Operating principles and functional overview of the unit
Procedurep. 10
Configuration and parameter setup procedures
Procedurep. 11
Communication interfaces and firmware update instructions
p. 12
Supplementary information and appendices
Overviewp. 13
Preface, safety notes and document conventions
Overviewpp. 14-21
Product introduction, intended use and technical basics
Procedurepp. 22-37
Installation, initial start-up and commissioning steps
Overviewpp. 38-50
Detailed functional description of operating modes
Procedurepp. 51-75
Detailed setup, calibration and configuration procedures
Procedurepp. 76-77
Standard operating procedures during normal use
Procedurepp. 78-85
Interface configuration and software update details
Procedurepp. 86-95
Maintenance, servicing and troubleshooting procedures
Procedurepp. 96-99
Safe shutdown and decommissioning instructions
Specpp. 100-102
Operating temperature range and technical specifications
p. 10table, 40 rows
p. 48table, 13 rows
p. 41table, 8 rows
p. 49table, 14 rows
p. 88table, 12 rows
p. 28table, 6 rows
p. 39table, 8 rows
p. 43table, 7 rows
p. 82table, 3 rows
p. 84table, 6 rows
Содержание (оглавление) раздела 4 Setup mode и раздела 5 Normal operation
Раздел
Название
Страница
4
Setup mode
51
4.1
Setup mode
51
4.1.1
Turning on the temperature control unit
51
4.1.2
Turning off the temperature control unit
51
4.1.3
Setting the overtemperature protection
52
4.1.3.1
General information on the overtemperature protection
52
4.1.3.2
Setting “OT limit: heating”
52
4.1.3.3
Setting “OT expansion vessel”
53
4.1.3.4
Setting “Process Safety”
53
4.1.3.5
Monitoring via “Display OT values”
53
4.1.4
Testing overtemperature protection for functionality
54
4.1.5
Adjusting the Delta T limiter
54
4.1.5.1
Changing the Delta T limiter
54
4.2
The temperature control circuit
54
4.2.1
Select temperature control: Internal or process
55
4.2.2
Temperature control to internal temperature
55
4.2.3
Temperature control to process temperature
56
4.2.4
Delta T limiter
56
4.2.5
Monitoring the Pt100 temperature sensors
56
4.2.6
Optimum control parameters for optimum temperature control
57
4.2.7
Sub-category: “Select auto/expert mode”.
57
4.2.8
Sub-category: “Configuration auto”
57
4.2.8.1
Sub-category: “Find parameters”
57
4.2.8.2
Sub-category: “Control Dynamics”
59
4.2.8.3
Sub-category: “Fluid Properties”
60
4.2.8.4
Sub-category: “Display parameters”
61
4.2.9
Sub-category: “Configuration manual”
62
4.2.9.1
Sub-category: “Change parameters”
62
4.2.9.2
Sub-category: “Display parameters”
63
4.2.9.3
Sub-category: “Control structure”
63
4.2.10
Sub-category: “Reset parameters”
63
4.2.11
Sub-category: “Display parameters”
64
4.2.12
Setting the setpoint thresholds
64
4.2.13
Setting the setpoint
65
4.3
Adjusting the cooling capacity
65
4.3.1
Cooling water connection
66
4.4
Filling, venting, degassing and draining
66
4.4.1
Externally closed application
66
4.4.1.1
Filling and venting externally closed application
67
4.4.1.2
Degassing externally closed applications
68
Restore factory settings — “X” = Value is reset, “–” = Value is not reset
Description
(A)
(B)
(C)
(D)
(E)
(F)
Acoustics
X
X
X
–
–
–
Display functions Warnings; Messages; Inactive Menu Items; Resolution Temperature
X
X
X
–
–
–
Display modes
X
X
X
–
–
–
AutoStart response
X
X
X
–
–
–
Limitations DeltaT limit; max. heating variable; max. cooling variable; max. permissible current consumption (for 230 V temperature control units)
X
X
X
–
–
–
Com.G@te (only if Com.G@te is connected) Analog Interface (configuration input) Meaning of AIF input; response on cable break; measurement range setting
X
X
X
–
–
X
Calibration
–
–
–
–
–
X
In case of analog error; current / voltage switching; filter constant
X
X
X
–
–
X
Configuration Output AIF output variable; Output range setting
(A) = All together; (B) = Device parameters without OT; (C) = Device Parameter; (D) Program Controller; (E) = Menu; (F) = Com.G@te
Overview of the E-grade variants
Temperature control units / E-grade
E-grade Basic
E-grade Exclusive
E-grade Professional
Brewing thermostats
–
X
O
Unistat temperature control units
–
–
X
UniCAL
–
–
X
Other temperature control units
X
O
O
E-grade “Explore” (only for “Unistat”-series temperature control units)
The E-grade includes E-grade “Professional” functionalities. In addition it includes: – Sensor calibration for return sensor – Return temperature display – Display of performance (estimated or calculated from volume flow) – Display mode “Explore” instead of “numerically large” – PB commands correspond to DV-E-grade, except: Temperatures only in 0.01 °C and volume flows only in 0.1 l/min.
–
–
O
E-grade “DV-E-grade” – All interface commands are enabled. – Temperatures in 0.001 C, volume flows in 0.001 l/min possible (cf. E-grade Explore) Note: Only the interface commands are enabled, not the corresponding menu items in Pilot ONE!
O
O
O
E-grade “OPC-UA” – OPC-UA interface via Ethernet. This E-grade additionally includes DV-E-grade functionalities.
O
O
O
X = Standard equipment, O = optional, – = not possible
Configuration manual parameters
Description
(A)
(B)
(C)
(D)
(E)
(F)
Automated compressor
X
X
X
–
–
–
Enter program
X
–
–
X
–
–
Pump settings Pump speed setpoint; Pump pressure setpoint; Control Mode
X
X
X
–
–
–
Fluid properties Thermal fluid; liter indicator
X
X
X
–
–
–
Protective function High Limit Alarm Internal; Low Limit Alarm Internal; High Limit Alarm Process; Low Limit Alarm Process; Hydrostatic Correction; Warning Time Limit (CC-E only); Min. Level (for temperature control units with analogue level sensor); Max. Level (for temperature controllers with analogue level sensor); Freeze Protection (optional)
X
X
X
–
–
–
Setpoint
X
X
X
–
–
–
Setpoint limitation Min. Setpoint; Max. Setpoint
X
X
X
–
–
–
Language
X
X
X
–
–
–
Temperature format
X
X
X
–
–
–
Temperature control mode
X
X
X
–
–
–
Favourites menu
X
–
–
–
X
–
Pilot ONE Ethernet IP address; Subnet Mask; Remote Access
X
X
X
–
–
–
Time format
X
X
X
–
–
–
2. Setpoint
X
X
X
–
–
–
Cooling*
Description
Maintenance interval
Comment
Person responsible
A/W
Check the power cable
Prior to switching on the temperature control unit or on relocation
Do not start the temperature control unit if the power cable is damaged.
Qualified electrician (BGV A3)
A
Clean the perforated sheet
As required
Clean the perforated sheet of the temperature control unit with a damp cloth
Operating company
A/W
Thermal fluid inspection
As required
–
Operating company and / or operators
U
Empty drip tray1
Monthly
→ Page 91, section »Emptying the drip tray (if present)«.
Operating company and / or operators
A/W
Check the mechanical seals (drip tray) 1
Monthly
→ Page 95, section »Inspect the mechanical seal«.
Operating company and / or operators
A
Check the liquefier fins
As required, after 3 months at the latest
→ Page 89, section »Clean liquefier fins (air-cooled temperature control unit)«.
Operating company and / or operators
W
Check the hat-type strainer (dirt trap)
As required, after 3 months at the latest
→ Page 90, section »Clean hat-type strainer (dirt trap) (water-cooled temperature control unit)«.
→ Page 54, section »Testing overtemperature protection for functionality«.
Operating company and / or operators
A/W
Check the temperature control unit for damage and stability
Every 12 months or after a change of location
–
Operating company and / or operators
W
Check the cooling water quality
Every 12 months
Descale the cooling water circuit as required. Documentation on water quality is available at: www.huber-online.com
Operating company and / or operators
A/W
Free-standing models: Check air filter mats
Determine according to ambient conditions.
Check all air filter mats of the temperature control unit. Replace the air filter mats as needed. → Page 17, section »Proper disposal of resources and consumables«.
Operating company and / or operators
A/W
Exchange safety-relevant electric and electromechanical components
20 years
Have the exchange only carried out by certified personnel (such as Huber service engineers). Please contact Customer Support. → Page 99, section »Contact data«.
Operating company
*A = Air cooling; W = Water cooling; U = Applicable only for Unistats 1 Not valid for magnetically coupled circulation pumps (entry "MC pump" in technical data sheet). Magnetically coupled circulation pumps are maintenance-free. Circulation pumps with mechanical seal have no entry under "Circulation pump" in the technical data sheet.
Side / Distance in cm (for operation in a tub)
Side
Air cooling
Water cooling
[A1] Top
Air outlet on top of unit: free standing
–
[A2] Top
can be located under a bench
can be located under a bench
[B] Left
min. 20
min. 20
[C] Right
min. 20
min. 20
[D] Front
min. 20
min. 20
[E] Rear
min. 20
min. 20
Designation / Specification
Designation
Specification
Calcium carbonate per liter
≤ 1.5 mmol/l; corresponds to a water hardness of: ≤ 8.4 °dH (soft)
PH value
between 6.0 and 8.5
Ultrapure water, distillates
Add 0.1 g of sodium carbonate (Na2CO3) per liter
Not approved water
Distilled, deionized, demineralized, chloric, ferruginous, ammoniacal, or contaminated river water or sea water
Thermal fluid: Water without ethylene glycol
Use
(+15) +65 °C to +90 °C
Thermal fluid: Water-ethylene glycol mixture
Use
do not use
Function description
Function
E-grade Basic
E-grade Exclusive
E-grade Professional
Storage/loading temperature control programs onto USB stick
–
X
X
Copying settings
–
–
X
Service data recorders (flight recorder)
X
X
X
PB commands1
X
X
X
Communication Watchdog
–
–
X
Process data recording directly to a USB stick: Setpoint, Actual Value Internal and Actual Value Process / Heating Capacity %, Cooling Capacity %, and Pump Pressure / Pump Speed, and VPC Pressure
–/–/–
X/X/–
X/X/X
Calendar start
–
–
X
1 See Data Communication Manual. Anything controllable via the graphical user interface of the Pilot ONE can be controlled by PB commands.
