Peter Huber Kältemaschinenbau Unichiller GL 400w
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller GL 400w

Peter Huber Kältemaschinenbau Unichiller GL 400w — Recirculating chillers.

The Unichiller GL 400w is a water-cooled recirculating chiller with a 30 l expansion volume, a fixed-speed pressure pump, and G1 1/4 male process connections, operating from -20 to 40 °C. It is designed for externally closed applications and requires a cooling water supply of 5–20 °C at 1–6 bar differential pressure. The Pilot ONE controller provides Ethernet, USB, and RS232 interfaces for integration into industrial process cooling systems.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Pump typeSafety classTemperature range, °CPower supplyExpansion volume, m³Minimum fill volume, m³
Unichiller GL 400w Current Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller 110T C3I / NFL-10–40400V 3~ 50Hz0.007—
Unichiller 110T-H C3III / FL-10–100400V 3~ 50Hz0.007—
Unichiller 110Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 130T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 130T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 130Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 160T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 160T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 160Tw C3I / NFL-20–40400V 3~ 50Hz0.01—
Unichiller 180-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 180T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 180T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 180Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 180Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 180w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 180w-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 200T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 200Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 210T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 210T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 210Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 230-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 230w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 230w-H E1III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller 250T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 250T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 250Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 260T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 260T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 260Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 260Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 300T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 300T-H D3III / FL-20–100460V 3~ 60Hz0.0390.014
Unichiller 300Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 300Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 350T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 350Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 500Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw-H D3III / FL-20–100400V 3~ 50Hz0.007—
Unichiller 700Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 800Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller GL 1000w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.05—
Unichiller GL 200w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 250w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 300w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 750w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.05—
Unichiller P180 E2I / NFL-20–40—0.0480.018
Unichiller P180-H E2III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller P180w E2I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller P180w-H E2III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller P230 E2I / NFL-20–40460V 3~ 60Hz0.0480.018
Unichiller P230-H E2III / FL-20–100460V 3~ 60Hz0.0480.018

Full specifications

Core Performance & Functional Specifications
Expansion volume 30 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type Pressure pump, speed not variable
External sensor connection Pt100
Temperature range -20...40 °C
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1 1/4 male
Cooling water differential pressure 1 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and Series Positioning

The Unichiller GL 400w is a water-cooled recirculating chiller within the Unichiller series. Unlike air-cooled siblings such as the Unichiller 110T (74 dB(A)) and 110Tw (65 dB(A)), this model requires an external cooling water circuit connected via G1 1/4 male fittings. The unit uses R-744 (CO2) refrigerant (safety group A1, GHS H280) and is classified under pressure equipment category II. The hydraulic circuit includes a pressure pump with fixed speed (not adjustable) and a 30 l expansion volume. Process connections are G1 1/4 AG male. The controller is the Pilot ONE, offering Ethernet, USB (host and device), and RS232 interfaces, plus a Pt100 external sensor connection. The unit is rated IP20 and must be installed indoors in a dry environment.

Behavior and Operating Limits

The chiller operates over a temperature range of -20 to 40 °C, with an ambient temperature limit of 5–40 °C. Cooling water inlet temperature must be between 5 and 20 °C; below the condensation point, cooling water pipes must be insulated. Cooling power decreases as cooling water temperature increases, and the required cooling water flow rate may rise accordingly. The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE, and EPDM, so only compatible cooling water should be used. The unit is intended for externally closed applications only. Models with “P” in the model name are suited to high pressure drop applications; this model does not carry that designation.

Installation, Control, and Operation

Electrical supply is 400 V 3~ 50 Hz with a voltage tolerance of ±5% and simultaneous frequency tolerance of ±2% (combinations like -5% voltage and +2% frequency are not allowed). Three-phase devices with current consumption below 63 A are supplied with cable but without a plug. The unit must not be installed in an ATEX zone. A wall clearance of 50 to 200 cm per side is required for maintenance on free-standing models. Leveling feet must be activated and parking brakes applied before startup. Electrical lines directly below the unit must be protected against liquid ingress from condensation. The Pilot ONE controller allows configuration of power-failure response (OFF/Standby, ON, or Power Failure Auto) and includes a Process Safety mode that releases maximum cooling capacity when the overtemperature trigger value is reached. Cooling water supply must be shut off even for short downtimes such as overnight. Refrigerant circuit work must only be performed by approved refrigeration/air-conditioning contractors.

