Induztr
Engineering Digest Document Viewer Related Models

Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®

Technical Reference Summary

Covered Models & Document Scope

Covers table-top models Unistat tango®, Unistat® 40x, Unistat® 705 and free-standing models Unistat® 4xx, 5xx, 6xx, 8xx, 9xx, 10x5. Model-name abbreviations: no suffix = air cooled, GL = refrigerant CO2, P = for applications with high pressure drop, w = water cooled, wl = water/air cooled. Controller is 'Pilot ONE' (standard) or 'Unistat Control ONE' (model-dependent). This documentation does not contain a device-specific technical appendix; full operating instructions with datasheet must be requested from Huber with model designation and serial number.

Section-by-Section Technical Breakdown

pp. 23-32
Commissioning: transport, positioning, ambient and installation conditions

Describes in-plant transport with or without lifting eyes, mounting/removing leveling feet, positioning units with or without casters, and the three transportation lock types (A, B, C) with their activation/deactivation procedures. It also covers unpacking, ambient conditions, EMC-specific notes, installation conditions, recommended temperature control and cooling water hoses, wrench sizes and torques, water-cooled units, and outdoor installation including winter operation.

  • Transportation locks Type A, Type B and Type C must be deactivated for operation and activated for transportation.
  • Leveling feet must be activated before operation and deactivated before transport.
  • Separate procedures for units with casters and without casters.
  • Dedicated section for temperature control units with water cooling and for outdoor installation incl. winter operation.
pp. 33-39
Preparations for operation and application/power connection

Covers activating leveling feet, opening/closing valves, checking knurled screws, locking/unlocking the expansion vessel (isolating sleeve, isolating kit, shut-off valves), checking the emergency stop switch switching status, connecting the exhaust air and functional ground. It also describes connecting externally closed and externally open (bath) applications and connecting to the power supply, including converting the power supply connection.

  • Expansion vessel [18] can be locked/unlocked via isolating sleeve, isolating kit or shut-off valves.
  • Emergency stop switch switching status must be checked before operation.
  • Exhaust air connection [105] and functional ground must be connected.
  • Power supply connection can be converted (Section 2.12.1).
pp. 40-52
Function description, heat transfer fluids and Pilot ONE controller

Explains general and other functions of the temperature control unit, information on heat transfer fluids, and what to note when planning the test. Describes the 'Pilot ONE®' controller functional overview, the Unistat® LED flow diagram, the clock/event function (rechargeable accumulator, alarm clock event, program event), touch-screen operation, display instruments (touchscreen [88], LED indicators temperature [90], flow chart [91], status [92]) and control instruments (touchbuttons, categories, sub-categories, dialogs).

  • LED flow diagram shows components and explains the LEDs.
  • Event functions: 'Alarm clock event' and 'Program event'.
  • Display instruments: touchscreen [88], LED-Indicator Temperature [90], LED-Indicator Flow Chart [91], LED-Indicator Status [92].
  • Function examples include displaying software version, Start & Stop, copying settings to/from USB flash drive, and restoring factory settings.
pp. 53-77
Setup mode: overtemperature protection, temperature control circuit and filling/venting

Describes activating/deactivating the emergency stop switch, turning the unit on/off, and setting the overtemperature protection (OT limit heating, OT expansion vessel, Process Safety, Display OT values) plus testing it for functionality. Covers the temperature control circuit (internal vs. process temperature, Delta T limiter, Pt100 monitoring, auto/expert mode, auto and manual configuration, control structure, reset parameters, freeze protection, setpoint thresholds and setpoint). Also covers filling, venting, degassing and draining for externally closed and externally open applications.

  • Overtemperature protection settings: 'OT limit: heating', 'OT expansion vessel', 'Process Safety'.
  • Delta T limiter can be adjusted and changed.
  • Auto configuration includes 'Find parameters', 'Control Dynamics', 'Fluid Properties'.
  • Manual configuration includes 'Change parameters', 'Display parameters', 'Control structure'.
  • Separate filling/venting/degassing/draining procedures for externally closed and externally open applications.
pp. 78-79
Normal operation: automatic operation and temperature control programs

Describes automatic operation, starting and ending the temperature control process, and temperature control via a created temperature control program including starting and ending/cancelling the program.

  • Starting and ending the temperature control process.
  • Starting and ending/cancelling a temperature control program.
pp. 80-81
Interfaces and software update

Covers removing the interface cover at the 'Unistat® Control ONE', swinging open the interface box [133], installing the Com.G@te® [46] and performing a firmware update.

  • Interface cover removal at 'Unistat® Control ONE'.
  • Interface box [133] swings open.
  • Com.G@te® [46] installation.
  • Firmware update procedure.
pp. 82-92
Maintenance, messages and component replacement

Covers messages from the temperature control unit, replacing the 'Pilot ONE®' or 'Unistat® Control ONE', function check and visual inspection, replacing temperature control or coolant hoses, cleaning liquefier fins, cleaning the hat-type strainer/dirt trap (including draining the cooling circuit and dismantling/installing the cooling water supply), emptying the drip tray, function test of the emergency stop switch, heat transfer fluid inspection/replacement/circuit rinsing, cleaning surfaces, inspecting the mechanical seal, plug contacts and decontamination before shipping.

  • Replacing 'Pilot ONE®' and 'Unistat® Control ONE' controllers.
  • Cleaning the hat-type strainer/dirt trap requires draining the cooling circuit and dismantling the cooling water supply.
  • Heat transfer fluid inspection, replacement and circuit rinsing.
  • Function test of the emergency stop switch.
  • Decontamination before shipping.
pp. 93-97
Shutting down, draining and shipping

Describes safety instructions and basic principles for shutdown, switch-off, draining the temperature control unit, draining the cooling water, uninstalling an external application, opening/closing valves, disassembling the exhaust air [105], deactivating leveling feet, mounting knurled screws, transportation lock, packing, shipping, disposal and contact data, plus the clearance certificate.

  • Draining the temperature control unit and the cooling water.
  • Uninstalling an external application.
  • Deactivating leveling feet and mounting knurled screws for transport.
  • Transportation lock, packing and shipping.
  • Clearance certificate.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
The temperature control unit must NOT be installed or started up within an ATEX zone; without an Ex px cabinet the unit/accessory is not protected against explosion. p.15 ↗ Site selection must exclude ATEX zones; Ex px cabinet operation requires the separate ATEX annex and only the model/serial number stated on the Ex px cabinet rating plate is suitable. whole series
The cooling circuit is permanently technically tight and was tested with an indicator device with a sensitivity from 10-4 Pa m3/s or better. p.17 ↗ Leak-tightness is factory-verified to this sensitivity; refrigerant handling and maintenance still require approved refrigeration/air-conditioning contractors. whole series
Refrigerant transport routes depend on UN ID and filling quantity: UN 3358 ≥0.1 kg to ≤11.99 kg — air freight not possible, road and sea possible; UN 3358 ≥12 kg — possible only as hazardous goods; UN 3358 SP A103 ≤99.9 g — air, road and sea possible; UN 2857 ≤11.99 kg — air, road and sea possible; UN 2857 ≥12 kg — air freight not possible, road/sea possible as hazardous goods. p.18 ↗ Shipping method must be selected from the refrigerant type and factory filling quantity; for air freight of UN 3358 ≥0.1 kg the unit must be filled with nitrogen and refrigerant filling on site is only permitted by Huber-certified partners. whole series
For FL-rated (flammable) heat transfer fluid, the fluid must have a fire point of ≥ 65 °C; if ethanol is used, only supervised operation is permitted. p.19 ↗ Fluid selection and operating supervision must respect the DIN 12876 FL rating and the 65 °C fire-point limit. whole series
Low level protection (independent version) operates via a pressure sensor in the heat transfer fluid circuit; air in the system (fill level too low, inadequately vented) prevents the pressure from reaching the specified value and interrupts temperature control and circulation. p.20 ↗ Thorough venting and correct fill level are mandatory; otherwise the unit will stop circulation. models with independent low level protection
In 'Process safety' mode, when the set overtemperature trigger value is reached the controller releases maximum cooling capacity instead of switching off all poles of the temperature control. p.20 ↗ Process safety keeps heat extraction available under exothermic conditions, reducing risk from glass breakage or uncontrolled phase transitions. whole series
Emergency strategy to interrupt the power grid connection depends on the installed switch combination: >Mains isolator< [36] (red-yellow or grey) turned to '0'; >Mains isolator< [36] plus >Appliance switch< [37] both to '0'; >Emergency stop switch< [70] pressed then >Mains isolator< [36] to '0'; >Power switch< [37] with plug pulled or building circuit breaker used, then switch to '0'. p.21 ↗ The lock-out/tag-out procedure must be written for the specific switch configuration shown in the connection sketch in the annex. whole series
After a power outage or when switching on, the response can be set via 'Pilot ONE': 'Off/Standby' (default, temperature control started only by manual input) or 'ON/Temperature control active' (temperature control always started after turning on, with an INFO shown for a few seconds allowing automatic startup to be disabled). p.21 ↗ Choose the power-interruption behavior according to whether unattended restart is acceptable for the process. whole series
The >Mains isolator< [36] (if present) can be locked in the off position to facilitate safe system installation and maintenance; a lock-out/tag-out procedure must be developed per local regulations (e.g. CFR 1910.147 for the US). p.17 ↗ The responsible body must document an energy-isolation procedure for maintenance. units with mains isolator
The responsible body must install a condensation water / heat transfer fluid drip tray below the temperature control unit; national legislation may prescribe it. p.17 ↗ Installation planning must include a drip tray and verification of local regulations. whole series
The temperature control units Unistat tango (w/wl), Unistat 405(w), Unistat 410(w) and Unistat 705(w) do not have a plate heat exchanger. p.16 ↗ The warning against very dynamic, high temperature changes recurring at short intervals (premature aging/fatigue of the plate heat exchanger) does not apply to these models in the same way. Unistat tango (w/wl), Unistat 405(w), Unistat 410(w), Unistat 705(w)
Use with medical devices (e.g. in vitro diagnostic procedure) or for direct foodstuff temperature control is NOT permissible. p.16 ↗ Application selection must exclude medical-device and direct-food use. whole series
The optional gas detection sensor can be mounted on an integrated mounting plate (stand-alone units only); assembly, electrical connection and function test are the responsibility of the operating company. p.18 ↗ Refrigerant-monitoring installation and commissioning must be planned and executed by the operating company. stand-alone units
The optional processing unit provides a potential-free switching contact to control a power disconnect relay and simultaneously supplies and analyzes the gas detection sensor; the operating company is responsible for dimensioning and on-site installation. p.18 ↗ Alarm integration into an alarm control unit must be engineered by the operating company. stand-alone units
Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. p.19 ↗ Refrigerant-circuit maintenance cannot be performed by general maintenance personnel. whole series
Operators must be at least 18 years old; persons under 18 may operate the unit only under supervision of a qualified specialist. p.19 ↗ Operating instructions and staffing must enforce the age/supervision requirement. whole series

Related models

Peter Huber Kältemaschinenbau Unistat 1005w Dynamisches Peter Huber Kältemaschinenbau Unistat 1015w Dynamisches Peter Huber Kältemaschinenbau Unistat 410w Dynamisches Peter Huber Kältemaschinenbau Unistat 425w Dynamisches Peter Huber Kältemaschinenbau Unistat 430w Dynamisches Peter Huber Kältemaschinenbau Unistat 510w Dynamisches Peter Huber Kältemaschinenbau Unistat 515w Dynamisches Peter Huber Kältemaschinenbau Unistat 520w Dynamisches Peter Huber Kältemaschinenbau Unistat 525 Dynamisches Peter Huber Kältemaschinenbau Unistat 525w Dynamisches Peter Huber Kältemaschinenbau Unistat 527w Dynamisches Peter Huber Kältemaschinenbau Unistat 530w Dynamisches Peter Huber Kältemaschinenbau Unistat 540w Dynamisches Peter Huber Kältemaschinenbau Unistat 545w Dynamisches Peter Huber Kältemaschinenbau Unistat 610w Dynamisches Peter Huber Kältemaschinenbau Unistat 615w Dynamisches Peter Huber Kältemaschinenbau Unistat 620w Dynamisches Peter Huber Kältemaschinenbau Unistat 625w Dynamisches Peter Huber Kältemaschinenbau Unistat 630w Dynamisches Peter Huber Kältemaschinenbau Unistat 635w Dynamisches Peter Huber Kältemaschinenbau Unistat 640w Dynamisches Peter Huber Kältemaschinenbau Unistat 645w Dynamisches Peter Huber Kältemaschinenbau Unistat 650w Dynamisches Peter Huber Kältemaschinenbau Unistat 680w Dynamisches Peter Huber Kältemaschinenbau Unistat 705w Dynamisches Peter Huber Kältemaschinenbau Unistat 815w Dynamisches Peter Huber Kältemaschinenbau Unistat 825w Dynamisches Peter Huber Kältemaschinenbau Unistat 905w Dynamisches Peter Huber Kältemaschinenbau Unistat 912w Dynamisches Peter Huber Kältemaschinenbau Unistat 915w Dynamisches Peter Huber Kältemaschinenbau Unistat 920w Dynamisches Peter Huber Kältemaschinenbau Unistat 930w Dynamisches Peter Huber Kältemaschinenbau Unistat 950w Dynamisches Peter Huber Kältemaschinenbau Unistat GL 530w Dynamisches Peter Huber Kältemaschinenbau Unistat GL 535w Dynamisches Peter Huber Kältemaschinenbau Unistat GL 540w Dynamisches Peter Huber Kältemaschinenbau Unistat GL 550w Dynamisches Peter Huber Kältemaschinenbau Unistat GL 560w Dynamisches Peter Huber Kältemaschinenbau Unistat GL 570w Dynamisches Peter Huber Kältemaschinenbau Unistat GL P530w Dynamisches Peter Huber Kältemaschinenbau Unistat GL P535w Dynamisches Peter Huber Kältemaschinenbau Unistat GL P540w Dynamisches Peter Huber Kältemaschinenbau Unistat GL P550w Dynamisches Peter Huber Kältemaschinenbau Unistat GL P560w Dynamisches Peter Huber Kältemaschinenbau Unistat GL P570w Dynamisches Peter Huber Kältemaschinenbau Unistat P405w Dynamisches Peter Huber Kältemaschinenbau Unistat P425 Dynamisches Peter Huber Kältemaschinenbau Unistat P425w Dynamisches Peter Huber Kältemaschinenbau Unistat P430 Dynamisches Peter Huber Kältemaschinenbau Unistat P430w Dynamisches Peter Huber Kältemaschinenbau Unistat P510 Dynamisches Peter Huber Kältemaschinenbau Unistat P510w Dynamisches Peter Huber Kältemaschinenbau Unistat P515w Dynamisches Peter Huber Kältemaschinenbau Unistat P520 Dynamisches Peter Huber Kältemaschinenbau Unistat P520w Dynamisches Peter Huber Kältemaschinenbau Unistat P525 Dynamisches Peter Huber Kältemaschinenbau Unistat P525w Dynamisches Peter Huber Kältemaschinenbau Unistat P527w Dynamisches Peter Huber Kältemaschinenbau Unistat P530w Dynamisches Peter Huber Kältemaschinenbau Unistat P540w Dynamisches Peter Huber Kältemaschinenbau Unistat P545w Dynamisches Peter Huber Kältemaschinenbau Unistat P610 Dynamisches Peter Huber Kältemaschinenbau Unistat P610w Dynamisches Peter Huber Kältemaschinenbau Unistat P615 Dynamisches Peter Huber Kältemaschinenbau Unistat P615w Dynamisches Peter Huber Kältemaschinenbau Unistat P620w Dynamisches Peter Huber Kältemaschinenbau Unistat P625w Dynamisches Peter Huber Kältemaschinenbau Unistat P630w Dynamisches Peter Huber Kältemaschinenbau Unistat P635w Dynamisches Peter Huber Kältemaschinenbau Unistat P645w Dynamisches Peter Huber Kältemaschinenbau Unistat P650w Dynamisches Peter Huber Kältemaschinenbau Unistat P815 Dynamisches Peter Huber Kältemaschinenbau Unistat P815w Dynamisches Peter Huber Kältemaschinenbau Unistat P825 Dynamisches Peter Huber Kältemaschinenbau Unistat P825w Dynamisches Peter Huber Kältemaschinenbau Unistat P905 Dynamisches Peter Huber Kältemaschinenbau Unistat P905w Dynamisches Peter Huber Kältemaschinenbau Unistat P912w Dynamisches Peter Huber Kältemaschinenbau Unistat P915w Dynamisches Peter Huber Kältemaschinenbau Unistat P920w Dynamisches Peter Huber Kältemaschinenbau Unistat P925w Dynamisches Peter Huber Kältemaschinenbau Unistat P930w Dynamisches Peter Huber Kältemaschinenbau Unistat P950w Dynamisches