Peter Huber Kältemaschinenbau Unichiller 500Tw
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 500Tw

Peter Huber Kältemaschinenbau Unichiller 500Tw — Recirculating chillers.

The Unichiller 500Tw is a water-cooled recirculating chiller with a Pilot ONE controller, covering -20 to 40 °C and requiring 400 V 3~ 50 Hz. Its D3 pump delivers up to 234 l/min at 4.9 bar through G1 1/4 male connections, with cooling water at 5–20 °C and 1.5–6 bar differential. It serves externally closed process loops needing high flow and pressure.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Pump typeSafety classTemperature range, °CPower supplyExpansion volume, m³Minimum fill volume, m³
Unichiller 500Tw Current D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 110T C3I / NFL-10–40400V 3~ 50Hz0.007—
Unichiller 110T-H C3III / FL-10–100400V 3~ 50Hz0.007—
Unichiller 110Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 130T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 130T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 130Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 160T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 160T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 160Tw C3I / NFL-20–40400V 3~ 50Hz0.01—
Unichiller 180-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 180T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 180T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 180Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 180Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 180w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 180w-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 200T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 200Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 210T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 210T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 210Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 230-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 230w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 230w-H E1III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller 250T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 250T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 250Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 260T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 260T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 260Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 260Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 300T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 300T-H D3III / FL-20–100460V 3~ 60Hz0.0390.014
Unichiller 300Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 300Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 350T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 350Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 600Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw-H D3III / FL-20–100400V 3~ 50Hz0.007—
Unichiller 700Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 800Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller GL 1000w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.05—
Unichiller GL 200w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 250w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 300w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 400w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 750w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.05—
Unichiller P180 E2I / NFL-20–40—0.0480.018
Unichiller P180-H E2III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller P180w E2I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller P180w-H E2III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller P230 E2I / NFL-20–40460V 3~ 60Hz0.0480.018
Unichiller P230-H E2III / FL-20–100460V 3~ 60Hz0.0480.018

Full specifications

Core Performance & Functional Specifications
Expansion volume 7 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type D3
External sensor connection Pt100
Temperature range -20...40 °C
Dimensions, Weight & Mechanical Design
Weight 634 kg
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1 1/4 male
Cooling Water Pressure 6 bar (max.)
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

Behavior under load and media

Installation, connection and operation

Integration & Operating Notes

Process Limits
Cooling water must be 5–20 °C at 1.5–6 bar differential

Verify the site cooling-water supply before installation: below 1.5 bar differential the unit cannot operate correctly, above 6 bar it may be damaged. Inlet temperature above 20 °C reduces cooling capacity and may demand higher flow; below the condensation point the internal cooling-water pipes must be insulated.

Unichiller 500Tw Technical Data, pages 1–2
Electrical
Three-phase supply delivered with cable but no plug

Units drawing less than 63 A are shipped with a cable only; the installer must provide a matching plug and verify the 400 V 3~ 50 Hz supply. Voltage tolerance is ±5% only with ±2% frequency tolerance, and the combination of -5% voltage with +2% frequency is explicitly not allowed.

Unichiller 500Tw Technical Data, page 2
Mounting
Plan 50–200 cm wall clearance per side for service

Free-standing installation requires 50 to 200 cm clearance on either side for maintenance access. The unit is for indoor use only at illuminance of at least 300 lx and up to 2000 m altitude; leveling feet or parking brakes must be set before switching on. Electrical lines directly below the water-cooled unit must be protected against condensate ingress.

Unichiller Operation Manual, pages 25 and 28
Process Limits
Externally closed circuits only; check pump curve against loop

The unit is designed for externally closed applications. The D3 pump delivers 234 l/min at 4.9 bar but only 40 l/min at 4.5 bar, so the external circuit pressure drop must be matched to the pump curve. Minimum filling capacity is 12.7 l with a 7 l expansion volume; pump connections are G1 1/4 male and must be counterheld at the maximum tightening torque.

Unichiller Operation Manual, page 34; Unichiller 500Tw Technical Data, page 1
Control & I/O
Configure power-failure response and Process Safety before use

The Pilot ONE allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto — the choice determines whether the process restarts automatically after an outage. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger value is reached instead of switching off all poles.

Unichiller Operation Manual, pages 18–19
Maintenance
Refrigerant circuit work restricted to approved contractors

Any work on the refrigerant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The refrigerant is R-449A (A1, H280, GWP 1397); charge and room size must be checked at installation to stay within the refrigerant limit per m³ of ambient air. Shut off cooling water even for overnight downtimes.

Unichiller Operation Manual, pages 15, 17 and 28
Environment
Not permitted inside ATEX zones; Class A EMC classification

The unit is not explosion-protected and must not be installed or started within an ATEX zone. It is classified for electromagnetic disturbance as Class A, Group 1 per EN55011, which may require additional EMC precautions in residential or light-industrial environments. Ambient temperature must stay within 5–40 °C.

Unichiller Operation Manual, page 13; Unichiller 500Tw Technical Data, page 2

Selection & Integration Q&A

When should I choose the water-cooled Unichiller 500Tw instead of an air-cooled sibling like the Unichiller 110T?

Pick the 500Tw when a cooling-water circuit at 5–20 °C and 1.5–6 bar is available and high pump flow (234 l/min at 4.9 bar) is needed; air-cooled siblings such as the 110T reject heat to room air and avoid the cooling-water connection but differ in capacity and noise.

What must the external circuit and site supply provide to integrate this chiller?

An externally closed loop with at least 12.7 l filling volume, G1 1/4 male mating fittings (hose couplings included), pressure drop matched to the D3 pump curve, cooling water at 5–20 °C and 1.5–6 bar differential, and a 400 V 3~ 50 Hz supply with a site-provided plug.

What installation and duty-cycle limits does the manual impose on the 500Tw?

Indoor use only, 5–40 °C ambient, ≥300 lx illuminance, up to 2000 m altitude, 50–200 cm wall clearance per side, no ATEX zones, leveling feet/parking brakes set before start-up, and cooling water shut off even for overnight downtimes.

What maintenance and accessory items are not included in the base delivery?

The three-phase cable is supplied without a plug, so a matching plug is a site item; refrigerant-circuit service must be performed by approved contractors. Hose couplings for pump and cooling-water connections are included, but external hoses, insulation for sub-dew-point cooling water, and condensate protection for cables below the unit are installer scope.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller 500Tw Technical Data
Datasheet ·18,2 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF