Peter Huber Kältemaschinenbau Unichiller 180Tw
Peter Huber Kältemaschinenbau Unichiller 180Tw — Recirculating chillers with Noise level 59 dB.
The Unichiller 180Tw is a water-cooled recirculating chiller with a Pilot ONE controller, rated for -20 to 40 °C and 18 kW cooling at 15 °C and 0 °C. It is the higher-capacity, lower-noise (59 dB(A)) modification in the Unichiller T series, distinguished from the 110T/110Tw siblings by its 180-size cooling capacity. It serves externally closed process loops requiring stable temperature control with water-cooled heat rejection.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Pump type | Safety class | Temperature range, °C | Power supply | Expansion volume, m³ | Minimum fill volume, m³ |
|---|---|---|---|---|---|---|
| Unichiller 180Tw Current | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 110T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 130T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 160T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.01 | — |
| Unichiller 180-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 180w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180w-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 200T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 200Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 210T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 230-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w-H | E1 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller 250T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 260T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 260Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 300T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 300T-H | D3 | III / FL | -20–100 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 300Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 300Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 350T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 350Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 500Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 700Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 800Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller GL 1000w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.05 | — |
| Unichiller GL 200w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 250w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 300w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 400w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 750w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.05 | — |
| Unichiller P180 | E2 | I / NFL | -20–40 | — | 0.048 | 0.018 |
| Unichiller P180-H | E2 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w | E2 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230 | E2 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
Full specifications
| Noise level | 59 dB(A) |
| Expansion volume | 22 l |
| Maximum delivery flow rate | 210 l/min (pressure pump max; also 4, 7 bar) |
| Minimum fill volume | 15 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | D3 |
| External sensor connection | Pt100 |
| Temperature range | -20...40 °C |
| Weight | 368 kg |
| Power supply | 460V 3~ 60Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G3/4 male |
| Cooling water consumption | 427 l/h (cooling water at 15 °C, flow -20 °C) cooling water at 15 °C, flow -10 °C: 553 l/h · cooling water at 15 °C, flow 0 °C: 666 l/h |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
- Water-cooled refrigeration circuit using R-449A (2.6 kg, CO2 equivalent 3.6 t), with a copper/stainless steel 1.4401/MS/PA/PPE/PTFE/EPDM cooling-water circuit.
- Pump type D3 with G1 1/4 male process connection and G3/4 male cooling-water connection; pump speed is not adjustable.
- Compared with the Unichiller 110T/110T-H (74 dB(A)) and 110Tw/110Tw-H (65 dB(A)), the 180Tw is the larger-capacity, 59 dB(A) variant; the 'w' suffix denotes water-cooled heat rejection.
- 22 l expansion volume and 15 l minimum fill volume define the hydraulic buffer.
Behavior under load
- Cooling capacity is 18 kW at 15 °C and 0 °C, 13 kW at -10 °C, and 6 kW at -20 °C; capacity falls as the cooling-water temperature rises, and required cooling-water flow may increase.
- Cooling-water consumption at 15 °C cooling water: 553 l/h at -10 °C flow, 666 l/h at 0 °C, 427 l/h at -20 °C.
- Temperature range -20 to 40 °C; ambient 5 to 40 °C.
Installation, control and operation
- Pilot ONE controller with Ethernet, USB (host and device) and RS232; external Pt100 sensor connection.
- Power-failure response configurable as OFF/Standby, ON, or Power Failure Auto; Process Safety mode releases maximum cooling when the overtemperature trigger is reached.
- Intended only for externally closed applications; not explosion-protected (no ATEX zones).
Integration & Operating Notes
Voltage tolerance of ±5% is permitted only with a simultaneous frequency tolerance of ±2%, and combinations such as -5% voltage with +2% frequency are not allowed. Verify site power quality against this coupled window before connection, since out-of-window supply can damage the 460 V 3~ 60 Hz unit.
External branch circuit protection according to UL 489 is required upstream. The three-phase unit (19.5 A, below 63 A) is supplied with cable but without a plug, so the installer must provide a suitable plug or hardwired connection.
The water-cooled unit requires a cooling-water supply with a minimum differential pressure of 1 bar and a maximum pressure of 6 bar. Design the supply to meet the flow demand at the required temperature (up to 666 l/h at 0 °C flow with 15 °C water) without exceeding the pressure limit.
The unit is designed for cooling-water intake between 5 °C and 20 °C. Pipes must be insulated if the water temperature can fall below the condensation point, and electrical lines directly below the unit must be protected against liquid ingress from condensation.
Free-standing installation requires a wall clearance of 50 to 200 cm on either side for maintenance access. The unit must be indoors at an illuminance of at least 300 lx and at altitudes up to 2000 m, and IP20 protection means a dry environment protected from water ingress.
Before the unit is put into operation, the parking brakes must be activated at the casters (if any) and/or the leveling feet must be unscrewed/activated. This prevents movement and crushing injuries during operation.
The unit is designed only for externally closed applications and must not be installed or started up inside an ATEX zone because it is not explosion-protected. Check refrigerant charge against room size to avoid exceeding the refrigerant limit per m³ of ambient air.
The Pilot ONE allows the power-failure response to be set as OFF/Standby, ON, or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles.
Work on the refrigerant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The unit contains R-449A (A1, H280), 2.6 kg, CO2 equivalent 3.6 t, so leak detection and environmental reporting obligations apply.
Shut off the cooling-water supply to the unit even for shorter downtimes such as overnight, to prevent water damage and corrosion during idle periods.
Selection & Integration Q&A
Choose the 180Tw when the process needs the higher 180-size cooling capacity (18 kW at 15 °C and 0 °C, 13 kW at -10 °C, 6 kW at -20 °C) and water-cooled heat rejection at 59 dB(A). The 110T/110Tw are smaller-capacity siblings; the 'w' suffix indicates water cooling, so pick a non-w sibling only if air-cooled heat rejection is acceptable.
Provide cooling water at 5–20 °C with a minimum differential pressure of 1 bar and maximum pressure of 6 bar, sized for up to 666 l/h at 0 °C flow (553 l/h at -10 °C, 427 l/h at -20 °C with 15 °C water). Connect via G3/4 male; the process pump connection is G1 1/4 male. Insulate pipes below the dew point and protect electrical lines below the unit from condensate.
Install indoors only, at 5–40 °C ambient, at least 300 lx illuminance, up to 2000 m altitude, with 50–200 cm wall clearance per side. IP20 requires a dry, water-protected location; no ATEX zones. Activate leveling feet/parking brakes before start-up, and shut off cooling water even for overnight stops.
The unit ships with cable but without a plug, so provide a plug or hardwired connection plus UL 489 branch protection. Refrigerant-circuit work must be by approved refrigeration contractors; R-449A charge is 2.6 kg (3.6 t CO2e). No consumables or cables beyond the supplied cable are included; external Pt100 sensor and interface cables are separate.