Peter Huber Kältemaschinenbau Unichiller 260T-H
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 260T-H

Peter Huber Kältemaschinenbau Unichiller 260T-H — Recirculating chillers.

The Unichiller 260T-H is a recirculating chiller with a 26 kW cooling capacity at 0 °C and a -20…100 °C working range, equipped with a non-variable D3 pressure pump and Pilot ONE controller. It is distinguished within the series by its higher capacity and heating option, intended for externally closed industrial process cooling circuits.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Pump typeSafety classTemperature range, °CPower supplyExpansion volume, m³Minimum fill volume, m³
Unichiller 260T-H Current D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 110T C3I / NFL-10–40400V 3~ 50Hz0.007—
Unichiller 110T-H C3III / FL-10–100400V 3~ 50Hz0.007—
Unichiller 110Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 130T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 130T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 130Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 160T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 160T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 160Tw C3I / NFL-20–40400V 3~ 50Hz0.01—
Unichiller 180-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 180T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 180T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 180Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 180Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 180w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 180w-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 200T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 200Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 210T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 210T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 210Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 230-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 230w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 230w-H E1III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller 250T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 250T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 250Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 260T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 260Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 260Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 300T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 300T-H D3III / FL-20–100460V 3~ 60Hz0.0390.014
Unichiller 300Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 300Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 350T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 350Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 500Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw-H D3III / FL-20–100400V 3~ 50Hz0.007—
Unichiller 700Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 800Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller GL 1000w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.05—
Unichiller GL 200w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 250w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 300w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 400w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 750w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.05—
Unichiller P180 E2I / NFL-20–40—0.0480.018
Unichiller P180-H E2III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller P180w E2I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller P180w-H E2III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller P230 E2I / NFL-20–40460V 3~ 60Hz0.0480.018
Unichiller P230-H E2III / FL-20–100460V 3~ 60Hz0.0480.018

Full specifications

Core Performance & Functional Specifications
Expansion volume 39 l
Heating capacity 2 kW
Minimum fill volume 14 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type D3
External sensor connection Pt100
Temperature range -20...100 °C
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling capacity 15 kW (at 15 °C)
at 0 °C: 26 kW · at -10 °C: 16 kW · at -20 °C: 10 kW
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and positioning

Performance and behavior

Installation, control and operation

Integration & Operating Notes

Mounting
Maintain 50–200 cm side clearance for service access

Free-standing models require a wall clearance of 50 to 200 cm on either side to allow maintenance and ensure adequate airflow. Insufficient clearance can lead to overheating and void warranty.

Operation Manual, Section 5 Installation
Electrical
Supply must meet combined voltage and frequency tolerance

The 400 V 3~ 50 Hz supply must be within ±5% voltage and ±2% frequency simultaneously. Verify site supply before commissioning; out-of-tolerance supply can damage the compressor and electronics.

Technical Datasheet, page 2
Electrical
Three-phase units below 63 A are delivered without plug

The unit is supplied with a cable but no plug. The installer must provide a suitable plug or hard-wired connection according to local electrical codes.

Technical Datasheet, page 2
Process Limits
Only externally closed circuits are permitted

The chiller is designed for externally closed applications. Using an open tank or unsealed circuit can cause overflow, contamination, and pump damage, and is not covered by warranty.

Operation Manual, Section 3.2 Intended Use
Process Limits
Cooling capacity derates with ambient temperature above 20 °C

Published cooling capacities are based on 20 °C room temperature. Higher ambient temperatures reduce capacity, so size the unit for the actual installation room conditions to avoid insufficient cooling.

Technical Datasheet, page 2
Control & I/O
Configure power-failure response to match process risk

The Pilot ONE controller allows selecting OFF/Standby, ON, or Power Failure Auto after a power outage. For exothermic processes, choose a mode that ensures safe restart or controlled shutdown.

Operation Manual, Section 4.5 Power Failure
Safety
Set overtemperature protection according to DIN 12876

The adjustable overtemperature protection must be set to a value that protects the process and equipment. In Process Safety mode, reaching the trigger releases maximum cooling instead of shutting down, which is critical for exothermic reactions.

Operation Manual, Section 4.4 Overtemperature Protection
Maintenance
Refrigerant circuit work requires certified contractors

Any work on the refrigerant circuit must be performed by approved refrigeration/air-conditioning system contractors. The refrigerant R-449A has a GWP of 1397, so leak checks and proper accounting are mandatory.

Operation Manual, Section 6 Maintenance
Environment
Indoor use only with minimum 300 lx illuminance

The unit must be installed indoors in a dry location with illuminance of at least 300 lx and altitude not exceeding 2000 m. It is not explosion-protected and must not be placed in an ATEX zone.

Operation Manual, Section 5 Installation

Selection & Integration Q&A

When should I choose the Unichiller 260T-H over the 110T or 110Tw models?

Choose the 260T-H when you need higher cooling capacity (26 kW at 0 °C) and heating capability (2 kW) for temperature control near ambient. The 110T series offers lower capacity and no heating; the 110Tw is water-cooled and quieter (65 dB(A)), while the 260T-H is air-cooled. Select based on required capacity, heating need, and noise constraints.

What are the requirements for connecting the 260T-H to my process loop?

The pump connection is G1 1/4 male, so use matching hose couplings with appropriate seals. The circuit must be externally closed; open tanks are not allowed. Ensure the pressure drop of your loop is compatible with the pump curve (202 l/min at 1.1 bar to 31 l/min at 4.5 bar). Provide a 400 V 3~ 50 Hz supply with ±5% voltage and ±2% frequency tolerance, and a suitable plug (not included).

What installation and duty-cycle limits apply to the 260T-H?

Install indoors only, with 5–40 °C ambient temperature, at least 300 lx illuminance, and max 2000 m altitude. Maintain 50–200 cm side clearance for maintenance. Engage leveling feet or parking brakes before operation. The unit is IP20 and not ATEX-rated. Cooling capacity derates above 20 °C ambient, so size accordingly.

What maintenance and accessories are required for the 260T-H?

Refrigerant circuit work must be by certified contractors; R-449A has GWP 1397, so leak checks are important. No consumables are included in base delivery; you must provide process hoses and couplings (G1 1/4 male), a power plug, and any external Pt100 sensor if needed. Regular maintenance includes checking refrigerant charge and cleaning the condenser (air-cooled).

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller 260T-H Technical Datasheet
Datasheet ·17,8 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF