 34
Control & I/O
Configure power-failure response and Process Safety mode

The Pilot ONE controller allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles. These settings must be configured according to the process risk assessment before commissioning.

Unichiller Operation Manual, page 18 and page 19
Maintenance
Refrigerant work only by approved contractors

Work on the refrigerant circuit must only be performed by approved refrigeration/air-conditioning system contractors. The refrigerant charge is 1.7 kg R-449A with a CO2 equivalent of 2.4 t, which triggers leak-check and F-gas record-keeping obligations. The refrigerant charge and room size must be checked at installation to avoid exceeding the refrigerant limit per m³ of ambient air.

Unichiller Operation Manual, page 17; Unichiller 180-H Datasheet, page 1
Environment
Do not install in ATEX zones; industrial EMC environment only

The unit is not explosion-protected and must not be installed or started up inside an ATEX zone. It emits Class A, Group 1 disturbances per EN55011, so it is intended for industrial environments and may cause interference in residential areas. Ambient temperature must stay within 5–40 °C; cooling capacity may drop above 20 °C.

Unichiller Operation Manual, page 13; Unichiller 180-H Datasheet, page 2

Selection & Integration Q&A

When should I choose the Unichiller 180-H over the smaller 110T or 110Tw siblings?

Choose the 180-H when your process requires up to 18 kW cooling at 20 °C and a -20 to 100 °C range with 2 kW heating. The 110T and 110Tw are smaller-capacity models; select them only if their lower cooling output and different noise levels (74 dB(A) and 65 dB(A)) suit your duty. The 180-H also provides 48 l expansion volume for larger external loops.

What are the requirements for the electrical supply and plug for this chiller?

The unit requires a 400 V 3~ 50 Hz supply with 3×16 A fusing. Because it is a three-phase device below 63 A, it is delivered with a cable but without a plug, so you must provide a suitable plug. Verify that site voltage and frequency tolerances do not combine -5% voltage with +2% frequency, as this is not allowed.

What installation clearances and environmental conditions are mandatory?

Install indoors only, with at least 300 lx illuminance and a wall clearance of 50–200 cm per side for maintenance. Ambient temperature must be 5–40 °C. The unit is IP20 and not ATEX-rated, so it must not be placed in hazardous zones. Engage leveling feet or parking brakes before operation.

What maintenance and service obligations apply, and what is not included in delivery?

Refrigerant circuit work must be performed by approved refrigeration contractors. The 1.7 kg R-449A charge (2.4 t CO2e) requires leak checks and F-gas records. The unit is delivered with a cable but no plug; mating hoses and fittings for the G1 1/4 male pump connection are not included. No consumables or accessories are specified in the base delivery.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller 180-H Datasheet
Datasheet ·18,0 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF