Peter Huber Kältemaschinenbau Unichiller GL 1000w

Peter Huber Kältemaschinenbau Unichiller GL 1000w — Recirculating chillers.

The Unichiller GL 1000w is a water-cooled recirculating chiller rated at 100 kW cooling at 20 °C, operating from -20 to 40 °C with a fixed-speed pressure pump (320 l/min max, 110 l/min at 4.0 bar). It is the 460 V/60 Hz three-phase variant in the Unichiller series, intended for externally closed process loops requiring high cooling capacity.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Pump typeSafety classTemperature range, °CPower supplyExpansion volume, m³Minimum fill volume, m³
Unichiller GL 1000w Current Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.05—
Unichiller 110T C3I / NFL-10–40400V 3~ 50Hz0.007—
Unichiller 110T-H C3III / FL-10–100400V 3~ 50Hz0.007—
Unichiller 110Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 130T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 130T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 130Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 160T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 160T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 160Tw C3I / NFL-20–40400V 3~ 50Hz0.01—
Unichiller 180-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 180T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 180T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 180Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 180Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 180w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 180w-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 200T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 200Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 210T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 210T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 210Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 230-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 230w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 230w-H E1III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller 250T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 250T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 250Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 260T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 260T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 260Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 260Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 300T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 300T-H D3III / FL-20–100460V 3~ 60Hz0.0390.014
Unichiller 300Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 300Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 350T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 350Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 500Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw-H D3III / FL-20–100400V 3~ 50Hz0.007—
Unichiller 700Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 800Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller GL 200w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 250w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 300w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 400w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 750w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.05—
Unichiller P180 E2I / NFL-20–40—0.0480.018
Unichiller P180-H E2III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller P180w E2I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller P180w-H E2III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller P230 E2I / NFL-20–40460V 3~ 60Hz0.0480.018
Unichiller P230-H E2III / FL-20–100460V 3~ 60Hz0.0480.018

Full specifications

Core Performance & Functional Specifications
Expansion volume 50 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type Pressure pump, speed not variable
External sensor connection Pt100
Temperature range -20...40 °C
Electrical, Power & Drive Specifications
Power supply 460V 3~ 60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1 1/4 male
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

The GL 1000w is a water-cooled (not air-cooled) recirculating chiller in the Unichiller series. Unlike the air-cooled Unichiller 110T/110Tw siblings, it uses a cooling-water circuit with G1 1/4 male connections and requires a facility cooling-water supply of 5–20 °C at minimum 1 bar differential pressure (maximum 6 bar). The pump is a fixed-speed pressure pump (not speed-variable) delivering up to 320 l/min, with 110 l/min available at 4.0 bar. The unit is controlled by the Pilot ONE controller and provides Ethernet, USB (host and device) and RS232 interfaces plus a Pt100 external sensor input. The 50 l expansion volume supports a substantial external loop.

Behavior and process limits

The unit is rated for 100 kW cooling at 20 °C and operates over a -20 to 40 °C temperature range. It is intended exclusively for externally closed applications; open external circuits are not permitted. Overtemperature protection according to DIN 12876 is adjustable, and the controller offers a Process Safety mode that releases maximum cooling capacity when the OT trigger value is reached rather than switching off all poles — relevant for exothermic processes. The power-failure response is configurable (OFF/Standby, ON, or Power Failure Auto) via the Pilot ONE.

Installation and operation

Electrical supply is 460 V three-phase 60 Hz; three-phase devices under 63 A are supplied with cable but without a plug. Voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%, and the combination of -5% voltage with +2% frequency is not allowed. Leakage current exceeds 3.5 mA, so protective earthing and RCD design must account for this; external branch circuit protection per UL 489 is required. The unit must be installed indoors at ≥300 lx and up to 2000 m altitude, outside any ATEX zone, with 50–200 cm wall clearance per side for maintenance. Leveling feet or parking brakes must be secured before startup. Cooling-water pipes must be insulated if water temperature falls below the condensation point, and electrical lines directly below the unit must be protected against liquid ingress. The cooling-water supply must be shut off even for overnight downtimes.

