Peter Huber Kältemaschinenbau Unichiller 260T
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 260T

Peter Huber Kältemaschinenbau Unichiller 260T — Recirculating chillers.

The Unichiller 260T is an air-cooled recirculating chiller with a fixed-speed D3 pressure pump and Pilot ONE controller, rated 10–16 kW cooling between -20 and 0 °C. Within the Unichiller series it is the higher-capacity, non-variable-pump modification, intended for externally closed process loops with moderate pressure drop.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Pump typeSafety classTemperature range, °CPower supplyExpansion volume, m³Minimum fill volume, m³
Unichiller 260T Current D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 110T C3I / NFL-10–40400V 3~ 50Hz0.007—
Unichiller 110T-H C3III / FL-10–100400V 3~ 50Hz0.007—
Unichiller 110Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 130T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 130T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 130Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 160T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 160T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 160Tw C3I / NFL-20–40400V 3~ 50Hz0.01—
Unichiller 180-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 180T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 180T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 180Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 180Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 180w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 180w-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 200T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 200Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 210T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 210T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 210Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 230-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 230w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 230w-H E1III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller 250T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 250T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 250Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 260T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 260Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 260Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 300T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 300T-H D3III / FL-20–100460V 3~ 60Hz0.0390.014
Unichiller 300Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 300Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 350T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 350Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 500Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw-H D3III / FL-20–100400V 3~ 50Hz0.007—
Unichiller 700Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 800Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller GL 1000w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.05—
Unichiller GL 200w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 250w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 300w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 400w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 750w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.05—
Unichiller P180 E2I / NFL-20–40—0.0480.018
Unichiller P180-H E2III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller P180w E2I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller P180w-H E2III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller P230 E2I / NFL-20–40460V 3~ 60Hz0.0480.018
Unichiller P230-H E2III / FL-20–100460V 3~ 60Hz0.0480.018

Full specifications

Core Performance & Functional Specifications
Expansion volume 39 l
Minimum fill volume 14 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type D3
External sensor connection Pt100
Temperature range -20...40 °C
Electrical, Power & Drive Specifications
Power supply 460V 3~ 60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling capacity 15 kW (at 0 °C)
at -10 °C: 16 kW · at -20 °C: 10 kW
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

Behavior under load and media

Installation, control and operation

Integration & Operating Notes

Process Limits
Match loop pressure drop to the fixed pump curve

Because pump speed is not adjustable, the achievable flow is determined solely by the external circuit's pressure drop: 210 l/min at zero head falls to 31 l/min at 4.5 bar. Oversized or restrictive loops will starve flow and degrade temperature stability; loops with high pressure drop should use a 'P' model instead.

Unichiller 260T Technical Data Sheet, Specifications; Operation Manual, Function Description
Electrical
Provide plug and UL 489 branch protection on site

The 460 V 3~ 60 Hz unit is supplied with cable but without a plug for currents under 63 A, and external branch circuit protection per UL 489 is mandatory. The installer must size and provide both from the electrical schematics, and must verify the combined voltage/frequency tolerance rule rather than voltage alone.

Unichiller 260T Technical Data Sheet, Installation notes
Mounting
Reserve 50–200 cm clearance per side and indoor siting

Air-cooled operation requires free airflow; the manual specifies 50–200 cm wall clearance on either side for maintenance, indoor use only at ≥300 lx and up to 2000 m altitude. IP20 offers no water protection, so the location must be dry and free of splashes or condensation.

Unichiller Operation Manual, Installation
Process Limits
Keep the external circuit closed and above 14 l fill

The unit is designed only for externally closed applications; an open loop risks evaporation and loss of control. The 39 l expansion volume with a 14 l minimum fill sets the smallest permissible loop volume, and underfilling or overfilling degrades control or causes overflow.

Unichiller 260T Technical Data Sheet, Specifications; Operation Manual, Function Description
Control & I/O
Choose power-failure response before connecting the process

The Pilot ONE can be set to OFF/Standby, ON or Power Failure Auto after a mains loss. For exothermic processes, Process Safety mode releases maximum cooling at the overtemperature trigger instead of de-energizing all poles; the correct mode must be selected and verified during commissioning.

Unichiller Operation Manual, Normal Operation / Setup Mode
Maintenance
Refrigerant work restricted to approved contractors

The R-449A circuit (A1, H280) may only be serviced by approved refrigeration/air-conditioning contractors. Refrigerant charge and room volume must be checked at installation against the limit per m³ of ambient air, and cooling water supply (where applicable) must be shut off even for overnight downtimes.

Unichiller Operation Manual, Maintenance / Safety

Selection & Integration Q&A

When should I choose the Unichiller 260T rather than a 110T or 110Tw sibling?

Pick the 260T when the process needs 10–16 kW between -20 and 0 °C and a fixed-speed pump is acceptable. The 110T/110Tw are lower-capacity; the 'w' variants are quieter (65 vs 74 dB(A)). For high-pressure-drop loops, a 'P' model is the correct choice instead.

What must I provide on the hydraulic and electrical side to connect this chiller to my process?

Mating hoses/fittings must match the G1 1/4 male pump connections, counterheld at the maximum tightening torque. Electrically, the 460 V 3~ 60 Hz unit needs a site-supplied plug (under 63 A) and UL 489 branch protection; verify the combined voltage/frequency tolerance.

What siting and duty-cycle limits apply to the 260T installation?

Indoors only, ambient 5–40 °C, ≥300 lx, up to 2000 m altitude, with 50–200 cm clearance per side for airflow and maintenance. IP20 requires a dry location; leveling feet or parking brakes must be set before start-up, and the unit must not be placed in an ATEX zone.

What is not included in base delivery and what service access does the unit require?

The plug and branch protection are not included, and no accessories or cables are listed in the verified facts. Refrigerant-circuit service requires an approved contractor; the 50–200 cm side clearance is the maintenance access space, and the Pilot ONE interfaces (Ethernet, USB, RS232, Pt100) are built in.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller 260T Technical Data Sheet
Datasheet ·17,8 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF