Peter Huber Kältemaschinenbau Unichiller 180Tw
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 180Tw

Peter Huber Kältemaschinenbau Unichiller 180Tw — Recirculating chillers with Noise level 59 dB.

The Unichiller 180Tw is a water-cooled recirculating chiller with a Pilot ONE controller, rated for -20 to 40 °C and 18 kW cooling at 15 °C and 0 °C. It is the higher-capacity, lower-noise (59 dB(A)) modification in the Unichiller T series, distinguished from the 110T/110Tw siblings by its 180-size cooling capacity. It serves externally closed process loops requiring stable temperature control with water-cooled heat rejection.

Noise level
59 dB

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Pump typeSafety classTemperature range, °CPower supplyExpansion volume, m³Minimum fill volume, m³
Unichiller 180Tw Current D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 110T C3I / NFL-10–40400V 3~ 50Hz0.007—
Unichiller 110T-H C3III / FL-10–100400V 3~ 50Hz0.007—
Unichiller 110Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 130T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 130T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 130Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 160T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 160T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 160Tw C3I / NFL-20–40400V 3~ 50Hz0.01—
Unichiller 180-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 180T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 180T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 180Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 180w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 180w-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 200T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 200Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 210T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 210T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 210Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 230-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 230w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 230w-H E1III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller 250T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 250T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 250Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 260T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 260T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 260Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 260Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 300T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 300T-H D3III / FL-20–100460V 3~ 60Hz0.0390.014
Unichiller 300Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 300Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 350T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 350Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 500Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw-H D3III / FL-20–100400V 3~ 50Hz0.007—
Unichiller 700Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 800Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller GL 1000w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.05—
Unichiller GL 200w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 250w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 300w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 400w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 750w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.05—
Unichiller P180 E2I / NFL-20–40—0.0480.018
Unichiller P180-H E2III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller P180w E2I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller P180w-H E2III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller P230 E2I / NFL-20–40460V 3~ 60Hz0.0480.018
Unichiller P230-H E2III / FL-20–100460V 3~ 60Hz0.0480.018

Full specifications

Core Performance & Functional Specifications
Noise level 59 dB(A)
Expansion volume 22 l
Maximum delivery flow rate 210 l/min (pressure pump max; also 4, 7 bar)
Minimum fill volume 15 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type D3
External sensor connection Pt100
Temperature range -20...40 °C
Dimensions, Weight & Mechanical Design
Weight 368 kg
Electrical, Power & Drive Specifications
Power supply 460V 3~ 60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Cooling water consumption 427 l/h (cooling water at 15 °C, flow -20 °C)
cooling water at 15 °C, flow -10 °C: 553 l/h · cooling water at 15 °C, flow 0 °C: 666 l/h
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

Behavior under load

Installation, control and operation

Integration & Operating Notes

Electrical
Supply tolerance is coupled: ±5% V with ±2% Hz only

Voltage tolerance of ±5% is permitted only with a simultaneous frequency tolerance of ±2%, and combinations such as -5% voltage with +2% frequency are not allowed. Verify site power quality against this coupled window before connection, since out-of-window supply can damage the 460 V 3~ 60 Hz unit.

Unichiller 180Tw Technical Datasheet, page 2
Electrical
Provide UL 489 branch protection and a plug or hardwire

External branch circuit protection according to UL 489 is required upstream. The three-phase unit (19.5 A, below 63 A) is supplied with cable but without a plug, so the installer must provide a suitable plug or hardwired connection.

Unichiller 180Tw Technical Datasheet, page 2
Cooling & Media
Cooling water: 1 bar min differential, 6 bar max

The water-cooled unit requires a cooling-water supply with a minimum differential pressure of 1 bar and a maximum pressure of 6 bar. Design the supply to meet the flow demand at the required temperature (up to 666 l/h at 0 °C flow with 15 °C water) without exceeding the pressure limit.

Unichiller 180Tw Technical Datasheet, page 1
Cooling & Media
Cooling water intake 5–20 °C; insulate below dew point

The unit is designed for cooling-water intake between 5 °C and 20 °C. Pipes must be insulated if the water temperature can fall below the condensation point, and electrical lines directly below the unit must be protected against liquid ingress from condensation.

Unichiller 180Tw Technical Datasheet, page 2; Operation Manual, page 28
Mounting
Allow 50–200 cm wall clearance per side

Free-standing installation requires a wall clearance of 50 to 200 cm on either side for maintenance access. The unit must be indoors at an illuminance of at least 300 lx and at altitudes up to 2000 m, and IP20 protection means a dry environment protected from water ingress.

Unichiller Operation Manual, page 25
Mounting
Secure leveling feet and parking brakes before start-up

Before the unit is put into operation, the parking brakes must be activated at the casters (if any) and/or the leveling feet must be unscrewed/activated. This prevents movement and crushing injuries during operation.

Unichiller Operation Manual, page 31
Process Limits
Externally closed circuits only; no ATEX zones

The unit is designed only for externally closed applications and must not be installed or started up inside an ATEX zone because it is not explosion-protected. Check refrigerant charge against room size to avoid exceeding the refrigerant limit per m³ of ambient air.

Unichiller Operation Manual, pages 13, 15, 34
Control & I/O
Configure power-failure response for the process

The Pilot ONE allows the power-failure response to be set as OFF/Standby, ON, or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles.

Unichiller Operation Manual, pages 18–19
Maintenance
Refrigerant work only by approved contractors

Work on the refrigerant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The unit contains R-449A (A1, H280), 2.6 kg, CO2 equivalent 3.6 t, so leak detection and environmental reporting obligations apply.

Unichiller Operation Manual, page 17; Technical Datasheet, page 1
Maintenance
Shut off cooling water even for overnight stops

Shut off the cooling-water supply to the unit even for shorter downtimes such as overnight, to prevent water damage and corrosion during idle periods.

Unichiller Operation Manual, page 28

Selection & Integration Q&A

When should I choose the Unichiller 180Tw over the 110T or 110Tw siblings?

Choose the 180Tw when the process needs the higher 180-size cooling capacity (18 kW at 15 °C and 0 °C, 13 kW at -10 °C, 6 kW at -20 °C) and water-cooled heat rejection at 59 dB(A). The 110T/110Tw are smaller-capacity siblings; the 'w' suffix indicates water cooling, so pick a non-w sibling only if air-cooled heat rejection is acceptable.

What cooling-water supply and piping must be provided for this water-cooled unit?

Provide cooling water at 5–20 °C with a minimum differential pressure of 1 bar and maximum pressure of 6 bar, sized for up to 666 l/h at 0 °C flow (553 l/h at -10 °C, 427 l/h at -20 °C with 15 °C water). Connect via G3/4 male; the process pump connection is G1 1/4 male. Insulate pipes below the dew point and protect electrical lines below the unit from condensate.

What are the installation environment and duty-cycle limits from the manual?

Install indoors only, at 5–40 °C ambient, at least 300 lx illuminance, up to 2000 m altitude, with 50–200 cm wall clearance per side. IP20 requires a dry, water-protected location; no ATEX zones. Activate leveling feet/parking brakes before start-up, and shut off cooling water even for overnight stops.

What maintenance and accessories are required beyond the base delivery?

The unit ships with cable but without a plug, so provide a plug or hardwired connection plus UL 489 branch protection. Refrigerant-circuit work must be by approved refrigeration contractors; R-449A charge is 2.6 kg (3.6 t CO2e). No consumables or cables beyond the supplied cable are included; external Pt100 sensor and interface cables are separate.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller 180Tw Technical Datasheet
Datasheet ·18,5 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF