Peter Huber Kältemaschinenbau Unichiller 230w-H
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 230w-H

Peter Huber Kältemaschinenbau Unichiller 230w-H — Recirculating chillers.

The Unichiller 230w-H is a water-cooled recirculating chiller with a 48 l expansion volume and 18 l minimum fill, operating from -20 to 100 °C with a 2 kW heater. It is distinguished within the series by its 460 V 3~ 60 Hz supply and G3/4 male cooling-water connection. It serves externally closed process loops requiring precise temperature control via the Pilot ONE controller.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Pump typeSafety classTemperature range, °CPower supplyExpansion volume, m³Minimum fill volume, m³
Unichiller 230w-H Current E1III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller 110T C3I / NFL-10–40400V 3~ 50Hz0.007—
Unichiller 110T-H C3III / FL-10–100400V 3~ 50Hz0.007—
Unichiller 110Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 130T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 130T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 130Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 160T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 160T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 160Tw C3I / NFL-20–40400V 3~ 50Hz0.01—
Unichiller 180-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 180T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 180T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 180Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 180Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 180w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 180w-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 200T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 200Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 210T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 210T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 210Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 230-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 230w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 250T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 250T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 250Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 260T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 260T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 260Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 260Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 300T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 300T-H D3III / FL-20–100460V 3~ 60Hz0.0390.014
Unichiller 300Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 300Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 350T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 350Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 500Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw-H D3III / FL-20–100400V 3~ 50Hz0.007—
Unichiller 700Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 800Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller GL 1000w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.05—
Unichiller GL 200w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 250w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 300w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 400w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 750w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.05—
Unichiller P180 E2I / NFL-20–40—0.0480.018
Unichiller P180-H E2III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller P180w E2I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller P180w-H E2III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller P230 E2I / NFL-20–40460V 3~ 60Hz0.0480.018
Unichiller P230-H E2III / FL-20–100460V 3~ 60Hz0.0480.018

Full specifications

Core Performance & Functional Specifications
Expansion volume 48 l
Heating capacity 2 kW
Minimum fill volume 18 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type E1
External sensor connection Pt100
Temperature range -20...100 °C
Electrical, Power & Drive Specifications
Power supply 460V 3~ 60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Cooling water consumption 324 l/h (cooling water 15°C, flow -20°C)
cooling water 15°C, flow -10°C: 414 l/h · cooling water 15°C, flow 0°C: 540 l/h
Cooling water differential pressure 1 bar
Cooling Water Pressure 6 bar
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and Series Positioning

The Unichiller 230w-H is a water-cooled recirculating chiller in the Unichiller series, using the Pilot ONE controller and an E1 pressure pump with fixed speed. Unlike sibling models such as the Unichiller 110T variants, this model is configured for a 460 V 3~ 60 Hz power supply and has a G3/4 male cooling-water connection. The cooling-water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE, and EPDM, which constrains water quality. The unit is rated IP20 and safety class III/FL, and is intended for indoor installation only.

Behavior Under Load and Media

Cooling-water consumption depends on the flow temperature: 540 l/h at 0 °C, 414 l/h at -10 °C, and 324 l/h at -20 °C, all with cooling water at 15 °C. A minimum cooling-water differential pressure of 1 bar is required, and the maximum cooling-water pressure is 6 bar. The cooling water inlet temperature must be between 5 °C and 20 °C. As cooling water temperature rises, cooling power drops and flow rate may need to increase. The unit is designed for externally closed applications only; models with 'P' in the model name are suited to high pressure drop, but this model does not carry that designation.

Installation, Control, and Operation

The Pilot ONE controller provides Ethernet, USB (Host and Device), and RS232 interfaces, and supports a Pt100 external sensor connection. Power-failure response is configurable as OFF/Standby, ON, or Power Failure Auto. Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached, rather than switching off all poles. The unit requires a three-phase 460 V 60 Hz supply with ±5% voltage tolerance and ±2% frequency tolerance; external branch circuit protection per UL 489 must be provided by the installer. Three-phase devices under 63 A are supplied with cable but no plug. A wall clearance of 50 to 200 cm per side is required for maintenance, and the unit must be installed indoors at an illuminance of at least 300 lx and up to 2000 m altitude. Leveling feet or parking brakes must be engaged before operation. Cooling water supply must be shut off even for short downtimes.

Integration & Operating Notes

Electrical
Site must provide 460 V 3~ 60 Hz with tight tolerances

The chiller requires a three-phase 460 V 60 Hz supply. Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed. Verify site power quality before installation to avoid warranty and operational issues.

Unichiller 230w-H Datasheet, page 2
Electrical
Installer must provide branch circuit protection and plug

External branch circuit protection according to UL 489 is required and must be provided by the installer. Three-phase devices with current consumption less than 63 A are supplied with cable but without a plug, so a suitable plug must be sourced separately.

Unichiller 230w-H Datasheet, page 2
Cooling & Media
Cooling water inlet must stay between 5 °C and 20 °C

The unit is designed to operate with cooling water intake temperature between 5 °C and 20 °C. Pipes must be insulated if below the condensation point. Cooling water pressure must not exceed 6 bar, and a minimum differential pressure of 1 bar is required for proper operation.

Unichiller 230w-H Datasheet, pages 1-2
Mounting
Maintain 50–200 cm wall clearance per side for maintenance

A wall clearance of 50 to 200 cm is required on either side for maintenance on free-standing models. The unit must be installed indoors at an illuminance of at least 300 lx and up to 2000 m altitude. It must not be installed or started up inside an ATEX zone.

Unichiller Operation Manual, pages 13 and 25
Mounting
Engage leveling feet or parking brakes before operation

The parking brakes must be activated at the casters (if any) and/or the leveling feet must be unscrewed/activated before the temperature control unit is put into operation. This prevents movement and crushing injuries during operation.

Unichiller Operation Manual, page 31
Process Limits
Unit is for externally closed applications only

This temperature control unit is designed for externally closed applications. Models with 'P' in the model name are particularly suitable for applications requiring high pressure drops; this model does not carry that designation, so verify the pressure drop of your external circuit.

Unichiller Operation Manual, page 34
Control & I/O
Configure power-failure response to match process restart needs

The power-failure response can be configured as OFF/Standby, ON, or Power Failure Auto via the Pilot ONE. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger value is reached instead of switching off all poles.

Unichiller Operation Manual, pages 18-19
Maintenance
Refrigerant circuit work requires approved contractor

Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The refrigerant R-449A has a GHS hazard H280 (pressurized gas) and a GWP of 1397. Observe the amount of refrigerant and room size to avoid exceeding the refrigerant limit per m³ of ambient air.

Unichiller Operation Manual, pages 15 and 17
Operation
Shut off cooling water supply even for short downtimes

Shut off the cooling water supply to the temperature control unit even for shorter downtimes (e.g., overnight). Electrical lines directly below a water-cooled unit must be protected against liquid ingress from condensation.

Unichiller Operation Manual, page 28

Selection & Integration Q&A

When should I choose the Unichiller 230w-H over the Unichiller 110T or 110Tw siblings?

Choose the 230w-H when you need a 460 V 3~ 60 Hz supply and a water-cooled unit with a G3/4 cooling-water connection. The 110T and 110Tw siblings are air-cooled (noise levels 74 dB(A) and 65 dB(A) respectively) and likely differ in voltage; select them if water cooling is unavailable or lower noise is required.

What are the requirements for the cooling water supply and the process loop connections?

Cooling water must be 5–20 °C inlet, max 6 bar, min 1 bar differential pressure, with consumption up to 540 l/h at 15 °C. The cooling-water connection is G3/4 male. The process loop connects via G1 1/4 AG male pump connections; ensure the external circuit is closed and compatible with the wetted materials (copper, stainless steel 1.4401, MS, PA, PPE, PTFE, EPDM).

What maintenance and accessories are required that are not included in the base delivery?

The base delivery includes a cable but no plug for the 460 V three-phase supply; a suitable plug must be sourced. External branch circuit protection per UL 489 is installer-supplied. Refrigerant circuit work must be by approved contractors. No other accessories or consumables are specified in the provided documents.