Peter Huber Kältemaschinenbau Unichiller 110T-H
Peter Huber Kältemaschinenbau Unichiller 110T-H — Recirculating chillers with Noise level 74 dB.
The Unichiller 110T-H is a recirculating chiller with a C3 pump and Pilot ONE controller, rated for -10 to 100 °C in externally closed circuits. Its 2 kW heating and staged cooling (11 kW at 15 °C, 6 kW at 0 °C, 2.7 kW at -10 °C) distinguish it from the quieter 110Tw variants. It serves as a temperature-control source for industrial processes requiring a 400 V three-phase supply.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Pump type | Safety class | Temperature range, °C | Power supply | Expansion volume, m³ | Minimum fill volume, m³ |
|---|---|---|---|---|---|---|
| Unichiller 110T-H Current | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 110Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 130T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 130Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | — | — |
| Unichiller 160T | C3 | I / NFL | -10–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160T-H | C3 | III / FL | -10–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 160Tw | C3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.01 | — |
| Unichiller 180-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 180Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 180Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 180w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 180w-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 200T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 200Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 210T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 210Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 230-H | E1 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w | E1 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller 230w-H | E1 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller 250T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 250Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 260T-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 260Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 260Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 300T | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.039 | 0.014 |
| Unichiller 300T-H | D3 | III / FL | -20–100 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 300Tw | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.022 | 0.015 |
| Unichiller 300Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 350T | D3 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.039 | 0.014 |
| Unichiller 350Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.022 | 0.015 |
| Unichiller 500Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 600Tw-H | D3 | III / FL | -20–100 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 700Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller 800Tw | D3 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.007 | — |
| Unichiller GL 1000w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.05 | — |
| Unichiller GL 200w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 250w | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz | 0.03 | — |
| Unichiller GL 300w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 400w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.03 | — |
| Unichiller GL 750w | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz | 0.05 | — |
| Unichiller P180 | E2 | I / NFL | -20–40 | — | 0.048 | 0.018 |
| Unichiller P180-H | E2 | III / FL | -20–100 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w | E2 | I / NFL | -20–40 | 400V 3~ 50Hz | 0.048 | 0.018 |
| Unichiller P180w-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230 | E2 | I / NFL | -20–40 | 460V 3~ 60Hz | 0.048 | 0.018 |
| Unichiller P230-H | E2 | III / FL | -20–100 | 460V 3~ 60Hz | 0.048 | 0.018 |
Full specifications
| Noise level | 74 dB(A) |
| Expansion volume | 7 l |
| Heating capacity | 2 kW |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | C3 |
| External sensor connection | Pt100 |
| Temperature range | -10...100 °C |
| Weight | 232 kg |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The 110T-H is a free-standing recirculating chiller with an atmospheric open expansion tank (7 l) and a non-adjustable C3 pump delivering up to 96 l/min at 5.6 bar.
- It is the heated variant of the 110T platform: 2 kW heating extends the upper working limit to 100 °C, while the base 110T lacks this heating capacity.
- Sibling models 110Tw and 110Tw-H are the low-noise versions (65 dB(A) vs 74 dB(A) here); the 110T-H trades acoustic comfort for heating capability.
- The 130T shares the 74 dB(A) noise level but is a different capacity class; the 110T-H is selected when the process window spans sub-ambient to 100 °C with moderate cooling demand.
Behavior under load and media
- Cooling capacity is strongly temperature-dependent: 11 kW at 15 °C, 6 kW at 0 °C, and 2.7 kW at -10 °C, all referenced to 20 °C room temperature. Sizing must use the capacity at the actual set point, not the nominal maximum.
- Pump delivery falls with rising system pressure (96 l/min max, 29 l/min at 5.0 bar), so external circuit pressure drop must be checked against the pump curve to avoid insufficient flow.
- The unit is designed for externally closed applications only; open-loop or unsealed circuits are outside its intended use.
- Overtemperature protection per DIN 12876 is adjustable and must be set for the application; Process Safety mode can release maximum cooling instead of tripping all poles when the OT trigger is reached.
Installation, control and operation
- The Pilot ONE controller provides Ethernet, USB (host and device) and RS232 interfaces, and allows configuration of the power-failure response (OFF/Standby, ON, or Power Failure Auto).
- Electrical supply is 400 V 3~ 50 Hz; three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide the mating plug.
- Voltage and frequency tolerances are linked: ±5% voltage only with ±2% frequency, and opposite-sign deviations are not permitted.
- The unit is IP20 and EMC Class A Group 1, intended for indoor industrial use at 5–40 °C ambient and at least 300 lx illuminance, up to 2000 m altitude.
- Pump connections are G1 1/4 male; maximum tightening torques must be observed and connections counterheld during hose installation.
Integration & Operating Notes
Free-standing installation requires a wall clearance of 50 to 200 cm on either side to allow service access to the refrigerant circuit and pump. Planning the footprint without this clearance will obstruct maintenance and may violate the manual's installation requirements.
The datasheet permits ±5% voltage only when frequency deviation is within ±2%, and explicitly forbids opposite-sign deviations (e.g., -5% voltage with +2% frequency). Site power quality must be checked against this combined envelope, not each parameter independently.
Devices below 63 A are delivered with a cable but no plug. The installer must procure and fit a suitable mating plug for the 400 V 3~ 50 Hz supply, and account for this in the electrical commissioning schedule.
The unit is designed exclusively for externally closed applications. Using it with an open tank or unsealed loop will cause evaporation, contamination and possible pump damage, and is outside the manual's intended use.
Cooling capacity falls from 11 kW at 15 °C to 2.7 kW at -10 °C, referenced to 20 °C room temperature. Selecting the unit on its nominal maximum will undersize the system for low-temperature operation, and rising ambient temperature further reduces available capacity.
The cooling water supply must be shut off even for short downtimes such as overnight to prevent water damage and corrosion. This should be included in the site's daily shutdown procedure, not only in long-term maintenance planning.
Any work on the refrigerant circuit must be performed by approved refrigeration/air-conditioning system contractors. The refrigerant is R-449A (GHS H280, GWP 1397), so leak response and handling must follow the corresponding safety rules and cannot be delegated to general maintenance staff.
The Pilot ONE allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. For exothermic processes, the wrong choice can cause either uncontrolled restart or loss of cooling; the response must be selected deliberately during commissioning.
In Process Safety mode, maximum cooling capacity is released when the overtemperature protection trigger value is reached, instead of switching off all poles. This behavior must be matched to the process risk: it protects against runaway exotherms but is not a substitute for a hardwired emergency stop.
Selection & Integration Q&A
Choose the 110T-H when the process needs heating up to 100 °C and a 74 dB(A) noise level is acceptable. The 110Tw-H is the low-noise alternative (65 dB(A)) for noise-sensitive rooms. The base 110T lacks the 2 kW heating, so it cannot cover the upper temperature range.
Provide a 400 V 3~ 50 Hz supply with a customer-supplied plug (cable delivered without plug), and G1 1/4 male mating fittings for the pump connections. The external circuit must be closed and its pressure drop checked against the pump curve (96 l/min max, 29 l/min at 5.0 bar).
Install indoors at 5–40 °C ambient, at least 300 lx illuminance, up to 2000 m altitude, with 50–200 cm side clearance for maintenance. IP20 requires a dry location. The unit is not ATEX-rated and must not be placed in a hazardous zone. Cooling water must be shut off overnight.
Refrigerant circuit work must be done by approved refrigeration contractors; the R-449A charge (GWP 1397) requires leak checks. The atmospheric expansion tank must be kept covered and level-checked. The mating electrical plug is not included, and no interface cables are listed in the delivery.