Female RS232 Serial (with adapter cable) — pin assignment
Pin
Signal
Description
2
RxD
Receive Data
3
TxD
Transmit Data
5
GND
Signal GND
Jack AIF Reg-E-Prog
Pin
Description
Signal
1
Current output, T external
0 a)/4 - 20 mA or 0 - 10 V
If using 0 - 10 V, integrate a
500 Ω resistor.
2
Current output, setpoint
0 a)/4 - 20 mA or 0 - 10 V
3
GND for analog outputs
GND
4
Analog input (programmable)
0 a)/4 - 20 mA or 0 - 10 V
Power input: 200 Ω working
resistance
Voltage input: 100 kΩ input
resistance
5
Current output, freely programmable
0 a)/4 - 20 mA or 0 - 10 V
If using 0 - 10 V, integrate a
500 Ω resistor.
6
GND for analog input
GND
a)Consult with our Customer Support.
Engineering Facts
14 facts · every claim linked to a page
Safety
The unit must not be installed or started inside an ATEX zone; explosion protection requires an Ex px cabinet.
Determines whether the unit can be placed in a hazardous area at all.
Source
“Do NOT install or start up the temperature control unit within an ATEX zone.”
Installation
The unit must be transported and stored upright only; horizontal transport damages the compressor.
Affects rigging, pallet handling and site logistics during installation.
Source
“Only transport the temperature control unit in an upright position.”
Installation
Only an emptied unit may be moved; filled transport risks thermal fluid overflow.
Dictates drain-before-move procedure and spill risk planning.
Source
“Only transport an emptied temperature control unit.”
Installation
Casters are not for transport and carry only 25% of total mass symmetrically; use an industrial truck.
Prevents using casters for relocation and informs floor/rigging planning.
Source
“The casters (if present) on the temperature control unit are not suitable for transport.”
Installation
Leveling feet must be unscrewed/activated and parking brakes engaged before commissioning.
Ensures stable, level installation and prevents creeping during operation.
Source
“The parking brakes at the casters (if present) must be activated and/or the leveling feet (if pre-sent) must be unscrewed/activated before the temperature control unit is put into operation.”
Installation
Allow about 2 hours acclimation when moving between cold and warm environments before power-on.
Prevents condensation-induced electrical faults at first start-up.
Source
“Allow the temperature control unit to acclimate for about 2 hours when changing from a cold to a warm environment (or vice versa). Do not turn on the temperature control unit before!”
Installation
For externally closed applications the expansion vessel must remain unlocked; locking it causes overpressure damage.
Critical valve setting depending on whether the application is open or closed.
Source
“Locking the >Expansion vessel< [18] will damage the externally closed application due to overpressure.”
Installation
For externally open applications the expansion vessel must be locked to prevent fluid loss.
Prevents leakage and ensures correct volume compensation for bath applications.
Source
“The >Expansion vessel< [18] must be locked!”
Installation
Hard-wired power connection must be done by an electrician; supply voltage/frequency must match the rating plate.
Defines qualification requirement and verification step for electrical hook-up.
Source
“Have the connection/adjustment to the power supply carried out by an electrician.”
Installation
At 230 V the max current is just under 16 A and the unit can be limited to 13 A or 10 A at first start-up.
Determines upstream fuse/breaker sizing and possible power derating.
Source
“The maximum current consumption at 230 V operation is just below 16 A.”
Operation
Process Safety mode requires the compressor set to Permanently ON to keep cooling active at the OT trigger.
Affects controller configuration for exothermic process protection.
Source
“This mode requires the compressor to be set to Permanently ON”
Operation
Power Failure Auto resumes temperature control automatically after a power outage if it was active.
Influences restart strategy and unattended operation risk assessment.
Source
“If temperature control is active during a power outage, the process will automatically continue after the power outage.”
Maintenance
Cooling water supply should be shut off even for short downtimes such as overnight.
Prevents water damage and unnecessary load on the cooling water circuit.
Source
“Shut off the cooling water supply to the temperature control unit even for shorter downtimes (e.g. overnight).”
Maintenance
Hoses and connections must be periodically checked for leaks and quality and replaced as needed.
Defines the preventive maintenance interval scope for the fluid circuit.
Source
“Check periodically for leaks and the quality of the hose and hose connections and take suitable measures (replace) as required.”