Integration & Operating Notes

Process Limits
Closed-loop process circuit only — no open tanks

The unit is designed exclusively for externally closed applications. Connecting an open external circuit can cause overflow, contamination, or unsafe operation, and voids the intended use. Verify the process loop is sealed and pressure-rated before commissioning.

Operation Manual, Function Description / Datasheet Product Overview
Cooling & Media
Cooling water must stay within 5–20 °C and ≥1 bar Δp

The water-cooled refrigeration circuit requires cooling-water inlet between 5 °C and 20 °C with at least 1 bar differential pressure (maximum 6 bar). Facility supply outside this envelope will degrade or prevent rated cooling. If water temperature falls below the condensation point, internal cooling-water pipes must be insulated.

Datasheet, Technical Specifications / Installation
Electrical
Leakage current >3.5 mA demands dedicated earthing design

Leakage current exceeds 3.5 mA, so the protective earth conductor and any residual current device must be sized for high leakage. External branch circuit protection per UL 489 is required. The unit is supplied with cable but without a plug for three-phase connection under 63 A.

Datasheet, Installation / Safety
Electrical
Do not combine -5% voltage with +2% frequency

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and the specific combination of -5% voltage with +2% frequency is explicitly not allowed. Verify supply quality at the installation point before connection to avoid operating outside the allowed envelope.

Datasheet, Installation
Mounting
Keep 50–200 cm clearance per side and secure the casters

Free-standing installation requires 50 to 200 cm wall clearance on either side for maintenance access. Leveling feet must be unscrewed/activated and parking brakes applied before switching on to prevent movement and crushing injury. Indoor use only, at ≥300 lx and up to 2000 m altitude, outside any ATEX zone.

Operation Manual, Installation / Commissioning
Maintenance
Refrigerant circuit work restricted to approved contractors

The refrigerant is R-744 (A1, H280, GWP 1, UN 2857). Any work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. Check refrigerant charge against room size at installation to avoid exceeding the limit per m³ of ambient air.

Operation Manual, Maintenance / Datasheet Refrigerant Data
Process Limits
Configure power-failure response and OT mode before run

The Pilot ONE allows power-failure response OFF/Standby, ON, or Power Failure Auto, and Process Safety mode releases maximum cooling at the OT trigger instead of switching off. For exothermic processes, select Process Safety and set the adjustable DIN 12876 overtemperature trigger for the application.

Operation Manual, Setup Mode / Safety
Cooling & Media
Shut off cooling water even for overnight downtime

The cooling-water supply must be shut off during shorter downtimes such as overnight to prevent water damage and corrosion. Electrical lines directly below the water-cooled unit must be protected against liquid ingress from condensation.

Operation Manual, Installation / Normal Operation

Selection & Integration Q&A

When should I choose the GL 1000w over the air-cooled Unichiller 110T/110Tw siblings?

Choose the GL 1000w when you need 100 kW cooling at 20 °C and have a facility cooling-water supply of 5–20 °C at ≥1 bar Δp; it is water-cooled and avoids the 65–74 dB(A) fan noise of the 110T/110Tw air-cooled siblings. Pick a sibling when no cooling water is available or when a smaller air-cooled unit suffices.

What pump and piping provisions are needed to connect the process loop?

The fixed-speed pressure pump delivers up to 320 l/min (110 l/min at 4.0 bar) through G1 1/4 male connections; size the external loop pressure drop against this curve. Mating hoses and fittings must be G1 1/4, and the maximum tightening torque must not be exceeded with connections counterheld.

What installation environment and duty limits apply from the manual?

Install indoors only, outside ATEX zones, at ≥300 lx and up to 2000 m altitude, with 5–40 °C ambient and 50–200 cm clearance per side. Secure leveling feet/parking brakes before startup. Shut off cooling water even overnight, and insulate cooling-water pipes if below the condensation point.

What is not included in base delivery and what service access is required?

Three-phase units under 63 A ship with cable but without a plug; the installer must provide a plug or hardwired connection plus UL 489 branch protection. Refrigerant-circuit service requires approved refrigeration contractors. No accessories or consumables are listed in the verified facts beyond the cable.