Integration & Operating Notes

Process Limits
Cooling water inlet must stay between 5 and 20 °C

Operating outside this range reduces cooling capacity or causes condensation. Below the condensation point, cooling water pipes must be insulated to prevent water damage and electrical hazards.

Datasheet, page 2
Electrical
Voltage and frequency tolerances must not be combined at extremes

The supply must be 400 V 3~ 50 Hz with ±5% voltage tolerance and ±2% frequency tolerance, but combinations such as -5% voltage with +2% frequency are not allowed. Exceeding these limits may damage the compressor or controller.

Datasheet, page 2
Mounting
Maintain 50–200 cm side clearance for maintenance access

Free-standing models require this wall clearance on either side to allow service access to the refrigerant circuit and pump. Insufficient clearance can prevent required maintenance and void warranty.

Operation Manual, page 25
Mounting
Unit must not be installed in an ATEX zone

The Unichiller GL 400w is not explosion-protected. Installing or starting it in a hazardous area risks ignition and is strictly prohibited.

Operation Manual, page 13
Control & I/O
Configure power-failure response to match process restart needs

The Pilot ONE controller can be set to OFF/Standby, ON, or Power Failure Auto. For processes that must resume automatically after a power outage, select Power Failure Auto; for safety-critical exothermic processes, consider Process Safety mode instead.

Operation Manual, page 19
Control & I/O
Process Safety mode releases maximum cooling on OT trip

When the overtemperature protection trigger value is reached, Process Safety mode releases maximum cooling capacity instead of switching off all poles. This is critical for exothermic reactions where loss of cooling could cause runaway temperatures.

Operation Manual, page 18
Maintenance
Refrigerant circuit work requires certified contractors

Only approved refrigeration/air-conditioning system contractors may perform work on the coolant circuit. Unauthorized intervention can release R-744 (CO2) refrigerant and violates pressure equipment regulations.

Operation Manual, page 17
Maintenance
Shut off cooling water during short downtimes

Even overnight shutdowns require closing the cooling water supply to prevent water damage and corrosion in the circuit. This also avoids unnecessary load on the building water system.

Operation Manual, page 28
Electrical
Three-phase units below 63 A are supplied without a plug

The installer must provide a suitable plug and connection point for the 400 V 3~ supply. Plan for a qualified electrician to terminate the cable according to local codes.

Datasheet, page 2
Process Limits
Only externally closed applications are permitted

The unit is designed for externally closed hydraulic circuits. Open-loop or vented applications may cause pump cavitation, fluid loss, or inadequate cooling.

Operation Manual, page 34

Selection & Integration Q&A

When should I choose the Unichiller GL 400w over the air-cooled Unichiller 110T or 110Tw?

Choose the GL 400w when a building cooling water circuit (5–20 °C, 1–6 bar differential) is available and lower noise is not critical. The 110T/110Tw are air-cooled and produce 74/65 dB(A); the GL 400w has no fan noise but requires water connections and a water supply.

What are the requirements for the cooling water circuit connected to this chiller?

The cooling water must be 5–20 °C, with a differential pressure of at least 1 bar and maximum 6 bar at the G1 1/4 male connections. Materials in the circuit include copper, stainless steel 1.4401, MS, PA, PPE, PTFE, and EPDM, so ensure water quality is compatible.

What installation and duty-cycle limits apply to the Unichiller GL 400w?

Install indoors only, at 5–40 °C ambient, with 50–200 cm side clearance for maintenance. The unit is IP20 and not ATEX-rated. Leveling feet must be activated and parking brakes applied before startup. Shut off cooling water even for overnight downtimes.

What maintenance and accessories are required that are not included in the base delivery?

The base unit includes a cable but no plug for the 400 V 3~ supply. Refrigerant circuit work must be performed by approved contractors. Regular maintenance includes checking cooling water quality and inspecting for condensation. No specific maintenance interval is stated in the provided documents.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller GL 400w
Datasheet ·30,3 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF