Peter Huber Kältemaschinenbau Unichiller 300T-H
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 300T-H

Peter Huber Kältemaschinenbau Unichiller 300T-H — Recirculating chillers.

The Unichiller 300T-H is a free-standing, water-cooled recirculating chiller with a Pilot ONE controller, rated -20 to 100 °C and 6–18 kW cooling capacity across that window. It is the high-capacity modification in the Unichiller T series, distinguished by its 460 V three-phase supply and G1 1/4 pump connection. It serves externally closed process loops requiring stable temperature control from sub-zero to near-boiling setpoints.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Pump typeSafety classTemperature range, °CPower supplyExpansion volume, m³Minimum fill volume, m³
Unichiller 300T-H Current D3III / FL-20–100460V 3~ 60Hz0.0390.014
Unichiller 110T C3I / NFL-10–40400V 3~ 50Hz0.007—
Unichiller 110T-H C3III / FL-10–100400V 3~ 50Hz0.007—
Unichiller 110Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 130T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 130T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 130Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 160T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 160T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 160Tw C3I / NFL-20–40400V 3~ 50Hz0.01—
Unichiller 180-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 180T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 180T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 180Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 180Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 180w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 180w-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 200T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 200Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 210T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 210T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 210Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 230-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 230w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 230w-H E1III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller 250T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 250T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 250Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 260T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 260T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 260Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 260Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 300T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 300Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 300Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 350T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 350Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 500Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw-H D3III / FL-20–100400V 3~ 50Hz0.007—
Unichiller 700Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 800Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller GL 1000w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.05—
Unichiller GL 200w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 250w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 300w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 400w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 750w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.05—
Unichiller P180 E2I / NFL-20–40—0.0480.018
Unichiller P180-H E2III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller P180w E2I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller P180w-H E2III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller P230 E2I / NFL-20–40460V 3~ 60Hz0.0480.018
Unichiller P230-H E2III / FL-20–100460V 3~ 60Hz0.0480.018

Full specifications

Core Performance & Functional Specifications
Expansion volume 39 l
Heating capacity 2 kW
Minimum fill volume 14 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type D3
External sensor connection Pt100
Temperature range -20...100 °C
Electrical, Power & Drive Specifications
Power supply 460V 3~ 60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling capacity 15 kW (at 30 °C)
at 0 °C: 18 kW · at -10 °C: 11 kW · at -20 °C: 6 kW
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

The 300T-H is a free-standing recirculating chiller with overall dimensions of 905 × 2172 × 1900 mm (W × D × H). It uses refrigerant R-449A (A1, H280) and is rated IP20. Within the Unichiller T series, the 300T-H is the larger-capacity, three-phase variant; the sibling 110T/110Tw models are smaller units with different noise profiles. The 300T-H carries a fixed-speed D3 pressure pump with a G1 1/4 male connection and no speed adjustment. The controller is the Pilot ONE, with Ethernet, USB (host and device) and RS232 interfaces and a Pt100 external sensor input.

Thermal behavior

Cooling capacity is strongly setpoint-dependent: 18 kW at 0 °C, 11 kW at -10 °C and 6 kW at -20 °C. Sizing must therefore use the capacity at the actual process setpoint, not the nominal figure. A 2 kW heating capacity is available for warm-side control. The published output data are valid at 20 °C room temperature; higher ambient reduces achievable cooling capacity. The unit is intended for externally closed applications only.

Hydraulics and control

The loop requires a minimum fill of 14 l, and the expansion volume is 39 l, which sets the minimum external circuit volume and expansion allowance. The pump connection is G1 1/4 male; mating hoses and couplings must match this thread and must be counterheld during tightening within the manual's torque limits. The Pilot ONE provides the power-failure response setting (OFF/Standby, ON, or Power Failure Auto) and the Process Safety mode, which releases maximum cooling capacity when the overtemperature trigger is reached rather than switching off all poles.

Integration & Operating Notes

Electrical
Supply must be 460 V three-phase 60 Hz with UL 489 breaker

The unit requires a 460 V 3~ 60 Hz supply. Voltage tolerance is ±5% and frequency ±2%, but the two must not combine in the unfavorable direction. External branch circuit protection per UL 489 is mandatory and must be provided by the installer. Three-phase devices under 63 A are delivered with cable but without a plug, so a suitable plug must be sourced on site.

Unichiller 300T-H Datasheet, page 2
Mounting
Plan 50–200 cm side clearance and indoor siting

Free-standing models require a wall clearance of 50 to 200 cm on either side for maintenance access. The unit may only be used indoors, at an illuminance of at least 300 lx and up to 2000 m altitude. Ambient temperature must remain between 5 and 40 °C. Leveling feet must be activated and parking brakes applied before startup.

Unichiller Operation Manual, page 25 and page 31
Process Limits
Externally closed loops only; size capacity at setpoint

The unit is designed for externally closed applications. Cooling capacity falls from 18 kW at 0 °C to 6 kW at -20 °C, so the loop heat load must be checked against the capacity at the actual operating setpoint. Minimum fill is 14 l and expansion volume is 39 l; the external circuit must accommodate both. Output data are valid at 20 °C room temperature and may drop at higher ambient.

Unichiller Operation Manual, page 34; Unichiller 300T-H Datasheet, page 1
Safety
Configure power-failure response and Process Safety mode

The Pilot ONE allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles. The chosen logic must match the process risk assessment before commissioning.

Unichiller Operation Manual, pages 18–19
Maintenance
Refrigerant circuit work restricted to approved contractors

Work on the refrigerant circuit must only be performed by approved refrigeration/air-conditioning system contractors. Refrigerant R-449A is classified A1 with GHS H280; charge and room size must be checked at installation to avoid exceeding the refrigerant limit per m³ of ambient air. Cooling water supply must be shut off even for short downtimes such as overnight.

Unichiller Operation Manual, pages 15, 17 and 28
Mounting
Not for ATEX zones; protect cables below unit

The unit is not explosion-protected and must not be installed or started up inside an ATEX zone. Electrical lines directly below a water-cooled unit must be protected against liquid ingress from condensation. The unit is EMC Class A Group 1 to EN55011, which may require attention in residential-like environments.

Unichiller Operation Manual, pages 13 and 28; Unichiller 300T-H Datasheet, page 2

Selection & Integration Q&A

When should I choose the 300T-H over the smaller 110T/110Tw siblings?

Choose the 300T-H when the process needs up to 18 kW at 0 °C and a -20 to 100 °C window on a 460 V three-phase supply. The 110T/110Tw siblings are smaller-capacity units; select them only when their lower capacity covers the load and their supply matches the site.

What must the external hydraulic circuit and power supply provide for this unit?

The loop must be externally closed with at least 14 l fill and allowance for 39 l expansion, connected via G1 1/4 male threads with counterheld fittings. Power must be 460 V 3~ 60 Hz within ±5% voltage and ±2% frequency, with a UL 489 breaker and a site-supplied plug.

What mounting and ambient limits apply to the 300T-H installation?

Install indoors only, at 5–40 °C ambient, at least 300 lx illuminance and up to 2000 m altitude, with 50–200 cm side clearance for maintenance. Activate leveling feet and parking brakes before startup. Do not place the unit in an ATEX zone; protect cables below it from condensation.

What service and accessory items are not included in base delivery?

The unit ships with cable but without a plug for three-phase supply under 63 A, so a plug and UL 489 breaker are site-supplied. Refrigerant circuit service must be performed by approved contractors. Cooling water supply must be shut off during downtimes; no external sensor or mating hoses are stated as included.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller 300T-H Datasheet
Datasheet ·17,9 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF